Design and development of automated beta irradiation and calibration system: effect of beam flattening filters on dose homogeneity using radiochromic films and angular beta dose dependency on ISO phantoms by measurement and MC simulation
Radiation Protection Dosimetry · 20.07.2026 · 202 (12) · 915-930
DOI: 10.1093/rpd/ncag031 ↗Kurz gefasst
Fachartikel in der Zeitschrift Radiation Protection Dosimetry, online erschienen am 05.04.2026. Verfasst von İbrahim Demirel und 1 weiteren Autor:innen. Beteiligte Einrichtung: Ankara University Institute of Nuclear Sciences, Besevler 10. Yıl campus, 06100 Tandogan, Ankara, Türkiye. Hauptthemen laut MEDLINE: Film Dosimetry, Beta Particles, Phantoms, Imaging, Radiation Protection. Die Arbeit nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: Scientific and Technological Research Council of Turkey.
Abstract (Originalsprache)
The present study is aimed at investigating dose homogeneity in a reference beta radiation field established for calibration of protection level devices and beta irradiation purposes as well as angular dose responses of the detectors on ISO rod and pillar phantoms. To this end, an automatic electromechanic system for beta irradiation and calibration (BIKS) was designed and manufactured with precise automatic control of the source shutter, stepper motors, detector rotation, and distance using microcontrollers. BIKS system enables to measure angular beta personal dose equivalents for skin-extremity Hp(0.07) and the lens of the eye Hp (3), including the corrections to standard reference conditions by using a multi-sensor. In beta radiation field, 2D spatial dose distributions were measured at 11 cm using radiochromic film dosimetry. Dose homogeneity is obtained better than ~91 ± 5%, on the average. Field uniformity was expanded using filters such as Hostaphan® PET films recommended by ISO 6980-1:2023 and other custom-designed filters. Angular dose responses of the detectors were measured over 0° to ±60° angle on ISO rod and pillar phantoms at 11, 20, and 30 cm. The measured differences in Hp(0.07) are <3% for all angles. BIKS system was also modelled using MC–based RayXpert® software. The estimated results are good agreement with those experimental ones.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Ankara University Institute of Nuclear Sciences, Besevler 10. Yıl campus, 06100 Tandogan, Ankara, Türkiye.
ORCID-Profile: Haluk Yücel
Förderung
- Scientific and Technological Research Council of Turkey
Zitieren
APA
Demirel, İ., & Yücel, H. (2026). Design and development of automated beta irradiation and calibration system: effect of beam flattening filters on dose homogeneity using radiochromic films and angular beta dose dependency on ISO phantoms by measurement and MC simulation. Radiation Protection Dosimetry, 202(12), 915-930. https://doi.org/10.1093/rpd/ncag031
Vancouver
Demirel İ, Yücel H. Design and development of automated beta irradiation and calibration system: effect of beam flattening filters on dose homogeneity using radiochromic films and angular beta dose dependency on ISO phantoms by measurement and MC simulation. Radiation Protection Dosimetry. 2026;202(12):915-930. doi:10.1093/rpd/ncag031
BibTeX
@article{Demirel2026,
author = {Demirel, İbrahim and Yücel, Haluk},
title = {Design and development of automated beta irradiation and calibration system: effect of beam flattening filters on dose homogeneity using radiochromic films and angular beta dose dependency on ISO phantoms by measurement and MC simulation},
journal = {Radiation Protection Dosimetry},
year = {2026},
volume = {202},
number = {12},
pages = {915--930},
doi = {10.1093/rpd/ncag031}
}
EUVIMED