Validation and predictive modeling of a portable bladder scanner for precision bladder management in pelvic radiotherapy
British Journal of Radiology · 2026
DOI: 10.1093/bjr/tqag138 ↗Kurz gefasst
Validierungsstudie in der Zeitschrift British Journal of Radiology, 2026. Verfasst von Wenzheng Huang und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Tongren Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai 200336, China. Hauptthemen laut MEDLINE: Urinary Bladder, Radiotherapy Planning, Computer-Assisted, Urinary Bladder Neoplasms. Die Arbeit nennt 42 Literaturangaben.
Abstract (Originalsprache)
Objectives
Validate the accuracy of a portable bladder scanner (Caresono PadScan HD5) against planning CT and demonstrate its clinical application to optimize bladder preparation in pelvic radiotherapy.
Methods
Two retrospective cohorts were analyzed: 114 patients assessed with HD5 and 111 historical controls. Agreement between HD5-measured volume (VHD5) and planning CT (VCT) was evaluated with correlation, regression, Bland–Altman, and paired tests. Initial CBCT repetition rates were compared between cohorts. Subgroup analyses (bladder volume [BV]<300 vs ≥300 mL; age<65 vs ≥ 65 years); sex and multivariable linear regression with interaction terms explored predictors of measurement error.
Results
V HD5 correlated with VCT (r = 0.723, P < .001) but underestimated volumes (mean bias −34.39 mL). Underestimation was markedly greater at VCT ≥300 mL (−71.46 vs −18.64 mL, P < .001). Multivariable regression confirmed BV as the primary driver, with an age-dependent interaction showing steeper underestimation in younger (<65 years) patients (P < .001). Following HD5 implementation, initial CBCT repetition rates decreased from 48.7% to 22.8% (P < .001).
Conclusions
HD5 enables reliable, though negatively biased, BV estimates. Embedding an age-stratified, moderate-fill workflow reduces repeat initial CBCT and standardizes bladder preparation. Advances in knowledge We quantify volume-dependent underestimation and an age interaction, and translate them into a predictive, age-stratified protocol that operationalizes HD5 measurements for routine pelvic RT bladder management.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology, Tongren Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai 200336, China.
- Department of Radiation Oncology, Shanghai Cancer Center Minhang Branch, Fudan University, 106 Ruili Road, Minhang District, Shanghai, 200240, China.
ORCID-Profile: Wenzheng Huang, Qunchao Hu
Zitieren
APA
Huang, W., Fu, W., Zhang, X., Sun, H., He, X., Li, F., & Hu, Q. (2026). Validation and predictive modeling of a portable bladder scanner for precision bladder management in pelvic radiotherapy. British Journal of Radiology. https://doi.org/10.1093/bjr/tqag138
Vancouver
Huang W, Fu W, Zhang X, Sun H, He X, Li F, et al. Validation and predictive modeling of a portable bladder scanner for precision bladder management in pelvic radiotherapy. British Journal of Radiology. 2026. doi:10.1093/bjr/tqag138
BibTeX
@article{Huang2026,
author = {Huang, Wenzheng and Fu, Weiping and Zhang, Xinlei and Sun, Hangbiao and He, Xuwei and Li, Feng and Hu, Qunchao},
title = {Validation and predictive modeling of a portable bladder scanner for precision bladder management in pelvic radiotherapy},
journal = {British Journal of Radiology},
year = {2026},
doi = {10.1093/bjr/tqag138}
}
EUVIMED