Ultra-low-dose photon-counting CT of paranasal sinus: an in vivo comparison of radiation dose and image quality to cone-beam CT
Dentomaxillofacial Radiology · 08.02.2024 · 53 (2) · 103-108
DOI: 10.1093/dmfr/twad010 ↗Kurz gefasst
Fachartikel in der Zeitschrift Dentomaxillofacial Radiology, online erschienen am 13.12.2023. Verfasst von Hanns Leonhard Kaatsch und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology and Neuroradiology, Bundeswehr Central Hospital Koblenz, Koblenz 56072, Germany. Hauptthemen laut MEDLINE: Tomography, X-Ray Computed, Paranasal Sinuses. Die Arbeit wurde bisher 21-mal zitiert (Crossref) und nennt 41 Literaturangaben.
Abstract (Originalsprache)
Purpose
This study investigated the differences in subjective and objective image parameters as well as dose exposure of photon-counting CT (PCCT) compared to cone-beam CT (CBCT) in paranasal sinus imaging for the assessment of rhinosinusitis and sinonasal anatomy.
Methods
This single-centre retrospective study included 100 patients, who underwent either clinically indicated PCCT or CBCT of the paranasal sinus. Two blinded experienced ENT radiologists graded image quality and delineation of specific anatomical structures on a 5-point Likert scale. In addition, contrast-to-noise ratio (CNR) and applied radiation doses were compared among both techniques.
Results
Image quality and delineation of bone structures in paranasal sinus PCCT was subjectively rated superior by both readers compared to CBCT (P < .001). CNR was significantly higher for photon-counting CT (P < .001). Mean effective dose for PCCT examinations was significantly lower than for CBCT (0.038 mSv ± 0.009 vs. 0.14 mSv ± 0.011; P < .001).
Conclusion
In a performance comparison of PCCT and a modern CBCT scanner in paranasal sinus imaging, we demonstrated that first-use PCCT in clinical routine provides higher subjective image quality accompanied by higher CNR at close to a quarter of the dose exposure compared to CBCT.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiology and Neuroradiology, Bundeswehr Central Hospital Koblenz, Koblenz 56072, Germany.
- Department of Vascular Surgery and Endovascular Surgery, Bundeswehr Central Hospital, Koblenz 56072, Germany.
- Bundeswehr Institute of Radiobiology affiliated to Ulm University, Munich 80937, Germany.
- Department of Neuroradiology, University Medical Center Mainz, Mainz 55131, Germany.
- Department of Radiology and Nuclear Medicine, University Medical Center Mannheim, Mannheim 68167, Germany.
Zitieren
APA
Kaatsch, H. L., Fulisch, F., Dillinger, D., Kubitscheck, L., Becker, B. V., Piechotka, J., Brockmann, M. A., Froelich, M. F., Schoenberg, S. O., Overhoff, D., & Waldeck, S. (2024). Ultra-low-dose photon-counting CT of paranasal sinus: an in vivo comparison of radiation dose and image quality to cone-beam CT. Dentomaxillofacial Radiology, 53(2), 103-108. https://doi.org/10.1093/dmfr/twad010
Vancouver
Kaatsch HL, Fulisch F, Dillinger D, Kubitscheck L, Becker BV, Piechotka J, et al. Ultra-low-dose photon-counting CT of paranasal sinus: an in vivo comparison of radiation dose and image quality to cone-beam CT. Dentomaxillofacial Radiology. 2024;53(2):103-108. doi:10.1093/dmfr/twad010
BibTeX
@article{Kaatsch2024,
author = {Kaatsch, Hanns Leonhard and Fulisch, Florian and Dillinger, Daniel and Kubitscheck, Laura and Becker, Benjamin V and Piechotka, Joel and Brockmann, Marc A and Froelich, Matthias F and Schoenberg, Stefan O and Overhoff, Daniel and Waldeck, Stephan},
title = {Ultra-low-dose photon-counting CT of paranasal sinus: an in vivo comparison of radiation dose and image quality to cone-beam CT},
journal = {Dentomaxillofacial Radiology},
year = {2024},
volume = {53},
number = {2},
pages = {103--108},
doi = {10.1093/dmfr/twad010}
}