Methodological quality of dental cone-beam CT diagnostic reference levels: a systematic review
Dentomaxillofacial Radiology · 2026
DOI: 10.1093/dmfr/twag036 ↗Kurz gefasst
Systematische Übersicht in der Zeitschrift Dentomaxillofacial Radiology, 2026. Verfasst von Bharathraj Arunmani und 4 weiteren Autor:innen. Beteiligte Einrichtung: Division of Oral & Maxillofacial Surgery, Medicine, Pathology & Radiology, Dublin Dental University Hospital, Trinity College Dublin, Dublin, D02 F85…. Hauptthemen laut MEDLINE: Cone-Beam Computed Tomography, Diagnostic Reference Levels, Radiography, Dental. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 41 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
Objective
To evaluate the methodological quality of dental cone-beam CT (CBCT) diagnostic reference levels (DRLs) studies and reports, identify factors that limit DRL comparison, assess the level and methods of dose optimization, and clarify obstacles, both in dose optimization and DRL establishment.
Methods
A systematic search was applied across 5 databases and gray literature with International Atomic Energy Agency (IAEA) member countries contacted, to identify dental CBCT DRL studies and national reports. Studies/reports containing DRL establishment procedures, insufficient methodological detail, and adopted DRLs were excluded. Methodological quality was assessed using a modified Effective Public Health Practice Project tool, aligned with a checklist developed in accordance with International Commission on Radiological Protection (ICRP) 135.
Results
Of 320 reports screened, 19 were included. Diagnostic reference levels were clinical indication or field of view (FOV) based, using dose/kerma-area-product (DAP/KAP) metrics. Overall, 63% of studies had moderate methodological quality; DRL comparability was limited due to gaps or inconsistencies in reporting of exposure parameters, FOV, and heterogeneous indication grouping. Substantial dose variability was observed within the same clinical indications (2- to 10-fold) and FOV categories (3- to 10-fold), reflecting routine use of large FOVs irrespective of indication, reliance on manufacturer default protocols, limited understanding/education/training of exposure settings or optimization strategies, non-standardized system FOV settings, and unauthorized CBCT operators.
Conclusions
The majority of CBCT DRL studies are not of high quality. Improving DRL reporting quality requires interdisciplinary collaboration; digitized-dose recording with robust retrospective management of protocols, dose, indications; national dose registries for periodic dose audit; improved clinician education/engagement, strengthened regulatory oversight, and potentially international standardized frameworks.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Division of Oral & Maxillofacial Surgery, Medicine, Pathology & Radiology, Dublin Dental University Hospital, Trinity College Dublin, Dublin, D02 F859, Ireland.
- Division of Dentistry, School of Medical Sciences, University of Manchester, Manchester, M15 6FH, United Kingdom.
- Department of Medical Physics & Bioengineering, St James's Hospital, Dublin, D08 NHY1, Ireland.
ORCID-Profile: Bharathraj Arunmani, Ronan Coleman, Isabel Fleischmann, Margarete B McGuigan
Zitieren
APA
Arunmani, B., Horner, K., Coleman, R., Fleischmann, I., & McGuigan, M. B. (2026). Methodological quality of dental cone-beam CT diagnostic reference levels: a systematic review. Dentomaxillofacial Radiology. https://doi.org/10.1093/dmfr/twag036
Vancouver
Arunmani B, Horner K, Coleman R, Fleischmann I, McGuigan MB. Methodological quality of dental cone-beam CT diagnostic reference levels: a systematic review. Dentomaxillofacial Radiology. 2026. doi:10.1093/dmfr/twag036
BibTeX
@article{Arunmani2026,
author = {Arunmani, Bharathraj and Horner, Keith and Coleman, Ronan and Fleischmann, Isabel and McGuigan, Margarete B},
title = {Methodological quality of dental cone-beam CT diagnostic reference levels: a systematic review},
journal = {Dentomaxillofacial Radiology},
year = {2026},
doi = {10.1093/dmfr/twag036}
}