Impact of Rubber Dam Thickness On Image Quality Parameters in Photostimulable Phosphor Plates
Dentomaxillofacial Radiology · 2026
DOI: 10.1093/dmfr/twag012 ↗Kurz gefasst
Fachartikel in der Zeitschrift Dentomaxillofacial Radiology, 2026. Verfasst von Şelale Özel und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Dentomaxillofacial Radiology, Dentistry Faculty, Altınbaş University, Istanbul 34147, Turkey. Hauptthemen laut MEDLINE: Rubber Dams, X-Ray Intensifying Screens, Radiography, Dental, Digital.
Abstract (Originalsprache)
Objective
The aim of this study is to investigate the effects of different rubber dam thicknesses on image quality in two exposure protocols.
Methods
For the study, three rubber dam thicknesses were used: thin (0.14 mm), medium (0.18 mm), and heavy (0.22 mm). To mimic clinical conditions, a rubber dam was used in two layers. Exposure was performed using two different protocols: Protocol 1: 0.080 s exposure time, 65 kV, 7 mA, and Protocol 2: 0.160 s exposure time, 65 kV, 7 mA. For each thickness and protocol, 47 measurements were taken (n = 47). Radiographic images were exported in TIFF and analyzed using ImageJ software. The region of interest was determined, and signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), Michelson, and Weber contrast values were obtained. One-way ANOVA, post hoc Tukey, and intra-class correlation were used for statistical analysis.
Results
No statistical difference was detected for protocol 1 between rubber dam thicknesses in SNR, CNR, Michelson, and Weber contrasts (p > 0.05). For protocol 2 (longer exposure time), heavy and medium thickness had lower SNR and CNR values than the thin one (p < 0.05). Michelson and Weber contrasts were statistically changed in different thicknesses of rubber dam (p < 0.05). ICC values were good and excellent.
Conclusions
The thick rubber dam reduced SNR and CNR values; likewise, Michelson and Weber contrasts were changed, which pointed a reduced image quality and negatively affected object visibility. Short exposure times are recommended to maintain image quality in clinical situations requiring the use of a thick rubber dam.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Dentomaxillofacial Radiology, Dentistry Faculty, Altınbaş University, Istanbul 34147, Turkey.
- Department of Dentomaxillofacial Radiology, Dentistry Faculty, İstanbul University-Cerrahpaşa, Istanbul 34098, Turkey.
- Department of Endodontics, Dentistry Faculty, Süleyman Demirel University, Isparta 32200, Turkey.
ORCID-Profile: Şelale Özel, Hakan Amasya, Deniz Yanık Nalbantoğlu
Zitieren
APA
Özel, Ş., Amasya, H., & Yanık Nalbantoğlu, D. (2026). Impact of Rubber Dam Thickness On Image Quality Parameters in Photostimulable Phosphor Plates. Dentomaxillofacial Radiology. https://doi.org/10.1093/dmfr/twag012
Vancouver
Özel Ş, Amasya H, Yanık Nalbantoğlu D. Impact of Rubber Dam Thickness On Image Quality Parameters in Photostimulable Phosphor Plates. Dentomaxillofacial Radiology. 2026. doi:10.1093/dmfr/twag012
BibTeX
@article{zel2026,
author = {Özel, Şelale and Amasya, Hakan and Yanık Nalbantoğlu, Deniz},
title = {Impact of Rubber Dam Thickness On Image Quality Parameters in Photostimulable Phosphor Plates},
journal = {Dentomaxillofacial Radiology},
year = {2026},
doi = {10.1093/dmfr/twag012}
}
EUVIMED