Emergency department use of a high-sensitivity point-of-care troponin assay reduces length of stay: an implementation study preliminary report

John W Pickering iD, Laura R Joyce iD, Christopher M Florkowski, Vanessa Buchan, Laura Hamill iD, Martin P Than iD

European Heart Journal: Acute Cardiovascular Care · 2024

DOI: 10.1093/ehjacc/zuae114 ↗

Kurz gefasst

Fachartikel in der Zeitschrift European Heart Journal: Acute Cardiovascular Care, 2024. Verfasst von John W Pickering und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medicine, University of Otago Christchurch, Christchurch, New Zealand. Hauptthemen laut MEDLINE: Emergency Service, Hospital, Length of Stay, Myocardial Infarction, Point-of-Care Systems. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 13 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NZ government Health Research Grant, Emergency Care Foundation.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Aims

Point-of-care (POC) high-sensitivity troponin (hs-cTn) assays within a clinical pathway may safely reduce length of stay (LoS) for patients presenting to the emergency department (ED) with possible acute myocardial infarction (AMI). In this early report, we present the first evaluation of a POC hs-cTn in real-life care. Methods and results In adult patients presenting to ED investigated for possible AMI, we compared the LoS in patients assessed with a troponin in the 8 weeks before (usual-care phase) and the 8 weeks following introduction of the Siemens Atellica VTLi POC hs-cTnI for decision-making (intervention phase). The VTLi replaced the laboratory (Beckman Coulter) assay as the default hs-cTn test within the clinical pathway. This was the only change to the pathway process. The safety outcome was first event AMI or cardiac death within 30 days. There were 2376 presentations in the usual-care phase with 188 individuals with AMI and 2392 in the intervention phase with 198 AMI. In the intervention phase, there was a mean (95% CI) reduction in LoS of 32 min (22–41 min) compared with the usual-care phase. This represents 21.4 fewer patient-hours in the ED each day (1196 in the 8-week period). In both phases, the pathway correctly identified all cases of AMI at index attendance. There were four follow-up events (two usual-care, two intervention) within 30 days.

Conclusion

The deployment of a hs-cTn POC analyser into a large ED safely reduced length of stay. If translatable to other EDs, this could represent an important advancement to patient care.

Trial registration

Australia New Zealand Clinical Trials Registry, No. ACTRN12619001189112.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Emergency Service, HospitalLength of StayMyocardial InfarctionPoint-of-Care SystemsBiomarkersHumansMaleFemaleAgedMiddle AgedTroponin

Stichwörter der Autor:innen

Accelerated diagnostic pathwayAcute myocardial infarctionEmergency departmentEmergency roomHigh-sensitivity cardiac troponinPoint of care

Substanzen

BiomarkersTroponin

Autor:innen und Institutionen

  • Department of Medicine, University of Otago Christchurch, Christchurch, New Zealand.
  • Emergency Department, Christchurch Hospital, Private Bag 4710, Christchurch 8140, New Zealand.
  • Department of Surgery and Critical Care, University of Otago Christchurch, Christchurch, New Zealand.
  • Canterbury Health Laboratories, Christchurch Hospital, Christchurch, New Zealand.
  • 24hr Surgery, Pegasus Health (Charitable), 401 Madras Street, Christchurch, New Zealand.
  • Department of Emergency Medicine, University of Kansas Medical Center, 4000 Cambridge Street Mailstop 1019, Kansas City, KS 66160, USA.

ORCID-Profile: John W Pickering, Laura R Joyce, Laura Hamill, Martin P Than

Förderung

  • NZ government Health Research Grant
  • Emergency Care Foundation

Zitieren

APA

Pickering, J. W., Joyce, L. R., Florkowski, C. M., Buchan, V., Hamill, L., & Than, M. P. (2024). Emergency department use of a high-sensitivity point-of-care troponin assay reduces length of stay: an implementation study preliminary report. European Heart Journal: Acute Cardiovascular Care. https://doi.org/10.1093/ehjacc/zuae114

Vancouver

Pickering JW, Joyce LR, Florkowski CM, Buchan V, Hamill L, Than MP. Emergency department use of a high-sensitivity point-of-care troponin assay reduces length of stay: an implementation study preliminary report. European Heart Journal: Acute Cardiovascular Care. 2024. doi:10.1093/ehjacc/zuae114

BibTeX
@article{Pickering2024,
  author = {Pickering, John W and Joyce, Laura R and Florkowski, Christopher M and Buchan, Vanessa and Hamill, Laura and Than, Martin P},
  title = {Emergency department use of a high-sensitivity point-of-care troponin assay reduces length of stay: an implementation study preliminary report},
  journal = {European Heart Journal: Acute Cardiovascular Care},
  year = {2024},
  doi = {10.1093/ehjacc/zuae114}
}
Kostenlos starten →