Evaluation of testicular shielding effectiveness for surface and internal photon dose using a 3D‐printed scrotal phantom

Joseph Speth, Josh Kilian-Meneghin, Salma K Jabbour, Yin Zhang, Xiao Wang iD, Zhenyu Xiong iD, Ke Nie, Ning J Yue, Xinxin Zhang iD

Journal of Applied Clinical Medical Physics · 2026

DOI: 10.1002/acm2.70842 ↗

Kurz gefasst

Evaluationsstudie in der Zeitschrift Journal of Applied Clinical Medical Physics, 2026. Verfasst von Joseph Speth und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Rutgers Cancer Institute of New Jersey, Rutgers Robert Wood Johnson Medical School, New Brunswick, New Jersey, USA. Hauptthemen laut MEDLINE: Phantoms, Imaging, Radiation Protection, Radiotherapy Planning, Computer-Assisted, Testis. Die Arbeit nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Testicular shielding has long been used to reduce testicular dose during radiation treatment. However, quantitative evaluations of the effectiveness of such shielding in the literature vary substantially.

Purpose

In this study, the dose reduction expected at both the testicular skin surface and the inside of the testes was evaluated under different irradiation conditions, to enable informed clinical decision‐making.

Methods

A 3D printed scrotal phantom was designed to simulate patient testes and was composed of three layers to enable dose measurement inside the phantom. Gafchromic EBT3 film was used for dose measurements. The scrotal phantom was attached to a RANDO anthropomorphic phantom for computed tomography scanning and measurements. A commercial testicular shield with a lead wall thickness of 1.27 cm and a diameter of 6 cm was used for shielded measurements. The measured absolute dose was normalized to the delivered MU for each measurement to obtain a dose‐to‐MU value (cGy/MU). Both dose‐to‐MU values and shielding effectiveness were evaluated with a stationary anterior‐posterior beam at different distances from field edge to scrotal phantom. In addition, the energy dependence of shielding effectiveness was assessed using two volumetric modulated arc therapy (VMAT) plans with identical parameters and modulation but using 6 and 15 MV energies.

Results

The shielded and unshielded dose‐to‐MU values generally decreased with increasing field‐to‐phantom distance for both surface and internal dose measurements. However, the shielding ratio remained relatively constant across distances, with a reduction factor of approximately 3.3 for surface dose and 2.7 for internal dose. Minimal energy dependence was observed in both internal and surface dose measurements, as well as in shielding effectiveness, across the two VMAT plans. The overall shielding ratio was approximately 2.3 for both plans.

Conclusions

Testicular shielding reduced both surface and internal testicular dose by approximately a factor of 2 to 3, demonstrating consistent effectiveness across different irradiation geometries and photon energies.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Phantoms, ImagingRadiation ProtectionRadiotherapy Planning, Computer-AssistedTestisPrinting, Three-DimensionalPhotonsOrgans at RiskScrotumMaleHumansRadiotherapy DosageRadiotherapy, Intensity-Modulated

Stichwörter der Autor:innen

out‐of‐field doseradiation therpaytesticular shielding

Autor:innen und Institutionen

  • Department of Radiation Oncology, Rutgers Cancer Institute of New Jersey, Rutgers Robert Wood Johnson Medical School, New Brunswick, New Jersey, USA.
  • Department of Radiation Oncology, Robert Wood Johnson Barnabas Health, Livingston, New Jersey, USA.

ORCID-Profile: Xiao Wang, Zhenyu Xiong, Xinxin Zhang

Zitieren

APA

Speth, J., Kilian-Meneghin, J., Jabbour, S. K., Zhang, Y., Wang, X., Xiong, Z., Nie, K., Yue, N. J., & Zhang, X. (2026). Evaluation of testicular shielding effectiveness for surface and internal photon dose using a 3D‐printed scrotal phantom. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70842

Vancouver

Speth J, Kilian-Meneghin J, Jabbour SK, Zhang Y, Wang X, Xiong Z, et al. Evaluation of testicular shielding effectiveness for surface and internal photon dose using a 3D‐printed scrotal phantom. Journal of Applied Clinical Medical Physics. 2026. doi:10.1002/acm2.70842

BibTeX
@article{Speth2026,
  author = {Speth, Joseph and Kilian-Meneghin, Josh and Jabbour, Salma K and Zhang, Yin and Wang, Xiao and Xiong, Zhenyu and Nie, Ke and Yue, Ning J and Zhang, Xinxin},
  title = {Evaluation of testicular shielding effectiveness for surface and internal photon dose using a 3D‐printed scrotal phantom},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2026},
  doi = {10.1002/acm2.70842}
}
Kostenlos starten →