Cardiac devices audit evaluating the use of 6 MV FFF beams following updated Radiotherapy Board guidance

Emily Russell iD, David Paynter, Peter Bownes, Vivian Cosgrove

British Journal of Radiology · 01.08.2026 · 99 (1184) · 1538-1540

DOI: 10.1093/bjr/tqag097 ↗

Kurz gefasst

Fachartikel in der Zeitschrift British Journal of Radiology, online erschienen am 29.04.2026. Verfasst von Emily Russell und 3 weiteren Autor:innen. Beteiligte Einrichtung: Leeds Cancer Centre, Leeds Teaching Hospitals NHS Trust, St James Hospital, Leeds LS9 7TF, United Kingdom. Hauptthemen laut MEDLINE: Pacemaker, Artificial, Radiotherapy, Intensity-Modulated, Defibrillators, Implantable. Die Arbeit nennt 6 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

Following guidance from the Radiotherapy Board for radiotherapy of patients with cardiac implantable electronic devices (CIED), a local audit of retrospective patient treatments was carried out. The guidance includes recommendations to avoid direct irradiation of the devices and the avoidance of Flattening Filter Free (FFF) beams due to the high dose rates associated with FFF beams, leading to higher risk of device malfunction.

Methods

Linac dose-rates were measured by noting the dose rate of the linac whilst delivering 6 MV FFF treatment plans, the local standard energy used for Volumetric Modulated Arc Therapy (VMAT). Plans were also assessed to determine whether treatment beams traveled through the CIED.

Results

It was found that 1 out of 540 patients with CIEDs was treated with the CIED receiving a dose rate above 8 Gy/minute for a 6 MV FFF beam. The plan for this treatment did not have the “protect function” for CIEDs, which is now standard planning practice.

Conclusions

The dose rate when using 6 MV FFF beams is sufficiently reduced, and therefore we have confidence in our current practice for the protection of patients with CIEDs. Advances in knowledge This technical note responds to recent guidance published on the safety of radiotherapy exposures for patients with cardiac devices. It audits the standard practices in a large NHS Trust, assessing whether these standard practices align with the guidance. This will be of interest to many other trusts aspiring to adhere to this guidance.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Pacemaker, ArtificialRadiotherapy, Intensity-ModulatedDefibrillators, ImplantableHumansRadiotherapy DosageRetrospective StudiesRadiotherapy Planning, Computer-AssistedRadiation ProtectionMedical AuditParticle Accelerators

Stichwörter der Autor:innen

CIEDRCR guidancecardiac deviceexternal beamflattening filter freeradiotherapyradiotherapy physics

Autor:innen und Institutionen

  • Leeds Cancer Centre, Leeds Teaching Hospitals NHS Trust, St James Hospital, Leeds LS9 7TF, United Kingdom.

ORCID-Profile: Emily Russell

Zitieren

APA

Russell, E., Paynter, D., Bownes, P., & Cosgrove, V. (2026). Cardiac devices audit evaluating the use of 6 MV FFF beams following updated Radiotherapy Board guidance. British Journal of Radiology, 99(1184), 1538-1540. https://doi.org/10.1093/bjr/tqag097

Vancouver

Russell E, Paynter D, Bownes P, Cosgrove V. Cardiac devices audit evaluating the use of 6 MV FFF beams following updated Radiotherapy Board guidance. British Journal of Radiology. 2026;99(1184):1538-1540. doi:10.1093/bjr/tqag097

BibTeX
@article{Russell2026,
  author = {Russell, Emily and Paynter, David and Bownes, Peter and Cosgrove, Vivian},
  title = {Cardiac devices audit evaluating the use of 6 MV FFF beams following updated Radiotherapy Board guidance},
  journal = {British Journal of Radiology},
  year = {2026},
  volume = {99},
  number = {1184},
  pages = {1538--1540},
  doi = {10.1093/bjr/tqag097}
}
Kostenlos starten →