Measurement of radiation dose and analysis of dose distribution in the thyroid and eye lens during digital panoramic tomography
Dentomaxillofacial Radiology · 2025
DOI: 10.1093/dmfr/twaf021 ↗Kurz gefasst
Fachartikel in der Zeitschrift Dentomaxillofacial Radiology, 2025. Verfasst von Franziska Schindler und 3 weiteren Autor:innen. Beteiligte Einrichtung: Clinic for Dental, Oral, and Maxillofacial Diseases, Polyclinic for Dental Prosthetics and Materials Science, University Medical Center of Johannes G…. Hauptthemen laut MEDLINE: Lens, Crystalline, Thyroid Gland, Radiation Dosage, Radiography, Panoramic. Die Arbeit nennt 23 Literaturangaben.
Abstract (Originalsprache)
This study investigates the radiation dose and its distribution in critical organs of the human head during the use of digital panoramic radiography devices. The aim was to quantify the doses accumulated in the thyroid and eye lenses, as well as to analyse the overall dose distribution across a selection of modern panoramic units. A female Alderson-Rando phantom, equipped with 3D-printed slabs and drilled holes for placing an ionisation chamber, was repeatedly exposed in four digital panoramic devices (Orthophos XG Plus, ProMax 2D Classic, Viso G7, and Veraview X800). Measurements were taken in both the orbital regions and the thyroid. Additionally, five radiochromic films were placed between the phantom slabs, from the orbital floor to the thyroid region, to assess dose distribution. Equivalent doses for the thyroid ranged between 0.30 µSv and 1.46 µSv, while the doses for the eye ranged from 0.88 µSv to 4.24 µSv. There were significant differences between the devices (pFriedman < 0.001). The dose distribution in the anterior part followed the typical trajectory of the rotational isocenter of the fan beam. The distribution varied across devices, with the highest peak observed in the vertical region of the external acoustic meatus. The absorbed doses to the thyroid and eye lenses in digital panoramic devices are generally low (in the microsievert range), and dose distribution varies between devices.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Clinic for Dental, Oral, and Maxillofacial Diseases, Polyclinic for Dental Prosthetics and Materials Science, University Medical Center of Johannes Gutenberg University Mainz, 55131 Mainz, Germany.
- Clinic and Policlinic for Oral and Maxillofacial Surgery, Plastic Operations, University Medical Center of Johannes Gutenberg University Mainz, 55131 Mainz, Germany.
- Clinic and Policlinic for Radiooncology and Radiation Therapy, University Medical Center of Johannes Gutenberg University Mainz, 55131 Mainz, Germany.
- Division of Oral Diagnostic Sciences, University of Bern, 3010 Bern, Switzerland.
Zitieren
APA
Schindler, F., Müller-Winter, D., Karle, H., & Schulze, R. (2025). Measurement of radiation dose and analysis of dose distribution in the thyroid and eye lens during digital panoramic tomography. Dentomaxillofacial Radiology. https://doi.org/10.1093/dmfr/twaf021
Vancouver
Schindler F, Müller-Winter D, Karle H, Schulze R. Measurement of radiation dose and analysis of dose distribution in the thyroid and eye lens during digital panoramic tomography. Dentomaxillofacial Radiology. 2025. doi:10.1093/dmfr/twaf021
BibTeX
@article{Schindler2025,
author = {Schindler, Franziska and Müller-Winter, Daniel and Karle, Heiko and Schulze, Ralf},
title = {Measurement of radiation dose and analysis of dose distribution in the thyroid and eye lens during digital panoramic tomography},
journal = {Dentomaxillofacial Radiology},
year = {2025},
doi = {10.1093/dmfr/twaf021}
}
EUVIMED