Validation of metrological traceability of the new generation of Abbott Alinity alkaline phosphatase assay

Giorgia Bianchi, Erika Frusciante, Giulia Colombo iD, Ilenia Infusino, Elena Aloisio iD, Mauro Panteghini

Clinical Chemistry and Laboratory Medicine (CCLM) · 2024

DOI: 10.1515/cclm-2023-0553 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Clinical Chemistry and Laboratory Medicine (CCLM), 2024. Verfasst von Giorgia Bianchi und 5 weiteren Autor:innen. Beteiligte Einrichtung: Clinical Pathology Unit, ASST Fatebenefratelli-Sacco, Milan, Italy. Hauptthemen laut MEDLINE: Alkaline Phosphatase, Clinical Enzyme Tests. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 36 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

Recently, Abbott Diagnostics marketed a new generation of Alinity enzyme assays, introducing a multiparametric calibrator [Consolidated Chemistry Calibrator (ConCC)] in place of or in addition to factor-based calibrations. For alkaline phosphatase (ALP), both calibration options are offered, i.e., with ConCC (ALP2) and with an experimental calibration factor (ALP2F). Both options are declared traceable to the 2011 IFCC reference measurement procedure (RMP). Before to replace the old generation (ALP1) with the new one, we decided to validate the trueness of ALP2/ALP2F.

Methods

Three approaches were employed: (a) preliminary comparison on 48 native frozen serum samples with ALP1, of which traceability to RMP was previously successfully verified; (b) examination of three banked serum pools (BSP) with values assigned by RMP; (c) direct comparison with RMP on a set of 24 fresh serum samples. Bias estimation and regression studies were performed, and the standard measurement uncertainty associated with ALP measurements on clinical samples (u result ) was estimated and compared with established analytical performance specifications (APS). ConCC commutability was also assessed.

Results

A positive proportional bias was found with both ALP2 and ALP2F when compared to ALP1 and RMP. This positive bias was confirmed on BSP: in average, +13.1 % for ALP2 and +10.0 % for ALP2F, respectively. u result were 13.28 % for ALP2 and 10.04 % for ALP2F, both not fulfilling the minimum APS of 4.0 %. Furthermore, ConCC was not commutable with clinical samples.

Conclusions

Our results unearth problems in the correct implementation of traceability of Alinity ALP2/ALP2F, with the risk for the new assay to be unfit for clinical purposes.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Alkaline PhosphataseClinical Enzyme TestsHumansSerumCalibrationReference Standards

Stichwörter der Autor:innen

alkaline phosphatasemeasurement uncertaintymetrological traceabilitystandardizationtrueness

Substanzen

Alkaline Phosphatase

Autor:innen und Institutionen

  • Clinical Pathology Unit, ASST Fatebenefratelli-Sacco, Milan, Italy.
  • Research Centre for Metrological Traceability in Laboratory Medicine (CIRME), University of Milan, Milan, Italy.

ORCID-Profile: Giulia Colombo, Elena Aloisio

Zitieren

APA

Bianchi, G., Frusciante, E., Colombo, G., Infusino, I., Aloisio, E., & Panteghini, M. (2024). Validation of metrological traceability of the new generation of Abbott Alinity alkaline phosphatase assay. Clinical Chemistry and Laboratory Medicine (CCLM). https://doi.org/10.1515/cclm-2023-0553

Vancouver

Bianchi G, Frusciante E, Colombo G, Infusino I, Aloisio E, Panteghini M. Validation of metrological traceability of the new generation of Abbott Alinity alkaline phosphatase assay. Clinical Chemistry and Laboratory Medicine (CCLM). 2024. doi:10.1515/cclm-2023-0553

BibTeX
@article{Bianchi2024,
  author = {Bianchi, Giorgia and Frusciante, Erika and Colombo, Giulia and Infusino, Ilenia and Aloisio, Elena and Panteghini, Mauro},
  title = {Validation of metrological traceability of the new generation of Abbott Alinity alkaline phosphatase assay},
  journal = {Clinical Chemistry and Laboratory Medicine (CCLM)},
  year = {2024},
  doi = {10.1515/cclm-2023-0553}
}
Kostenlos starten →