Comparison of Bladder Volume Measurement Accuracy between Portable Ultrasound Diagnostic Equipment and CBCT for Pelvic Prostate Radiotherapy

Takahiro Saika, Kenji Matsumoto

Japanese Journal of Radiological Technology · 2024

DOI: 10.6009/jjrt.2024-1461 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Japanese Journal of Radiological Technology, 2024. Verfasst von Takahiro Saika und 1 weiteren Autor:innen. Beteiligte Einrichtung: Radiology Center, Kindai University Hospital. Hauptthemen laut MEDLINE: Prostatic Neoplasms, Urinary Bladder, Cone-Beam Computed Tomography, Ultrasonography. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 10 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE: We evaluated the measurement accuracy of the CubeScan BioCon-900 (here after BioCon-900), a portable ultrasound imaging diagnostic device capable of measuring bladder volume, to determine if it can accurately assess bladder volume before intensity-modulated radiation therapy (IMRT) for prostate cancer. METHODS: Bladder volume was measured from kV-Cone Beam computed tomography (CBCT) images obtained immediately before the administration of IMRT for prostate cancer using Halcyon. The bladder volume measured from kV-CBCT images (23 patients, total number of scans: 139) immediately before the IMRT procedure was used as the reference value. The difference between the bladder volume measured by the BioCon-900 was then calculated. RESULTS: The bladder volume measured from kV-CBCT images was 117.5±49.4 cc. In contrast, the bladder volume obtained with BioCon-900 was 104.1±48.6 ml, resulting in an absolute mean difference of 18.4% and a correlation coefficient of 0.881. The measurements by BioCon-900 tended to be, on average, 11% smaller than the bladder volume measured by kV-CBCT imaging. CONCLUSION: kV-CBCT images demonstrated a strong positive correlation between bladder volume and bladder urine output obtained with BioCon-900. BioCon-900 can be used before kV-CBCT imaging to accurately and conveniently assess bladder volume.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Prostatic NeoplasmsUrinary BladderCone-Beam Computed TomographyUltrasonographyHumansMaleAgedMiddle AgedRadiotherapy, Intensity-ModulatedOrgan SizeAged, 80 and over

Stichwörter der Autor:innen

CUBESCANbladder volumeprostateradiotherapyultrasound

Autor:innen und Institutionen

  • Radiology Center, Kindai University Hospital.

Zitieren

APA

Saika, T., & Matsumoto, K. (2024). Comparison of Bladder Volume Measurement Accuracy between Portable Ultrasound Diagnostic Equipment and CBCT for Pelvic Prostate Radiotherapy. Japanese Journal of Radiological Technology. https://doi.org/10.6009/jjrt.2024-1461

Vancouver

Saika T, Matsumoto K. Comparison of Bladder Volume Measurement Accuracy between Portable Ultrasound Diagnostic Equipment and CBCT for Pelvic Prostate Radiotherapy. Japanese Journal of Radiological Technology. 2024. doi:10.6009/jjrt.2024-1461

BibTeX
@article{Saika2024,
  author = {Saika, Takahiro and Matsumoto, Kenji},
  title = {Comparison of Bladder Volume Measurement Accuracy between Portable Ultrasound Diagnostic Equipment and CBCT for Pelvic Prostate Radiotherapy},
  journal = {Japanese Journal of Radiological Technology},
  year = {2024},
  doi = {10.6009/jjrt.2024-1461}
}
Kostenlos starten →