Impact of Rubber Dam Thickness On Image Quality Parameters in Photostimulable Phosphor Plates

Şelale Özel iD, Hakan Amasya iD, Deniz Yanık Nalbantoğlu iD

Dentomaxillofacial Radiology · 2026

DOI: 10.1093/dmfr/twag012 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Dentomaxillofacial Radiology, 2026. Verfasst von Şelale Özel und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Dentomaxillofacial Radiology, Dentistry Faculty, Altınbaş University, Istanbul 34147, Turkey. Hauptthemen laut MEDLINE: Rubber Dams, X-Ray Intensifying Screens, Radiography, Dental, Digital.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

The aim of this study is to investigate the effects of different rubber dam thicknesses on image quality in two exposure protocols.

Methods

For the study, three rubber dam thicknesses were used: thin (0.14 mm), medium (0.18 mm), and heavy (0.22 mm). To mimic clinical conditions, a rubber dam was used in two layers. Exposure was performed using two different protocols: Protocol 1: 0.080 s exposure time, 65 kV, 7 mA, and Protocol 2: 0.160 s exposure time, 65 kV, 7 mA. For each thickness and protocol, 47 measurements were taken (n = 47). Radiographic images were exported in TIFF and analyzed using ImageJ software. The region of interest was determined, and signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), Michelson, and Weber contrast values were obtained. One-way ANOVA, post hoc Tukey, and intra-class correlation were used for statistical analysis.

Results

No statistical difference was detected for protocol 1 between rubber dam thicknesses in SNR, CNR, Michelson, and Weber contrasts (p > 0.05). For protocol 2 (longer exposure time), heavy and medium thickness had lower SNR and CNR values than the thin one (p < 0.05). Michelson and Weber contrasts were statistically changed in different thicknesses of rubber dam (p < 0.05). ICC values were good and excellent.

Conclusions

The thick rubber dam reduced SNR and CNR values; likewise, Michelson and Weber contrasts were changed, which pointed a reduced image quality and negatively affected object visibility. Short exposure times are recommended to maintain image quality in clinical situations requiring the use of a thick rubber dam.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Rubber DamsX-Ray Intensifying ScreensRadiography, Dental, DigitalSignal-To-Noise RatioHumans

Stichwörter der Autor:innen

CNRSNRcontrastimage qualityphotostimulable phosphor platesrubber dam

Autor:innen und Institutionen

  • Department of Dentomaxillofacial Radiology, Dentistry Faculty, Altınbaş University, Istanbul 34147, Turkey.
  • Department of Dentomaxillofacial Radiology, Dentistry Faculty, İstanbul University-Cerrahpaşa, Istanbul 34098, Turkey.
  • Department of Endodontics, Dentistry Faculty, Süleyman Demirel University, Isparta 32200, Turkey.

ORCID-Profile: Şelale Özel, Hakan Amasya, Deniz Yanık Nalbantoğlu

Zitieren

APA

Özel, Ş., Amasya, H., & Yanık Nalbantoğlu, D. (2026). Impact of Rubber Dam Thickness On Image Quality Parameters in Photostimulable Phosphor Plates. Dentomaxillofacial Radiology. https://doi.org/10.1093/dmfr/twag012

Vancouver

Özel Ş, Amasya H, Yanık Nalbantoğlu D. Impact of Rubber Dam Thickness On Image Quality Parameters in Photostimulable Phosphor Plates. Dentomaxillofacial Radiology. 2026. doi:10.1093/dmfr/twag012

BibTeX
@article{zel2026,
  author = {Özel, Şelale and Amasya, Hakan and Yanık Nalbantoğlu, Deniz},
  title = {Impact of Rubber Dam Thickness On Image Quality Parameters in Photostimulable Phosphor Plates},
  journal = {Dentomaxillofacial Radiology},
  year = {2026},
  doi = {10.1093/dmfr/twag012}
}
Kostenlos starten →