Validation of metrological traceability of the new generation of Abbott Alinity alkaline phosphatase assay
Clinical Chemistry and Laboratory Medicine (CCLM) · 2024
DOI: 10.1515/cclm-2023-0553 ↗Kurz gefasst
Fachartikel in der Zeitschrift Clinical Chemistry and Laboratory Medicine (CCLM), 2024. Verfasst von Giorgia Bianchi und 5 weiteren Autor:innen. Beteiligte Einrichtung: Clinical Pathology Unit, ASST Fatebenefratelli-Sacco, Milan, Italy. Hauptthemen laut MEDLINE: Alkaline Phosphatase, Clinical Enzyme Tests. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 36 Literaturangaben.
Abstract (Originalsprache)
Objectives
Recently, Abbott Diagnostics marketed a new generation of Alinity enzyme assays, introducing a multiparametric calibrator [Consolidated Chemistry Calibrator (ConCC)] in place of or in addition to factor-based calibrations. For alkaline phosphatase (ALP), both calibration options are offered, i.e., with ConCC (ALP2) and with an experimental calibration factor (ALP2F). Both options are declared traceable to the 2011 IFCC reference measurement procedure (RMP). Before to replace the old generation (ALP1) with the new one, we decided to validate the trueness of ALP2/ALP2F.
Methods
Three approaches were employed: (a) preliminary comparison on 48 native frozen serum samples with ALP1, of which traceability to RMP was previously successfully verified; (b) examination of three banked serum pools (BSP) with values assigned by RMP; (c) direct comparison with RMP on a set of 24 fresh serum samples. Bias estimation and regression studies were performed, and the standard measurement uncertainty associated with ALP measurements on clinical samples (u result ) was estimated and compared with established analytical performance specifications (APS). ConCC commutability was also assessed.
Results
A positive proportional bias was found with both ALP2 and ALP2F when compared to ALP1 and RMP. This positive bias was confirmed on BSP: in average, +13.1 % for ALP2 and +10.0 % for ALP2F, respectively. u result were 13.28 % for ALP2 and 10.04 % for ALP2F, both not fulfilling the minimum APS of 4.0 %. Furthermore, ConCC was not commutable with clinical samples.
Conclusions
Our results unearth problems in the correct implementation of traceability of Alinity ALP2/ALP2F, with the risk for the new assay to be unfit for clinical purposes.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Clinical Pathology Unit, ASST Fatebenefratelli-Sacco, Milan, Italy.
- Research Centre for Metrological Traceability in Laboratory Medicine (CIRME), University of Milan, Milan, Italy.
ORCID-Profile: Giulia Colombo, Elena Aloisio
Zitieren
APA
Bianchi, G., Frusciante, E., Colombo, G., Infusino, I., Aloisio, E., & Panteghini, M. (2024). Validation of metrological traceability of the new generation of Abbott Alinity alkaline phosphatase assay. Clinical Chemistry and Laboratory Medicine (CCLM). https://doi.org/10.1515/cclm-2023-0553
Vancouver
Bianchi G, Frusciante E, Colombo G, Infusino I, Aloisio E, Panteghini M. Validation of metrological traceability of the new generation of Abbott Alinity alkaline phosphatase assay. Clinical Chemistry and Laboratory Medicine (CCLM). 2024. doi:10.1515/cclm-2023-0553
BibTeX
@article{Bianchi2024,
author = {Bianchi, Giorgia and Frusciante, Erika and Colombo, Giulia and Infusino, Ilenia and Aloisio, Elena and Panteghini, Mauro},
title = {Validation of metrological traceability of the new generation of Abbott Alinity alkaline phosphatase assay},
journal = {Clinical Chemistry and Laboratory Medicine (CCLM)},
year = {2024},
doi = {10.1515/cclm-2023-0553}
}
EUVIMED