LeserbriefOpen Access · CC BY-NC 4.0

Leadless epicardial pacing at the left ventricular apex: an animal study

David Backhoff iD, Matthias J Müller iD, Yannic Wilberg iD, Katja Eildermann iD, Thomas Paul iD, Dieter Zenker iD, Ulrich Krause iD

Europace · 2023

DOI: 10.1093/europace/euad303 ↗

Kurz gefasst

Leserbrief in der Zeitschrift Europace, 2023. Verfasst von David Backhoff und 6 weiteren Autor:innen. Beteiligte Einrichtung: Pediatric Heart Center, Justus-Liebig University Giessen, Feulgenstrasse 10-12, 35385 Gießen, Germany. Hauptthemen laut MEDLINE: Pacemaker, Artificial. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 9 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: Medtronic External Research.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Aims

State-of-the-art pacemaker implantation technique in infants and small children consists of pace/sense electrodes attached to the epicardium and a pulse generator in the abdominal wall with a significant rate of dysfunction during growth, mostly attributable to lead failure. In order to overcome lead-related problems, feasibility of epicardial implantation of a leadless pacemaker at the left ventricular apex in a growing animal model was studied. Methods and results Ten lambs (median body weight 26.8 kg) underwent epicardial implantation of a Micra transcatheter pacing system (TPS) pacemaker (Medtronic Inc., Minneapolis, USA). Using a subxyphoid access, the Micra was introduced through a short, thick-walled tube to increase tissue contact and to prevent tilting from the epicardial surface. The Micra's proprietary delivery system was firmly pressed against the heart, while the Micra was pushed forward out of the sheath allowing the tines to stick into the left ventricular apical epimyocardium. Pacemakers were programmed to VVI 30/min mode. Pacemaker function and integrity was followed for 4 months after implantation. After implantation, median intrinsic R-wave amplitude was 5 mV [interquartile range (IQR) 2.8–7.5], and median pacing impedance was 2235 Ω (IQR 1725–2500), while the median pacing threshold was 2.13 V (IQR 1.25–2.9) at 0.24 ms. During follow-up, 6/10 animals had a significant increase in pacing threshold with loss of capture at maximum output at 0.24 ms in 2/10 animals. After 4 months, median R-wave amplitude had dropped to 2.25 mV (IQR 1.2–3.6), median pacing impedance had decreased to 595 Ω (IQR 575–645), and median pacing threshold had increased to 3.3 V (IQR 1.8–4.5) at 0.24 ms. Explantation of one device revealed deep penetration of the Micra device into the myocardium.

Conclusion

Short-term results after epicardial implantation of the Micra TPS at the left ventricular apex in lambs were satisfying. During mid-term follow-up, however, pacing thresholds increased, resulting in loss of capture in 2/10 animals. Penetration of one device into the myocardium was of concern. The concept of epicardial leadless pacing seems very attractive, and the current shape of the Micra TPS makes the device unsuitable for epicardial placement in growing organisms.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Pacemaker, ArtificialHumansChildAnimalsSheepTreatment OutcomeEquipment DesignHeart VentriclesMyocardiumCardiac Pacing, Artificial

Stichwörter der Autor:innen

ChildrenEpicardial pacingLeadless pacemakerMicra

Autor:innen und Institutionen

  • Pediatric Heart Center, Justus-Liebig University Giessen, Feulgenstrasse 10-12, 35385 Gießen, Germany.
  • Department of Pediatric Cardiology, Intensive Care Medicine and Neonatology, University Medical Center, Georg-August-University Göttingen, Robert-Koch Str. 40, 37075 Göttingen, Germany.
  • Department of Thoracic and Cardiac Surgery, University Medical Center, Georg-August-University Göttingen, Robert-Koch Str. 40, 37075 Göttingen, Germany.

ORCID-Profile: David Backhoff, Matthias J Müller, Yannic Wilberg, Katja Eildermann, Thomas Paul, Dieter Zenker, Ulrich Krause

Förderung

  • Medtronic External Research

Zitieren

APA

Backhoff, D., Müller, M. J., Wilberg, Y., Eildermann, K., Paul, T., Zenker, D., & Krause, U. (2023). Leadless epicardial pacing at the left ventricular apex: an animal study. Europace. https://doi.org/10.1093/europace/euad303

Vancouver

Backhoff D, Müller MJ, Wilberg Y, Eildermann K, Paul T, Zenker D, et al. Leadless epicardial pacing at the left ventricular apex: an animal study. Europace. 2023. doi:10.1093/europace/euad303

BibTeX
@article{Backhoff2023,
  author = {Backhoff, David and Müller, Matthias J and Wilberg, Yannic and Eildermann, Katja and Paul, Thomas and Zenker, Dieter and Krause, Ulrich},
  title = {Leadless epicardial pacing at the left ventricular apex: an animal study},
  journal = {Europace},
  year = {2023},
  doi = {10.1093/europace/euad303}
}
Kostenlos starten →