Comparison of whole-blood sirolimus concentrations measured by EMIT-based Siemens Viva-ProE® System and LC-MS/MS in Chinese transplant patients

Rongqi Lin, Yipeng Cai, Yingbin Huang, Xueyong Li, Yiying Chen, Bo Chen, Kaixian Lai, Junnan Wu, Yu Cheng, Maobai Liu, Yisheng Chen, Hongqiang Qiu

Journal of Pharmaceutical and Biomedical Analysis · 2024

DOI: 10.1016/j.jpba.2024.116271 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Journal of Pharmaceutical and Biomedical Analysis, 2024. Verfasst von Rongqi Lin und 11 weiteren Autor:innen. Beteiligte Einrichtung: Department of Pharmacy, Shanghang County Hospital, Shanghang 364200, People's Republic of China; Department of Pharmacy, Fujian Medical University Un…. Hauptthemen laut MEDLINE: Drug Monitoring, Immunosuppressive Agents, Organ Transplantation, Sirolimus. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 32 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Sirolimus (SRL) is commonly used in transplant patients to prevent organ transplant rejection. The current guidelines recommend to perform SRL therapeutic drug monitoring regularly to improve treatment outcomes and avoid adverse effects. Consequently, a precise and accurate method for determining SRL is crucial in clinical practice. Currently, liquid chromatography-tandem mass spectrometry (LC-MS/MS) and immunoassays have been widely adopted for determining SRL concentrations. However, previous studies have shown that immunoassays exhibit a positive bias compared to LC-MS/MS. As the new updated version of the EMIT-based Viva-E® System (SVPS), this study aims to compare SRL blood concentrations measured by the SVPS and LC-MS/MS. The residual whole-blood samples obtained from transplant patients were simultaneously analyzed using the SVPS and LC-MS/MS, respectively. The correlation between the two assays was analyzed using the linear regression analysis and Deming linear regression. The Pearson correlation coefficient and Intraclass correlation coefficient (ICC) analysis were executed. The Paired Wilcoxon test and Bland-Altman analysis were performed to assess the concordance between the two methods. The SVPS considerably increased SRL concentration value by 46.62 % as compared to the LC-MS/MS method. When SRL concentrations measured by the SVPS were above 4.0 ng/mL, there was no significant difference between the corrected SVPS concentrations after using the Deming linear regression equation, indicating their interchangeability. Given the significant disparities observed between EMIT and LC-MS/MS, it is crucial to indicate the methodology and instruments in both TDM reports and future clinical guidelines. Our study also provides the conversion formulas between the SVPS and LC-MS/MS, which can be applied as a reference for different clinical centers.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Drug MonitoringImmunosuppressive AgentsOrgan TransplantationSirolimusTandem Mass SpectrometryAdultFemaleHumansMaleMiddle AgedChinaChromatography, LiquidEast Asian PeopleImmunoassayLiquid Chromatography-Mass SpectrometryReproducibility of Results

Stichwörter der Autor:innen

EMITImmunosuppressantsLC-MS/MSSirolimusTransplant patientsViva-ProE

Substanzen

Immunosuppressive AgentsSirolimus

Autor:innen und Institutionen

  • Department of Pharmacy, Shanghang County Hospital, Shanghang 364200, People's Republic of China; Department of Pharmacy, Fujian Medical University Union Hospital, Fuzhou 350001, People's Republic of China; The School of…
  • Department of Pharmacy, Fujian Medical University Union Hospital, Fuzhou 350001, People's Republic of China.
  • Department of Pharmacy, Fujian Medical University Union Hospital, Fuzhou 350001, People's Republic of China; The School of Pharmacy, Fujian Medical University, Fuzhou 350004, People's Republic of China.
  • Department of Pharmacy, Pinghe County Hospital, Pinghe 363700, People's Republic of China.
  • Department of Pharmacy, Shanghang County Hospital, Shanghang 364200, People's Republic of China. Electronic address: fjshys@fjmu.edu.cn.
  • Department of Pharmacy, Fujian Medical University Union Hospital, Fuzhou 350001, People's Republic of China; The School of Pharmacy, Fujian Medical University, Fuzhou 350004, People's Republic of China. Electronic addre…

Zitieren

APA

Lin, R., Cai, Y., Huang, Y., Li, X., Chen, Y., Chen, B., Lai, K., Wu, J., Cheng, Y., Liu, M., Chen, Y., & Qiu, H. (2024). Comparison of whole-blood sirolimus concentrations measured by EMIT-based Siemens Viva-ProE® System and LC-MS/MS in Chinese transplant patients. Journal of Pharmaceutical and Biomedical Analysis. https://doi.org/10.1016/j.jpba.2024.116271

Vancouver

Lin R, Cai Y, Huang Y, Li X, Chen Y, Chen B, et al. Comparison of whole-blood sirolimus concentrations measured by EMIT-based Siemens Viva-ProE® System and LC-MS/MS in Chinese transplant patients. Journal of Pharmaceutical and Biomedical Analysis. 2024. doi:10.1016/j.jpba.2024.116271

BibTeX
@article{Lin2024,
  author = {Lin, Rongqi and Cai, Yipeng and Huang, Yingbin and Li, Xueyong and Chen, Yiying and Chen, Bo and Lai, Kaixian and Wu, Junnan and Cheng, Yu and Liu, Maobai and Chen, Yisheng and Qiu, Hongqiang},
  title = {Comparison of whole-blood sirolimus concentrations measured by EMIT-based Siemens Viva-ProE® System and LC-MS/MS in Chinese transplant patients},
  journal = {Journal of Pharmaceutical and Biomedical Analysis},
  year = {2024},
  doi = {10.1016/j.jpba.2024.116271}
}
Kostenlos starten →