Fachartikel in der Zeitschrift Pediatric Radiology, 2023. Verfasst von Touko Kaasalainen und 6 weiteren Autor:innen. Beteiligte Einrichtung: HUS Diagnostic Center, Radiology, University of Helsinki and Helsinki University Hospital, P.O. Box 340, 00290, Helsinki, Finland. touko.kaasalainen@…. Hauptthemen laut MEDLINE: Surgery, Computer-Assisted, Craniosynostoses. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 35 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: University of Helsinki including Helsinki University Central Hospital.
Abstract (Originalsprache)
Background
When postoperative multi-slice computed tomography (MSCT) imaging of patients with craniosynostosis is used, it is usually performed a few days after surgery in a radiology department. This requires additional anesthesia for the patient. Recently, intraoperative mobile cone-beam CT (CBCT) devices have gained popularity for orthopedic and neurosurgical procedures, which allows postoperative CT imaging in the operating room.
Objective
This single-center retrospective study compared radiation dose and image quality of postoperative imaging performed using conventional MSCT scanners and O-arm CBCT. Materials and methods A total of 104 pediatric syndromic and non-syndromic patients who were operated on because of single- or multiple-suture craniosynostosis were included in this study. The mean volumetric CT dose index (CTDI vol ) and dose-length product (DLP) values of optimized craniosynostosis CT examinations (58 MSCT and 46 CBCT) were compared. Two surgeons evaluated the subjective image quality.
Results
CBCT resulted in significantly lower CTDI vol (up to 14%) and DLP (up to 33%) compared to MSCT. Multi-slice CT image quality was considered superior to CBCT scans. However, all scans were considered to be of sufficient quality for diagnosis.
Conclusion
The O-arm device allowed for an immediate postoperative CBCT examination in the operating theater using the same anesthesia induction. Radiation exposure was lower in CBCT compared to MSCT scans, thus further encouraging the use of O-arms. Cone-beam CT imaging with an O-arm is a feasible method for postoperative craniosynostosis imaging, yielding less anesthesia to patients, lower health costs and the possibility to immediately evaluate results of the surgical operation.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- HUS Diagnostic Center, Radiology, University of Helsinki and Helsinki University Hospital, P.O. Box 340, 00290, Helsinki, Finland. touko.kaasalainen@hus.fi.
- Department of Oral and Maxillofacial Surgery, Lahti Central Hospital, Päijät-Häme Joint Authority for Health and Wellbeing, Lahti, Finland.
- HUS Diagnostic Center, Radiology, University of Helsinki and Helsinki University Hospital, P.O. Box 340, 00290, Helsinki, Finland.
- Department of Physics, University of Helsinki, Helsinki, Finland.
- Cleft Palate and Craniofacial Center, Department of Plastic Surgery, Helsinki University Hospital and University of Helsinki, Helsinki, Finland.
- Faculty of Medicine, University of Helsinki, Helsinki, Finland.
Förderung
- University of Helsinki including Helsinki University Central Hospital
Zitieren
APA
Kaasalainen, T., Männistö, V., Mäkelä, T., Suojanen, J., Nuorala, A., Heliövaara, A., & Leikola, J. (2023). Postoperative computed tomography imaging of pediatric patients with craniosynostosis: radiation dose and image quality comparison between multi-slice computed tomography and O-arm cone-beam computed tomography. Pediatric Radiology. https://doi.org/10.1007/s00247-023-05644-3
Vancouver
Kaasalainen T, Männistö V, Mäkelä T, Suojanen J, Nuorala A, Heliövaara A, et al. Postoperative computed tomography imaging of pediatric patients with craniosynostosis: radiation dose and image quality comparison between multi-slice computed tomography and O-arm cone-beam computed tomography. Pediatric Radiology. 2023. doi:10.1007/s00247-023-05644-3
BibTeX
@article{Kaasalainen2023,
author = {Kaasalainen, Touko and Männistö, Ville and Mäkelä, Teemu and Suojanen, Juho and Nuorala, Antton and Heliövaara, Arja and Leikola, Junnu},
title = {Postoperative computed tomography imaging of pediatric patients with craniosynostosis: radiation dose and image quality comparison between multi-slice computed tomography and O-arm cone-beam computed tomography},
journal = {Pediatric Radiology},
year = {2023},
doi = {10.1007/s00247-023-05644-3}
}
EUVIMED