EUVIMEDEUVIMED entdecken
Multizentrische StudieOpen Access · CC BY 4.0

Application of a circular-shaped pulsed field ablation catheter with magnetic sensors for pulmonary vein isolation: a multi-centre clinical study report

Yan Wang iD, Heng Li Lai iD, Qi Chen iD, Hao Liu iD, Qi Ming Liu iD, Wei Bin Huang iD, Yu Tao, Qingmei Xiong iD, Ning Zhou iD, Chunxia Zhao iD, Jie Qiu, Ziqin Xu und 1 weitere

Europace · 2024

DOI: 10.1093/europace/euae068 ↗

Kurz gefasst

Multizentrische Studie in der Zeitschrift Europace, 2024. Verfasst von Yan Wang und 12 weiteren Autor:innen. Beteiligte Einrichtung: Division of Cardiology, Department of Internal Medicine, Tongji Hospital, Tongji Medical College, Huazhong University of Science and Technology, Jief…. Hauptthemen laut MEDLINE: Pulmonary Veins, Atrial Fibrillation, Catheter Ablation. Die Arbeit wurde bisher 11-mal zitiert (Crossref) und nennt 23 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Sichuan Jinjiang Electronic Medical Device Technology Co., Ltd, Chengdu, China.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Aims

A few studies have reported the effect and safety of pulsed field ablation (PFA) catheters for ablating atrial fibrillation (AF), which were mainly based on basket-shaped or flower-shaped designs. However, the clinical application of a circular-shaped multi-electrode catheter with magnetic sensors is very limited. To study the efficacy and safety of a PFA system in patients with paroxysmal AF using a circular-shaped multi-electrode catheter equipped with magnetic sensors for pulmonary vein isolation (PVI). Methods and results A novel proprietary bipolar PFA system was used for PVI, which utilized a circular-shaped multi-electrode catheter with magnetic sensors and allowed for three-dimensional model reconstruction, mapping, and ablation in one map. To evaluate the efficacy, efficiency, and safety of this PFA system, a prospective, multi-centre, single-armed, pre-market clinical study was performed. From July 2021 to December 2022, 151 patients with paroxysmal AF were included and underwent PVI. The study examined procedure time, immediate success rate, procedural success rate at 12 months, and relevant complications. In all 151 patients, all the pulmonary veins were acutely isolated using the studied system. Pulsed field ablation delivery was 78.4 ± 41.8 times and 31.3 ± 16.7 ms per patient. Skin-to-skin procedure time was 74.2 ± 29.8 min, and fluoroscopy time was 13.1 ± 7.6 min. The initial 11 (7.2%) cases underwent procedures with deep sedation anaesthesia, and the following cases underwent local anaesthesia. In the initial 11 cases, 4 cases (36.4%) presented transient vagal responses, and the rest were all successfully preventatively treated with atropine injection and rapid fluid infusion. No severe complications were found during or after the procedure. During follow-up, 3 cases experienced atrial flutter, and 11 cases had AF recurrence. The estimated 12-month Kaplan–Meier of freedom from arrhythmia was 88.4%.

Conclusion

The PFA system, comprised of a circular PFA catheter with magnetic sensors, could rapidly achieve PVI under three-dimensional guidance and demonstrated excellent safety with comparable effects.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Pulmonary VeinsAtrial FibrillationCatheter AblationHumansTreatment OutcomeProspective StudiesCathetersMagnetic PhenomenaRecurrence

Stichwörter der Autor:innen

Atrial fibrillationCatheter ablationElectroporationPulmonary vein isolationPulsed field ablation

Autor:innen und Institutionen

  • Division of Cardiology, Department of Internal Medicine, Tongji Hospital, Tongji Medical College, Huazhong University of Science and Technology, Jiefang Avenue 1095#, 430030 Wuhan, China.
  • Division of Cardiology, Jiangxi Provincial People's Hospital, Nanchang, China.
  • Department of Cardiovascular Medicine, The Second Affiliated Hospital of Nanchang University, Nanchang, China.
  • Division of Cardiology, The Second Affiliated Hospital of Guangxi Medical University, Nanning, China.
  • Division of Cardiology, Xiamen University Affiliated Zhongshan Hospital, Xiamen, China.
  • Division of Cardiology, The Second Xiangya Hospital of Central South University, Changsha, China.

ORCID-Profile: Yan Wang, Heng Li Lai, Qi Chen, Hao Liu, Qi Ming Liu, Wei Bin Huang, Qingmei Xiong, Ning Zhou, Chunxia Zhao, Dao Wen Wang

Förderung

  • Sichuan Jinjiang Electronic Medical Device Technology Co., Ltd, Chengdu, China

Zitieren

APA

Wang, Y., Lai, H. L., Chen, Q., Liu, H., Liu, Q. M., Huang, W. B., Tao, Y., Xiong, Q., Zhou, N., Zhao, C., Qiu, J., Xu, Z., & Wang, D. W. (2024). Application of a circular-shaped pulsed field ablation catheter with magnetic sensors for pulmonary vein isolation: a multi-centre clinical study report. Europace. https://doi.org/10.1093/europace/euae068

Vancouver

Wang Y, Lai HL, Chen Q, Liu H, Liu QM, Huang WB, et al. Application of a circular-shaped pulsed field ablation catheter with magnetic sensors for pulmonary vein isolation: a multi-centre clinical study report. Europace. 2024. doi:10.1093/europace/euae068

BibTeX
@article{Wang2024,
  author = {Wang, Yan and Lai, Heng Li and Chen, Qi and Liu, Hao and Liu, Qi Ming and Huang, Wei Bin and Tao, Yu and Xiong, Qingmei and Zhou, Ning and Zhao, Chunxia and Qiu, Jie and Xu, Ziqin and Wang, Dao Wen},
  title = {Application of a circular-shaped pulsed field ablation catheter with magnetic sensors for pulmonary vein isolation: a multi-centre clinical study report},
  journal = {Europace},
  year = {2024},
  doi = {10.1093/europace/euae068}
}
Kostenlos starten →