Development of an LC-MRM-MS-Based Candidate Reference Measurement Procedure for Standardization of Serum Apolipoprotein (a) Tests
Clinical Chemistry · 01.03.2023 · 69 (3) · 251-261
DOI: 10.1093/clinchem/hvac204 ↗Kurz gefasst
Fachartikel in der Zeitschrift Clinical Chemistry, online erschienen am 16.01.2023. Verfasst von L Renee Ruhaak und 18 weiteren Autor:innen. Beteiligte Einrichtung: Department of Clinical Chemistry and Laboratory Medicine, Leiden University Medical Center, Leiden, The Netherlands. Hauptthemen laut MEDLINE: Serum, Peptides. Die Arbeit wurde bisher 55-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: NIDDK NIH HHS.
Abstract (Originalsprache)
Background
Medical results generated by European CE Marking for In Vitro Diagnostic or in-house tests should be traceable to higher order reference measurement systems (RMS), such as International Federation of Clinical Chemistry and Laboratory Medicine (IFCC)-endorsed reference measurement procedures (RMPs) and reference materials. Currently, serum apolipoprotein (a) [apo(a)] is recognized as a novel risk factor for cardiovascular risk assessment and patient management. The former RMS for serum apo(a) is no longer available; consequently, an International System of Units (SI)-traceable, ideally multiplexed, and sustainable RMS for apo(a) is needed.
Methods
A mass spectrometry (MS)-based candidate RMP (cRMP) for apo(a) was developed using quantitative bottom-up proteomics targeting 3 proteotypic peptides. The method was provisionally validated according to ISO 15193 using a single human serum based calibrator traceable to the former WHO-IFCC RMS.
Results
The quantitation of serum apo(a) was by design independent of its size polymorphism, was linear from 3.8 to 456 nmol/L, and had a lower limit of quantitation for apo(a) of 3.8 nmol/L using peptide LFLEPTQADIALLK. Interpeptide agreement showed Pearson Rs of 0.987 and 0.984 for peptides GISSTVTGR and TPENYPNAGLTR, and method comparison indicated good correspondence (slopes 0.977, 1.033, and 1.085 for LFLEPTQADIALLK, GISSTVTGR, and TPENYPNAGLTR). Average within-laboratory imprecision of the cRMP was 8.9%, 11.9%, and 12.8% for the 3 peptides.
Conclusions
A robust, antibody-independent, MS-based cRMP was developed as higher order RMP and an essential part of the apo(a) traceability chain and future RMS. The cRMP fulfils predefined analytical performance specifications, making it a promising RMP candidate in an SI-traceable MS-based RMS for apo(a).
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Substanzen
Autor:innen und Institutionen
- Department of Clinical Chemistry and Laboratory Medicine, Leiden University Medical Center, Leiden, The Netherlands.
- Institute of Laboratory Medicine, Clinical Chemistry and Molecular Diagnostics, University Hospital Leipzig, Leipzig, Germany.
- Division of Laboratory Sciences, Centers for Disease Control and Prevention, Atlanta, GA, USA.
- Division Clinical Mass Spectrometry of the German Society of Clinical Chemistry and Laboratory Medicine (DGKL), Berlin, Germany.
- Department of Laboratory Medicine and Pathology, University of Washington, Seattle, WA, USA.
- Siemens Healthcare Diagnostics Products GmbH, Marburg, Germany.
ORCID-Profile: L Renee Ruhaak, Uta Ceglarek, Andrew N Hoofnagle, Stefan Coassin, Liesbet Deprez, Ioannis Dikaios, Gerhard M Kostner, Florian Kronenberg, Alicia Lyle, Christa M Cobbaert
Förderung
- NIDDK NIH HHS
Zitieren
APA
Ruhaak, L. R., Romijn, F. P. H. T. M., Begcevic Brkovic, I., Kuklenyik, Z., Dittrich, J., Ceglarek, U., Hoofnagle, A. N., Althaus, H., Angles-Cano, E., Coassin, S., Delatour, V., Deprez, L., Dikaios, I., Kostner, G. M., Kronenberg, F., Lyle, A., Prinzing, U., Vesper, H. W., & Cobbaert, C. M. (2023). Development of an LC-MRM-MS-Based Candidate Reference Measurement Procedure for Standardization of Serum Apolipoprotein (a) Tests. Clinical Chemistry, 69(3), 251-261. https://doi.org/10.1093/clinchem/hvac204
Vancouver
Ruhaak LR, Romijn FPHTM, Begcevic Brkovic I, Kuklenyik Z, Dittrich J, Ceglarek U, et al. Development of an LC-MRM-MS-Based Candidate Reference Measurement Procedure for Standardization of Serum Apolipoprotein (a) Tests. Clinical Chemistry. 2023;69(3):251-261. doi:10.1093/clinchem/hvac204
BibTeX
@article{Ruhaak2023,
author = {Ruhaak, L Renee and Romijn, Fred P H T M and Begcevic Brkovic, Ilijana and Kuklenyik, Zsusanna and Dittrich, Julia and Ceglarek, Uta and Hoofnagle, Andrew N and Althaus, Harald and Angles-Cano, Eduardo and Coassin, Stefan and Delatour, Vincent and Deprez, Liesbet and Dikaios, Ioannis and Kostner, Gerhard M and Kronenberg, Florian and Lyle, Alicia and Prinzing, Urban and Vesper, Hubert W and Cobbaert, Christa M},
title = {Development of an LC-MRM-MS-Based Candidate Reference Measurement Procedure for Standardization of Serum Apolipoprotein (a) Tests},
journal = {Clinical Chemistry},
year = {2023},
volume = {69},
number = {3},
pages = {251--261},
doi = {10.1093/clinchem/hvac204}
}
EUVIMED