EUVIMEDEUVIMED entdecken
Crossref

National dose survey and discussion on establishing diagnostic reference levels for dental imaging in Korea

Jo-Eun Kim iD, Han-Gyeol Yeom iD, Jae Joon Hwang iD, Yoon Joo Choi iD, Jin-Woo Han, Seo-Young An iD, Gyu-Tae Kim iD, Jae-Seo Lee iD, Jin-Soo Kim iD, Kyung-A Kim iD, Won-Jeong Han, Juhee Kang iD und 1 weitere

Dentomaxillofacial Radiology · 2025

DOI: 10.1093/dmfr/twaf014 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Dentomaxillofacial Radiology, 2025. Verfasst von Jo-Eun Kim und 12 weiteren Autor:innen. Beteiligte Einrichtung: Department of Oral and Maxillofacial Radiology, School of Dentistry and Dental Research Institute, Seoul National University, Daehak-ro 101, Jongro-g…. Hauptthemen laut MEDLINE: Radiation Dosage, Radiography, Dental, Diagnostic Reference Levels. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 18 Literaturangaben. Gefördert von: Korea Disease Control and Prevention Agency.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

This study aimed to establish updated diagnostic reference levels (DRLs) for dental imaging modalities in South Korea.

Methods

In cooperation with the Korea Disease Control and Prevention Agency, various types of institutions (dental clinics, dental hospitals, and dental university hospitals) were selected to investigate the status of diagnostic radiation equipment use. Subsequently, over 300 units were randomly selected for each imaging device type (intraoral, panoramic, and cone-beam CT [CBCT]) as measurement samples. DRLs were defined as the 75th percentile of the dose area product distribution. The differences in dose were analysed based on the type of institution, age of use, country of manufacture, and presence of a multifunction device.

Results

The national DRLs for dental imaging established in this survey were as follows: intraoral imaging at 48 mGy·cm2 for adults and 31 mGy·cm2 for children; panoramic imaging at 354 mGy·cm2 for adults and 224 mGy·cm2 for children; and CBCT at 1856 mGy·cm2 for adults and 1350 mGy·cm2 for children. Private dental clinics and hospitals recorded approximately twice the dose levels of university dental hospitals. CBCT devices in dental hospitals and those that have been in used for 5-10 years showed significantly high radiation doses.

Conclusions

The DRLs established through this study were found to be significantly increased, especially in adult and paediatric panoramic radiographs and paediatric CBCT images, compared with those in previous surveys; moreover, they were higher than those in other countries. The findings of this study can serve as a basis for national dose reduction efforts.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiation DosageRadiography, DentalDiagnostic Reference LevelsHumansRepublic of KoreaCone-Beam Computed TomographyAdultRadiography, PanoramicChildSurveys and Questionnaires

Stichwörter der Autor:innen

dental imagingdiagnostic reference leveldose area productnational survey

Autor:innen und Institutionen

  • Department of Oral and Maxillofacial Radiology, School of Dentistry and Dental Research Institute, Seoul National University, Daehak-ro 101, Jongro-gu, Seoul, 03080, Korea.
  • Department of Oral and Maxillofacial Radiology and Wonkwang Dental Research Institute, College of Dentistry, Wonkwang University, Iksan, 54538, Korea.
  • Department of Oral and Maxillofacial Radiology, School of Dentistry, Pusan National University, Yangsan, 50612, Korea.
  • Department of Oral and Maxillofacial Radiology, Yonsei University College of Dentistry, Seoul, 03722, Korea.
  • Department of Oral and Maxillofacial Radiology, College of Dentistry, Gangneung-Wonju National University, Gangneung, 25457, Korea.
  • Department of Oral and Maxillofacial Radiology, School of Dentistry, Kyungpook National University, Daegu, 41940, Korea.

ORCID-Profile: Jo-Eun Kim, Han-Gyeol Yeom, Jae Joon Hwang, Yoon Joo Choi, Seo-Young An, Gyu-Tae Kim, Jae-Seo Lee, Jin-Soo Kim, Kyung-A Kim, Juhee Kang, Min-Suk Heo

Förderung

  • Korea Disease Control and Prevention Agency

Zitieren

APA

Kim, J. E., Yeom, H. G., Hwang, J. J., Choi, Y. J., Han, J. W., An, S. Y., Kim, G. T., Lee, J. S., Kim, J. S., Kim, K. A., Han, W. J., Kang, J., & Heo, M. S. (2025). National dose survey and discussion on establishing diagnostic reference levels for dental imaging in Korea. Dentomaxillofacial Radiology. https://doi.org/10.1093/dmfr/twaf014

Vancouver

Kim JE, Yeom HG, Hwang JJ, Choi YJ, Han JW, An SY, et al. National dose survey and discussion on establishing diagnostic reference levels for dental imaging in Korea. Dentomaxillofacial Radiology. 2025. doi:10.1093/dmfr/twaf014