Vergleichsstudie in der Zeitschrift Clinical Chemistry, online erschienen am 07.01.2026. Verfasst von Salomé Coppens und 6 weiteren Autor:innen. Beteiligte Einrichtung: PPC-LBPC, Univ Montpellier, INM, INSERM, IRMB, CHU Montpellier, Montpellier, France. Hauptthemen laut MEDLINE: Neurofilament Proteins, Liquid Chromatography-Mass Spectrometry. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: European Union's Horizon 2020, Marie Curie ITN "MIRIADE", NEuroBioStand.
Abstract (Originalsprache)
Background
Neurofilament light chain (Nf-L) is a key early biomarker for axonal damage and neurodegeneration, increasingly used in clinical practice for diagnosis, prognosis, and treatment monitoring. To ensure reliable clinical implementation, standardized measurement procedures and calibrators traceable to the International System of Units (SI) are needed. Although a few mass-spectrometry methods for Nf-L quantification in cerebrospinal fluid CSF) and plasma have been described, none currently use properly defined SI-traceable calibrators and existing harmonization efforts rely solely on immunoassays. This study presents a validated immunoprecipitation (IP)-LC-MS/MS assay using an SI-traceable calibrator and compares it with Lumipulse and Simoa platforms to assess agreement and bias.
Methods
A new IP-LC-MS/MS method was developed based on SI-traceable calibrator quantification and analytically validated using International Council for Harmonisation (ICH) guidelines. Analysis of 69 CSF samples by MS assay and Lumipulse (Fujirebio®) was performed as well as head-to-head comparisons between MS, Simoa (Quanterix®) and Lumipulse assays on 12 CSF pools.
Results
The method relying on 3 peptides was validated analytically. Significant results (P < 0.05) were obtained between amyloid positive to negative group when using the LC-MS/MS assay. MS and Lumipulse results were correlated. Head-to-head comparison of the 3 methods showed great correlation (r2 > 0.98) but systematic bias was identified between all techniques.
Conclusion
A new IP-LC-MS/MS method using a SI-traceable calibrator was developed, and analytically and clinically validated. Comparison between available immunoassays resulted in great correlation but biases were identified reinforcing the need of standardization for Nf-L measurement.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Substanzen
Autor:innen und Institutionen
- PPC-LBPC, Univ Montpellier, INM, INSERM, IRMB, CHU Montpellier, Montpellier, France.
- National Measurement Laboratory, LGC, Guildford, United Kingdom.
ORCID-Profile: Salomé Coppens, Sylvain Lehmann
Förderung
- European Union's Horizon 2020
- Marie Curie ITN "MIRIADE"
- NEuroBioStand
- European Partnership on Metrology
- European Union's Horizon Europe Research and Innovation Programme
- European Union’s Horizon 2020
- Marie Curie ITN “MIRIADE”
- European Union’s Horizon Europe Research and Innovation Programme
Zitieren
APA
Coppens, S., Vialaret, J., Mondésert, E., Drinkwater, G., Luckau, L., Lehmann, S., & Hirtz, C. (2026). Development of a Novel Liquid Chromatography Coupled to Multiple Reaction Monitoring (LC-MRM) Assay for the Quantification of Neurofilament Light Chain in Cerebrospinal Fluid and Comparison with Ultra-Sensitive Immunoassay: A Step toward Standardization. Clinical Chemistry, 72(4), 478-487. https://doi.org/10.1093/clinchem/hvaf180
Vancouver
Coppens S, Vialaret J, Mondésert E, Drinkwater G, Luckau L, Lehmann S, et al. Development of a Novel Liquid Chromatography Coupled to Multiple Reaction Monitoring (LC-MRM) Assay for the Quantification of Neurofilament Light Chain in Cerebrospinal Fluid and Comparison with Ultra-Sensitive Immunoassay: A Step toward Standardization. Clinical Chemistry. 2026;72(4):478-487. doi:10.1093/clinchem/hvaf180
BibTeX
@article{Coppens2026,
author = {Coppens, Salomé and Vialaret, Jérôme and Mondésert, Etienne and Drinkwater, Giles and Luckau, Luise and Lehmann, Sylvain and Hirtz, Christophe},
title = {Development of a Novel Liquid Chromatography Coupled to Multiple Reaction Monitoring (LC-MRM) Assay for the Quantification of Neurofilament Light Chain in Cerebrospinal Fluid and Comparison with Ultra-Sensitive Immunoassay: A Step toward Standardization},
journal = {Clinical Chemistry},
year = {2026},
volume = {72},
number = {4},
pages = {478--487},
doi = {10.1093/clinchem/hvaf180}
}
EUVIMED