Safety and performance of vibration-assisted extraction of chronic cardiac implantable electronic device leads: A preclinical ovine study
Heart Rhythm · 2026
DOI: 10.1016/j.hrthm.2026.06.042 ↗Kurz gefasst
Fachartikel in der Zeitschrift Heart Rhythm, 2026. Verfasst von Yuval Ramot und 7 weiteren Autor:innen. Beteiligte Einrichtung: Faculty of Medicine, Hebrew University of Jerusalem, Jerusalem, Israel; Department of Dermatology, Hadassah Medical Center, Jerusalem, Israel. Hauptthemen laut MEDLINE: Device Lead Extraction, Vibration, Pacemaker, Artificial, Defibrillators, Implantable. Die Arbeit nennt 19 Literaturangaben.
Abstract (Originalsprache)
BACKGROUND: Transvenous extraction of chronic cardiac implantable electronic device leads is challenged by fibrotic adhesions and procedural risks. Novel technologies aim to improve safety and efficacy. A high-frequency mechanical vibration system aims to disrupt adhesions. OBJECTIVE: This study aimed to evaluate the initial safety, performance, and usability of a novel vibration-assisted lead extraction system in a preclinical Good Laboratory Practice ovine model. METHODS: 9 leads implanted for >2 years (range 2.9-5.25 years) were extracted from 6 sheep using the vibration system under fluoroscopy. The internal lumen of the leads was used to deliver a customized locking stylet. The stylet was used to deliver vibration energy to the lead, applied for up to 20 minutes/animal. Assessments included procedural success, lead integrity, intraprocedural monitoring, 3-day clinical follow-up, laboratory tests, necropsy, and histopathology (heart, lungs, central veins). RESULTS: Procedural success (lead removal) was 100% without adjunctive tools: external cutting sheath was not required for adhesion dissection in all of the cases. Lead release occurred within ∼1 minute in 8 of 9 leads. Primary safety endpoints were met: no central vein or myocardial perforation and no pulmonary embolism. No blood pressure drops, significant arrhythmias, or worsening tricuspid regurgitation occurred. 1 lead fracture (1 of 9) resulted in a retained distal screw (
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Faculty of Medicine, Hebrew University of Jerusalem, Jerusalem, Israel; Department of Dermatology, Hadassah Medical Center, Jerusalem, Israel.
- Faculty of Medicine, Hebrew University of Jerusalem, Jerusalem, Israel; Department of Cardiology, Hadassah Medical Center, Jerusalem, Israel.
- Pre-clinical Research and Development Unit, Shamir Medical Center, Zerifin, Israel.
- Consultant in Toxicologic Pathology, Tel Aviv and Tel Aviv University, Israel. Electronic address: anyska@nyska.net.
Zitieren
APA
Ramot, Y., Tomer, O., Zilberman Barzilai, D., Willenz, U., Beer, O., Tov, Y. S., Luria, D., & Nyska, A. (2026). Safety and performance of vibration-assisted extraction of chronic cardiac implantable electronic device leads: A preclinical ovine study. Heart Rhythm. https://doi.org/10.1016/j.hrthm.2026.06.042
Vancouver
Ramot Y, Tomer O, Zilberman Barzilai D, Willenz U, Beer O, Tov YS, et al. Safety and performance of vibration-assisted extraction of chronic cardiac implantable electronic device leads: A preclinical ovine study. Heart Rhythm. 2026. doi:10.1016/j.hrthm.2026.06.042
BibTeX
@article{Ramot2026,
author = {Ramot, Yuval and Tomer, Orr and Zilberman Barzilai, Deniz and Willenz, Udi and Beer, Oren and Tov, Yariv Siman and Luria, David and Nyska, Abraham},
title = {Safety and performance of vibration-assisted extraction of chronic cardiac implantable electronic device leads: A preclinical ovine study},
journal = {Heart Rhythm},
year = {2026},
doi = {10.1016/j.hrthm.2026.06.042}
}
EUVIMED