Design and development of automated beta irradiation and calibration system: effect of beam flattening filters on dose homogeneity using radiochromic films and angular beta dose dependency on ISO phantoms by measurement and MC simulation

İbrahim Demirel, Haluk Yücel iD

Radiation Protection Dosimetry · 20.07.2026 · 202 (12) · 915-930

DOI: 10.1093/rpd/ncag031 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Radiation Protection Dosimetry, online erschienen am 05.04.2026. Verfasst von İbrahim Demirel und 1 weiteren Autor:innen. Beteiligte Einrichtung: Ankara University Institute of Nuclear Sciences, Besevler 10. Yıl campus, 06100 Tandogan, Ankara, Türkiye. Hauptthemen laut MEDLINE: Film Dosimetry, Beta Particles, Phantoms, Imaging, Radiation Protection. Die Arbeit nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: Scientific and Technological Research Council of Turkey.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The present study is aimed at investigating dose homogeneity in a reference beta radiation field established for calibration of protection level devices and beta irradiation purposes as well as angular dose responses of the detectors on ISO rod and pillar phantoms. To this end, an automatic electromechanic system for beta irradiation and calibration (BIKS) was designed and manufactured with precise automatic control of the source shutter, stepper motors, detector rotation, and distance using microcontrollers. BIKS system enables to measure angular beta personal dose equivalents for skin-extremity Hp(0.07) and the lens of the eye Hp (3), including the corrections to standard reference conditions by using a multi-sensor. In beta radiation field, 2D spatial dose distributions were measured at 11 cm using radiochromic film dosimetry. Dose homogeneity is obtained better than ~91 ± 5%, on the average. Field uniformity was expanded using filters such as Hostaphan® PET films recommended by ISO 6980-1:2023 and other custom-designed filters. Angular dose responses of the detectors were measured over 0° to ±60° angle on ISO rod and pillar phantoms at 11, 20, and 30 cm. The measured differences in Hp(0.07) are <3% for all angles. BIKS system was also modelled using MC–based RayXpert® software. The estimated results are good agreement with those experimental ones.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Film DosimetryBeta ParticlesPhantoms, ImagingRadiation ProtectionCalibrationMonte Carlo MethodEquipment DesignHumansRadiation DosageComputer SimulationAutomation

Autor:innen und Institutionen

  • Ankara University Institute of Nuclear Sciences, Besevler 10. Yıl campus, 06100 Tandogan, Ankara, Türkiye.

ORCID-Profile: Haluk Yücel

Förderung

  • Scientific and Technological Research Council of Turkey

Zitieren

APA

Demirel, İ., & Yücel, H. (2026). Design and development of automated beta irradiation and calibration system: effect of beam flattening filters on dose homogeneity using radiochromic films and angular beta dose dependency on ISO phantoms by measurement and MC simulation. Radiation Protection Dosimetry, 202(12), 915-930. https://doi.org/10.1093/rpd/ncag031

Vancouver

Demirel İ, Yücel H. Design and development of automated beta irradiation and calibration system: effect of beam flattening filters on dose homogeneity using radiochromic films and angular beta dose dependency on ISO phantoms by measurement and MC simulation. Radiation Protection Dosimetry. 2026;202(12):915-930. doi:10.1093/rpd/ncag031

BibTeX
@article{Demirel2026,
  author = {Demirel, İbrahim and Yücel, Haluk},
  title = {Design and development of automated beta irradiation and calibration system: effect of beam flattening filters on dose homogeneity using radiochromic films and angular beta dose dependency on ISO phantoms by measurement and MC simulation},
  journal = {Radiation Protection Dosimetry},
  year = {2026},
  volume = {202},
  number = {12},
  pages = {915--930},
  doi = {10.1093/rpd/ncag031}
}
Kostenlos starten →