Pulsed field ablation using focal contact force-sensing catheters for treatment of atrial fibrillation: acute and 90-day invasive remapping results
Europace · 02.06.2023 · 25 (6)
DOI: 10.1093/europace/euad147 ↗Kurz gefasst
Multizentrische Studie in der Zeitschrift Europace, online erschienen am 19.06.2023. Verfasst von Ante Anić und 9 weiteren Autor:innen. Beteiligte Einrichtung: Heart and vascular center, Klinički Bolnički Centar (KBC) Split, Spinčićeva ul. 1, Split 21000, Croatia. Hauptthemen laut MEDLINE: Atrial Fibrillation, Catheter Ablation, Pulmonary Veins. Die Arbeit wurde bisher 77-mal zitiert (Crossref) und nennt 54 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: Galvanize Therapeutics.
Abstract (Originalsprache)
Aims
Pulsed field ablation (PFA) has emerged as a promising alternative to thermal ablation for treatment of atrial fibrillation (AF). We report performance and safety using the CENTAURI™ System (Galvanize Therapeutics) with three commercial, focal ablation catheters. Methods and results ECLIPSE AF (NCT04523545) was a prospective, single-arm, multi-centre study evaluating safety and acute and chronic pulmonary vein isolation (PVI) durability using the CENTAURI System in conjunction with the TactiCath SE, StablePoint, and ThermoCool ST ablation catheters. Patients with paroxysmal or persistent AF were treated at two centres. Patients were analysed in five cohorts based upon ablation settings, catheter, and mapping system. Pulsed field ablation was performed in 82 patients (74% male, 42 paroxysmal AF). Pulmonary vein isolation was achieved in 100% of pulmonary veins (322/322) with first-pass isolation in 92.2% (297/322). There were four serious adverse events of interest (three vascular access complications and one lacunar stroke). Eighty patients (98%) underwent invasive remapping. Pulsed field ablation development Cohorts 1 and 2 showed a per-patient isolation rate of 38% and 26% and a per-PV isolation rate of 47% and 53%, respectively. Optimized PFA Cohorts 3–5 showed a per-patient isolation rate of 60%, 73%, and 81% and a per-PV isolation rate of 84%, 90%, and 92%, respectively.
Conclusion
ECLIPSE AF demonstrated that optimized PFA using the CENTAURI System with three commercial, contact force-sensing, solid-tip focal ablation catheters resulted in transmural lesion formation and high proportion of durable PVI with a favourable safety profile, thus providing a viable treatment option for AF that integrates with contemporary focal ablation workflows.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Heart and vascular center, Klinički Bolnički Centar (KBC) Split, Spinčićeva ul. 1, Split 21000, Croatia.
- Jessa Ziekenhuis, Hasselt, Belgium.
- Galvanize Therapeutics, Inc. (previously Galaxy Medical), Redwood City, CA, USA.
ORCID-Profile: Ante Anić, Thomas Phlips, Toni Brešković, Pieter Koopman, Vikramaditya Mediratta, Zrinka Jurišić, Lucija Lisica, Johan Vijgen
Förderung
- Galvanize Therapeutics
Zitieren
APA
Anić, A., Phlips, T., Brešković, T., Koopman, P., Girouard, S., Mediratta, V., Jurišić, Z., Sikirić, I., Lisica, L., & Vijgen, J. (2023). Pulsed field ablation using focal contact force-sensing catheters for treatment of atrial fibrillation: acute and 90-day invasive remapping results. Europace, 25(6). https://doi.org/10.1093/europace/euad147
Vancouver
Anić A, Phlips T, Brešković T, Koopman P, Girouard S, Mediratta V, et al. Pulsed field ablation using focal contact force-sensing catheters for treatment of atrial fibrillation: acute and 90-day invasive remapping results. Europace. 2023;25(6). doi:10.1093/europace/euad147
BibTeX
@article{Ani2023,
author = {Anić, Ante and Phlips, Thomas and Brešković, Toni and Koopman, Pieter and Girouard, Steven and Mediratta, Vikramaditya and Jurišić, Zrinka and Sikirić, Ivan and Lisica, Lucija and Vijgen, Johan},
title = {Pulsed field ablation using focal contact force-sensing catheters for treatment of atrial fibrillation: acute and 90-day invasive remapping results},
journal = {Europace},