Film measurements of kilovoltage x-ray percentage depth doses for different energies and field sizes

Tze Yee Lim iD, Dershan Luo, Jamshid Moradi-Kurdestany, Xin Wang, Ramesh Tailor

British Journal of Radiology · 01.08.2025 · 98 (1172) · 1298-1304

DOI: 10.1093/bjr/tqaf107 ↗

Kurz gefasst

Fachartikel in der Zeitschrift British Journal of Radiology, online erschienen am 27.05.2025. Verfasst von Tze Yee Lim und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, TX 77030, United States. Hauptthemen laut MEDLINE: Film Dosimetry, Radiation Dosage, X-Ray Therapy. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 36 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

Kilovoltage x-ray therapy, which includes superficial x-rays and orthovoltage x-rays, is particularly suited for treatments near the skin surface. Radiation dose deposition at depth is typically characterized by percentage depth dose (PDD) curves. However, PDD data are difficult to obtain in kilovoltage x-ray therapy, hence clinics often rely on British Journal of Radiology Supplement 25 (BJR-25) archival data. This study measured PDD for different energies and field sizes on the Xstrahl 300 kilovoltage x-ray therapy unit for comparison with BJR-25 data.

Methods

We irradiated EBT3 film in water using different energy and field size combinations, then, compared the acquired PDD curves with BJR-25 data. We used 75, 125, and 250 kVp beams, for 10 × 10 cm2 open field, 4 × 4 cm2 closed-ended square applicator, and open-ended circular applicators of 1.5, 2.0, 2.8, and 4.7 cm diameter.

Results

PDD for different energies and field sizes on the Xstrahl 300 kilovoltage x-ray therapy unit agreed well with BJR-25 data. The average difference between measured and BJR-25 data was −0.3 ± 1.8%; 85.9% of the individual local differences were within ±3%, 94.8% within ±4%, and 99.5% within ±5%.

Conclusions

The PDD curves presented here, together with BJR-25 data, can serve as useful comparison datasets for the commissioning of kilovoltage x-ray machines. Advances in knowledge Our film-based measurement technique provides high spatial resolution PDD datapoints for a modern kilovoltage x-ray unit, adding to the limited published data in the kilovoltage energy range.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Film DosimetryRadiation DosageX-Ray TherapyHumansRadiotherapy DosageX-RaysPhantoms, Imaging

Stichwörter der Autor:innen

kilovoltage x-rayorthovoltagepercentage depth dosesuperficial

Autor:innen und Institutionen

  • Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, TX 77030, United States.
  • Department of Radiation Oncology, Mercy Coletta Cancer Center, Oklahoma City, OK 73120, United States.

ORCID-Profile: Tze Yee Lim

Zitieren

APA

Lim, T. Y., Luo, D., Moradi-Kurdestany, J., Wang, X., & Tailor, R. (2025). Film measurements of kilovoltage x-ray percentage depth doses for different energies and field sizes. British Journal of Radiology, 98(1172), 1298-1304. https://doi.org/10.1093/bjr/tqaf107

Vancouver

Lim TY, Luo D, Moradi-Kurdestany J, Wang X, Tailor R. Film measurements of kilovoltage x-ray percentage depth doses for different energies and field sizes. British Journal of Radiology. 2025;98(1172):1298-1304. doi:10.1093/bjr/tqaf107

BibTeX
@article{Lim2025,
  author = {Lim, Tze Yee and Luo, Dershan and Moradi-Kurdestany, Jamshid and Wang, Xin and Tailor, Ramesh},
  title = {Film measurements of kilovoltage x-ray percentage depth doses for different energies and field sizes},
  journal = {British Journal of Radiology},
  year = {2025},
  volume = {98},
  number = {1172},
  pages = {1298--1304},
  doi = {10.1093/bjr/tqaf107}
}
Kostenlos starten →