EUVIMEDEUVIMED entdecken
Klinische StudieLeserbriefStudienprotokollOpen Access · CC BY-NC 4.0

Deep sedation protocol during atrial fibrillation ablation using a novel variable-loop biphasic pulsed field ablation catheter

Massimo Grimaldi iD, Federico Quadrini iD, Nicola Caporusso, Federica Troisi iD, Nicola Vitulano, Vito Delmonte, Antonio Di Monaco iD

Europace · 2023

DOI: 10.1093/europace/euad222 ↗

Kurz gefasst

Klinische Studie in der Zeitschrift Europace, 2023. Verfasst von Massimo Grimaldi und 6 weiteren Autor:innen. Beteiligte Einrichtung: Dipartimento di Cardiologia, Ospedale Generale Regionale 'F. Miulli', Acquaviva delle Fonti, Bari, Italy. Hauptthemen laut MEDLINE: Atrial Fibrillation, Catheter Ablation, Deep Sedation, Pulmonary Veins. Die Arbeit wurde bisher 50-mal zitiert (Crossref) und nennt 15 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Pulsed field ablation (PFA) is a novel, largely nonthermal ablative modality that, by virtue of its putative preferential action on myocardial tissue through the process of irreversible electroporation (IRE), may replace conventional thermal ablation for atrial fibrillation (AF). The recent inspIRE study confirmed safety and effectiveness of a fully integrated biphasic PFA system with a variable loop circular catheter for the treatment of paroxysmal AF. The majority of PFA procedures were performed using general anesthesia. However, due to the risks of general anesthesia we report the data regarding our sedation protocol used during inspIRE study.

Methods

A total of 29 patients (mean age 55±9 years; 72% male) were enrolled as part of this analysis within the inspIRE trial. The sedation protocol is reported in the manuscript. The Richmond Agitation-Sedation Scale (RASS), the Visual Analogue Scale (VAS) and the Patient State Index were collected during sedation. Each patient was monitored using the Masimo Sedline. At the end of ablation, the Likert Scale Questionnaire (LSQ) was used to assess the patients' satisfaction with intraoperative analgesia-sedation.

Results

No procedural complications were documented. Sufficient oxygen saturation was maintained in all patients during procedure. Non-invasive ventilation or tracheal intubation were not required for any patient. The RAAS score between -1 to -5 was obtained in 27 patients (93%) while the value 0 was obtained in 2 patients (7%). The VAS score between 0 to 2 was obtained in 24 patients (83%); the VAS score 3 in 3 patients (10%) and the VAS score 4 in 2 patients (7%). The PSI score <50 was achieved in 16 patients (55%) while the PSI between 50 and 70 was achieved in 9 patients (31%). Positive answers to LSQ were obtained in most patients.

Conclusion

During PFA ablation procedures with the variable-loop circular catheter and its accompanying biphasic pulse, our deep sedation protocol is a valid alternative to general anesthesia.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Atrial FibrillationCatheter AblationDeep SedationPulmonary VeinsFemaleHumansMaleMiddle AgedCathetersHeart RateTreatment Outcome

Stichwörter der Autor:innen

Atrial fibrillationCatheter ablationElectroporationPulmonary vein isolationPulsed field ablationSedation protocol

Autor:innen und Institutionen

  • Dipartimento di Cardiologia, Ospedale Generale Regionale 'F. Miulli', Acquaviva delle Fonti, Bari, Italy.
  • Reparto di Anestesia e Rianimazione, Ospedale Generale Regionale 'F. Miulli', Acquaviva delle Fonti, Bari, Italy.

ORCID-Profile: Massimo Grimaldi, Federico Quadrini, Federica Troisi, Antonio Di Monaco

Zitieren

APA

Grimaldi, M., Quadrini, F., Caporusso, N., Troisi, F., Vitulano, N., Delmonte, V., & Di Monaco, A. (2023). Deep sedation protocol during atrial fibrillation ablation using a novel variable-loop biphasic pulsed field ablation catheter. Europace. https://doi.org/10.1093/europace/euad222

Vancouver

Grimaldi M, Quadrini F, Caporusso N, Troisi F, Vitulano N, Delmonte V, et al. Deep sedation protocol during atrial fibrillation ablation using a novel variable-loop biphasic pulsed field ablation catheter. Europace. 2023. doi:10.1093/europace/euad222

BibTeX
@article{Grimaldi2023,
  author = {Grimaldi, Massimo and Quadrini, Federico and Caporusso, Nicola and Troisi, Federica and Vitulano, Nicola and Delmonte, Vito and Di Monaco, Antonio},
  title = {Deep sedation protocol during atrial fibrillation ablation using a novel variable-loop biphasic pulsed field ablation catheter},
  journal = {Europace},
  year = {2023},
  doi = {10.1093/europace/euad222}
}
Kostenlos starten →