Comparison of radiographic imaging quality/accuracy using photostimulable phosphor plates and metal oxide semiconductor receptors
Journal of Dental Education · 2025
DOI: 10.1002/jdd.13812 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Journal of Dental Education, 2025. Verfasst von Roberto Schwerter-Medina und 4 weiteren Autor:innen. Beteiligte Einrichtung: Instituto de Odontoestomatología, Escuela de Odontología, Facultad de Medicina, Universidad Austral de Chile, Valdivia, Chile. Hauptthemen laut MEDLINE: Semiconductors, Radiography, Dental, Digital, X-Ray Intensifying Screens, Radiography, Dental. Die Arbeit nennt 16 Literaturangaben.
Abstract (Originalsprache)
Aim
To compare thequality and accuracy of radiographic images obtained by dentistry students from a Chilean university using photostimulable phosphor plates (PSP) and complementary metal oxide semiconductor (CMOS) receptors. Materials and methods An experimental study was carried out in which 31 dental students used PSP and CMOS receptors to acquire radiographic images with the aid of a phantom. The time required to generate a radiograph was recorded. Image quality included sharpness/definition and brightness/contrast analysis. Accuracy assessed placement error, angulation errors, and cone cutting. A three‐point scale was used for each variable: 0 = undiagnosable, 1 = acceptable diagnosis with minor errors, and 2 = perfect diagnostic quality.
Results
The required time to generate a radiograph was faster with PSP than with CMOS (1.43 ± 0.28 min and 1.52 ± 0.61 min, respectively). Image quality analysis revealed that PSP was superior in sharpness/definition and brightness/contrast, whereas no significant differences were observed compared to CMOS ( p > 0.05). Moreover, no differences were noticed in technique accuracy regarding angulation errors and cone cutting ( p > 0.05). However, placement errors with PSP had significantly higher average scores than CMOS (1.11 ± 0.77 and 0.67 ± 0.85, respectively; p < 0.01).
Conclusion
Dental students would benefit from being trained on both receptors to be able to adapt to a diversified workplace. Radiographic images obtained by dentistry students demonstrate that PSP and CMOS exhibit similar parameters regarding quality and accuracy, except in placement where PSP demonstrates greater accuracy.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Instituto de Odontoestomatología, Escuela de Odontología, Facultad de Medicina, Universidad Austral de Chile, Valdivia, Chile.
- Laboratorio de Polímeros, Instituto de Química, Facultad de Ciencias, Universidad Austral de Chile, Valdivia, Chile.
- Cirujano Dentista EDF, Servicio de Salud de Tarapacá, Iquique, Chile.
ORCID-Profile: Roberto Schwerter-Medina, Pablo Romero-Araya, Felipe-Rodrigo Aguilera
Zitieren
APA
Schwerter-Medina, R., Ríos-Kremer, S., Rosas, C., Romero-Araya, P., & Aguilera, F. R. (2025). Comparison of radiographic imaging quality/accuracy using photostimulable phosphor plates and metal oxide semiconductor receptors. Journal of Dental Education. https://doi.org/10.1002/jdd.13812
Vancouver
Schwerter-Medina R, Ríos-Kremer S, Rosas C, Romero-Araya P, Aguilera FR. Comparison of radiographic imaging quality/accuracy using photostimulable phosphor plates and metal oxide semiconductor receptors. Journal of Dental Education. 2025. doi:10.1002/jdd.13812
BibTeX
@article{SchwerterMedina2025,
author = {Schwerter-Medina, Roberto and Ríos-Kremer, Sebastián and Rosas, Cristian and Romero-Araya, Pablo and Aguilera, Felipe-Rodrigo},
title = {Comparison of radiographic imaging quality/accuracy using photostimulable phosphor plates and metal oxide semiconductor receptors},
journal = {Journal of Dental Education},
year = {2025},
doi = {10.1002/jdd.13812}
}
EUVIMED