Characterization of a new low‐dose and low‐energy Gafchromic film LD‐V1

Oliva Masella, Kevin J Murphy, Magdalena Bazalova-Carter iD

Journal of Applied Clinical Medical Physics · 2024

DOI: 10.1002/acm2.14531 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Oliva Masella und 2 weiteren Autor:innen. Beteiligte Einrichtung: Physics and Astronomy Department, University of Victoria, Victoria, British Columbia, Canada. Hauptthemen laut MEDLINE: Film Dosimetry, Radiation Dosage. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 35 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NSERC Discovery Grant and Discovery Accelerator Supplement, Canada Research Chairs program, Canada Foundation for Innovation.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To characterize the dose‐response, energy dependence, postexposure changes, orientation dependence, and spatial capabilities of LD‐V1, a new low‐dose Gafchromic film for low‐energy x‐ray dosimetry.

Methods

A single sheet of LD‐V1 Gafchromic film was cut into 15 × 20 mm 2 rectangles with a notch to track orientation. Eight different doses between 5 and 320 mGy were delivered by an MXR‐160/22 x‐ray tube using x‐ray beams of 90, 100, and 120 kVp filtered with 3 mm of Al and 2 mm of Ti. The 120 kVp films were scanned at 1, 1.5, 2, 3, 12, 24, 48, 72, and 168 h postexposure in portrait orientation and additionally scanned in landscape orientation at 24 h. The 90 and 100 kVp films were scanned at 24 h postexposure in portrait orientation. Lastly, a 20 × 200 mm 2 strip of film was irradiated using a thin‐slit imaging collimator and scanned 24 h postexposure to test the film performance in an x‐ray imaging application.

Results

Of the three color channels, the red channel was found to produce a dose‐response curve with a large range of net optical density (netOD) values across the considered dose range. A prominent energy dependence was discovered, resulting in dose discrepancies on the scale of 17 mGy between 90 and 120 kVp for a dose of 80 mGy. The measured postexposure changes suggest that the calibration irradiation‐to‐scan time should be longer than 12 h with a ± 4 h scanning time window for dose errors of

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Film DosimetryRadiation DosageHumansX-RaysCalibrationPhantoms, Imaging

Stichwörter der Autor:innen

dosimetryfilmlow energy x‐rays

Autor:innen und Institutionen

  • Physics and Astronomy Department, University of Victoria, Victoria, British Columbia, Canada.

ORCID-Profile: Magdalena Bazalova-Carter

Förderung

  • NSERC Discovery Grant and Discovery Accelerator Supplement
  • Canada Research Chairs program
  • Canada Foundation for Innovation
  • British Columbia Knowledge Fund

Zitieren

APA

Masella, O., Murphy, K. J., & Bazalova-Carter, M. (2024). Characterization of a new low‐dose and low‐energy Gafchromic film LD‐V1. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14531

Vancouver

Masella O, Murphy KJ, Bazalova-Carter M. Characterization of a new low‐dose and low‐energy Gafchromic film LD‐V1. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14531

BibTeX
@article{Masella2024,
  author = {Masella, Oliva and Murphy, Kevin J and Bazalova-Carter, Magdalena},
  title = {Characterization of a new low‐dose and low‐energy Gafchromic film LD‐V1},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2024},
  doi = {10.1002/acm2.14531}
}
Kostenlos starten →