EUVIMEDEUVIMED entdecken
Crossref

Converting dose-area product to effective dose in dental cone-beam computed tomography using organ-specific deep learning

Ruben Pauwels iD

Dentomaxillofacial Radiology · 2025

DOI: 10.1093/dmfr/twae067 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Dentomaxillofacial Radiology, 2025. Verfasst von Ruben Pauwels. Beteiligte Einrichtung: Department of Dentistry and Oral Health, Aarhus University, 8000 Aarhus, Denmark. Hauptthemen laut MEDLINE: Cone-Beam Computed Tomography, Deep Learning, Radiation Dosage, Radiography, Dental. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 44 Literaturangaben. Gefördert von: European Union Horizon 2020 Research and Innovation, Marie Skłodowska-Curie, Aarhus University Research Foundation.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

To develop an accurate method for converting dose-area product (DAP) to patient dose for dental cone-beam computed tomography (CBCT) using deep learning.

Methods

A total of 24 384 CBCT exposures of an adult phantom were simulated with PCXMC 2.0, using permutations of tube voltage, filtration, source-isocenter distance, beam width/height, and isocenter position. Equivalent organ doses as well as DAP values were recorded. Next, using the aforementioned scan parameters as inputs, neural networks (NN) were trained using Keras for estimating the equivalent dose per DAP for each organ. Two methods were explored for positional input features: (1) “Coordinate” mode, which uses the (continuous) XYZ coordinates of the isocentre, and (2) “AP/JAW” mode, which uses the (categorical) anteroposterior and craniocaudal position. Each network was trained, validated, and tested using a 3/1/1 data split. Effective dose (ED) was calculated from the combination of NN outputs using ICRP 103 tissue weighting factors. The performance of the resulting NN models for estimating ED/DAP was compared with that of a multiple linear regression (MLR) model as well as direct conversion coefficients (CC).

Results

The mean absolute error (MAE) for organ dose/DAP on the test data ranged from 0.18% (bone surface) to 2.90% (oesophagus) in “Coordinate” mode and from 2.74% (red bone marrow) to 14.13% (brain) in “AP/JAW” mode. The MAE for ED was 0.23% and 4.30%, respectively, for the two modes, vs. 5.70% for the MLR model and 20.19%-32.67% for the CCs.

Conclusions

NNs allow for an accurate estimation of patient dose based on DAP in dental CBCT.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Cone-Beam Computed TomographyDeep LearningRadiation DosageRadiography, DentalHumansPhantoms, ImagingAdultNeural Networks, Computer

Stichwörter der Autor:innen

cone-beam computed tomographydeep learningdental radiographyradiation dosimetryradiation protection

Autor:innen und Institutionen

  • Department of Dentistry and Oral Health, Aarhus University, 8000 Aarhus, Denmark.
  • Aarhus Institute of Advanced Studies, Aarhus University, 8000 Aarhus, Denmark.
  • Department of Radiology, Faculty of Dentistry, Chulalongkorn University, 10330 Bangkok, Thailand.

ORCID-Profile: Ruben Pauwels

Förderung

  • European Union Horizon 2020 Research and Innovation
  • Marie Skłodowska-Curie
  • Aarhus University Research Foundation

Zitieren

APA

Pauwels, R. (2025). Converting dose-area product to effective dose in dental cone-beam computed tomography using organ-specific deep learning. Dentomaxillofacial Radiology. https://doi.org/10.1093/dmfr/twae067

Vancouver

Pauwels R. Converting dose-area product to effective dose in dental cone-beam computed tomography using organ-specific deep learning. Dentomaxillofacial Radiology. 2025. doi:10.1093/dmfr/twae067

BibTeX
@article{Pauwels2025,
  author = {Pauwels, Ruben},
  title = {Converting dose-area product to effective dose in dental cone-beam computed tomography using organ-specific deep learning},
  journal = {Dentomaxillofacial Radiology},
  year = {2025},
  doi = {10.1093/dmfr/twae067}
}
Kostenlos starten →