Diagnostic performance of a novel wireless direct-conversion intraoral sensor versus complementary metal-oxide semiconductor and photostimulable storage phosphor systems for proximal caries detection: an ex vivo study
Dentomaxillofacial Radiology · 2026
DOI: 10.1093/dmfr/twag031 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Dentomaxillofacial Radiology, 2026. Verfasst von Matheus Barros-Costa und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Oral Diagnosis, Division of Oral Radiology, Piracicaba Dental School, University of Campinas, Piracicaba, Sao Paulo 13414-903, Brazil. Hauptthemen laut MEDLINE: Semiconductors, Dental Caries, Radiography, Dental, Digital, Wireless Technology. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 18 Literaturangaben. Gefördert von: Coordenação de Aperfeiçoamento de Pessoal de Nível Superior - Brasil (CAPES) - Finance Code 001, Coordenação de Aperfeiçoamento de Pessoal de Nível Superior – Brasil (CAPES) – Finance Code 001.
Abstract (Originalsprache)
Objectives
To compare the diagnostic accuracy of a new wireless direct-conversion digital intraoral sensor to current indirect-conversion complementary metal-oxide semiconductor (CMOS) and photostimulable storage phosphor (PSP) sensors for detecting proximal caries.
Methods
The sample in this ex vivo study were 28 human premolars with 21 micro-CT verified non-cavitated proximal dental carious lesions. Images were obtained using 3 digital intraoral systems: a wireless direct-conversion sensor, a wired indirect-conversion CMOS sensor, and a PSP sensor. A custom phantom was constructed providing a limited dental arch segment comprising a molar, first premolar, and canine. The premolar was substituted and the phantom imaged using each intraoral system with (filtered) and without (raw) the application of a manufacturer-recommended enhancements, providing 168 images. Eight oral radiologists assessed the images using a 5-point scale. Receiver operating characteristic analysis was performed to calculate sensitivity, specificity, and the area under the curve (AUC) and compared across modalities using 2-way analysis of variance (ANOVA) with post hoc Tukey’s test at an a priori level of significance of 0.05.
Results
The AUC ranged from 0.58 to 0.73 (sensitivity range: 0.21-0.68, specificity range: 0.69-0.92). The wireless direct-conversion sensor (0.72 ± 0.06) showed diagnostic performance comparable to CMOS (0.72 ± 0.06) and superior to PSP (0.58 ± 0.07) for raw images (P < .001). The direct-conversion sensor without image enhancement (0.68 ± 0.18) demonstrated the highest sensitivity, whereas the PSP without image enhancement (0.21 ± 0.09) had the lowest sensitivity. Application of image enhancement improved PSP performance (0.48 ± 0.16), approximating that of the other systems.
Conclusions
Based on this ex vivo study, the wireless direct-conversion and CMOS sensors demonstrated similar diagnostic performance, superior to PSP in unfiltered conditions. These findings should be interpreted in light of the ex vivo design and modest sample size.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Oral Diagnosis, Division of Oral Radiology, Piracicaba Dental School, University of Campinas, Piracicaba, Sao Paulo 13414-903, Brazil.
- Department of Diagnosis & Oral Health, University of Louisville School of Dentistry, Louisville, KY 40202, United States.
- Department of Biosciences, Piracicaba Dental School, University of Campinas, Piracicaba, São Paulo 13414-903, Brazil.
ORCID-Profile: Matheus Barros-Costa, Gustavo Machado Santaella, Gabriela Gavilan Hadid, Christiano Oliveira-Santos, Deborah Queiroz Freitas, Francisco Carlos Groppo
Förderung
- Coordenação de Aperfeiçoamento de Pessoal de Nível Superior - Brasil (CAPES) - Finance Code 001
- Coordenação de Aperfeiçoamento de Pessoal de Nível Superior – Brasil (CAPES) – Finance Code 001
Zitieren
APA
Barros-Costa, M., Santaella, G. M., Hadid, G. G., Oliveira-Santos, C., Freitas, D. Q., Groppo, F. C., & Scarfe, W. C. (2026). Diagnostic performance of a novel wireless direct-conversion intraoral sensor versus complementary metal-oxide semiconductor and photostimulable storage phosphor systems for proximal caries detection: an ex vivo study. Dentomaxillofacial Radiology. https://doi.org/10.1093/dmfr/twag031
Vancouver
Barros-Costa M, Santaella GM, Hadid GG, Oliveira-Santos C, Freitas DQ, Groppo FC, et al. Diagnostic performance of a novel wireless direct-conversion intraoral sensor versus complementary metal-oxide semiconductor and photostimulable storage phosphor systems for proximal caries detection: an ex vivo study. Dentomaxillofacial Radiology. 2026. doi:10.1093/dmfr/twag031
BibTeX
@article{BarrosCosta2026,
author = {Barros-Costa, Matheus and Santaella, Gustavo Machado and Hadid, Gabriela Gavilan and Oliveira-Santos, Christiano and Freitas, Deborah Queiroz and Groppo, Francisco Carlos and Scarfe, William C},
title = {Diagnostic performance of a novel wireless direct-conversion intraoral sensor versus complementary metal-oxide semiconductor and photostimulable storage phosphor systems for proximal caries detection: an ex vivo study},
journal = {Dentomaxillofacial Radiology},
year = {2026},
doi = {10.1093/dmfr/twag031}
}
EUVIMED