Commutability study on three CRMs evaluating their suitability for calibration, trueness verification and statistical quality control of methods measuring metal concentrations in human blood

Thomas P J Linsinger, Guy Auclair, Liesbet Deprez

Analytical and Bioanalytical Chemistry · 2025

DOI: 10.1007/s00216-025-05751-0 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Analytical and Bioanalytical Chemistry, 2025. Verfasst von Thomas P J Linsinger und 2 weiteren Autor:innen. Beteiligte Einrichtung: Joint Research Centre, European Commission, Retieseweg 111, 2440, Geel, Belgium. thomas.linsinger@ec.europa.eu. Hauptthemen laut MEDLINE: Metals, Mass Spectrometry, Spectrophotometry, Atomic. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: HORIZON EUROPE Non-nuclear direct actions of the Joint Research Centre.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The commutability of three certified reference materials for Cd, Cr, Hg, Ni, Pb and Tl in blood was evaluated for the combination of digestion ICP-MS with dilution ICP-MS or dilution GFAAS using the difference in bias approach. Maximum non-commutability bias (MANCB) was derived from the median of the tolerance limits set by six proficiency testing (PT) providers. The evaluation showed that this criteria are too strict and that either unfeasible high number of measurements or a larger MANCB is required to demonstrate commutability. The effect of increasing the MANCB on the maximum tolerable measurement uncertainty was found small to moderate demonstrating the commutability of the materials when used as a calibrator. Also the inclusion of the uncertainty of non-commutability into the measurement uncertainty resulted in only small increases of the measurement uncertainty, demonstrating the suitability of the materials for trueness control and method validation. The bias of the three CRMs was for most elements within the 99% confidence limits of the bias of the patient samples, demonstrating the suitability of the materials for statistical quality control. Graphical abstract

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

MetalsMass SpectrometrySpectrophotometry, AtomicHumansCalibrationQuality ControlReference StandardsUncertaintyReproducibility of Results

Stichwörter der Autor:innen

Certified reference materialCommutabilityMetrology

Substanzen

Metals

Autor:innen und Institutionen

  • Joint Research Centre, European Commission, Retieseweg 111, 2440, Geel, Belgium. thomas.linsinger@ec.europa.eu.
  • Joint Research Centre, European Commission, Retieseweg 111, 2440, Geel, Belgium.

Förderung

  • HORIZON EUROPE Non-nuclear direct actions of the Joint Research Centre

Zitieren

APA

Linsinger, T. P. J., Auclair, G., & Deprez, L. (2025). Commutability study on three CRMs evaluating their suitability for calibration, trueness verification and statistical quality control of methods measuring metal concentrations in human blood. Analytical and Bioanalytical Chemistry. https://doi.org/10.1007/s00216-025-05751-0

Vancouver

Linsinger TPJ, Auclair G, Deprez L. Commutability study on three CRMs evaluating their suitability for calibration, trueness verification and statistical quality control of methods measuring metal concentrations in human blood. Analytical and Bioanalytical Chemistry. 2025. doi:10.1007/s00216-025-05751-0

BibTeX
@article{Linsinger2025,
  author = {Linsinger, Thomas P J and Auclair, Guy and Deprez, Liesbet},
  title = {Commutability study on three CRMs evaluating their suitability for calibration, trueness verification and statistical quality control of methods measuring metal concentrations in human blood},
  journal = {Analytical and Bioanalytical Chemistry},
  year = {2025},
  doi = {10.1007/s00216-025-05751-0}
}
Kostenlos starten →