Evaluation of neutron dose equivalent in prostate stereotactic body radiotherapy by bubble detector

Sali Di iD, Brandon VanGenderen iD, Marcus Sonier, Joseph Awotwi-Pratt

Biomedical Physics & Engineering Express · 2026

DOI: 10.1088/2057-1976/aeaba9 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Biomedical Physics & Engineering Express, 2026. Verfasst von Sali Di und 3 weiteren Autor:innen. Beteiligte Einrichtung: BC Cancer Abbotsford, Abbotsford, British Columbia, Canada. Hauptthemen laut MEDLINE: Radiosurgery, Prostatic Neoplasms, Neutrons, Radiometry. Die Arbeit nennt 33 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Stereotactic body radiotherapy (SBRT) for prostate cancer often uses high-energy photon beams, which can produce photoneutrons through interaction in the linac head, raising concerns regarding neutron-related exposure. Therefore, it is important to estimate neutron dose equivalent in SBRT plans when comparing beam energies and treatment delivery techniques. BD-PND fast neutron detectors were used in this study to quantify neutron dose equivalent in prostate SBRT plans. We performed dosimetry measurements using a Varian TrueBeam Linac and assessed neutron production at energies of 2.5 MV, 6 MV, 6 MV flattening filter free (FFF), 10 MV, 10 MV FFF and 15 MV. Bubble counts were converted to neutron dose equivalent using the manufacturer-provided calibration factor. The measured neutron dose equivalent was highest at the centre and decreased at peripheral measurement locations farther from the primary beam. Significant differences in neutron dose equivalent were found between flattening filter (FF) and FFF beams at the central beam position after multiple comparison correction. In contrast, peripheral positions showed no statistically significant differences. Additionally, low bubble counts were observed under 6 MV and 6 MV FFF beams, suggesting a low-level detector response under the present measurement conditions. Overall, beam energy, FF or FFF delivery and measurement positions all affect the measured neutron dose equivalent in prostate SBRT. The low-energy observation should be interpreted as preliminary, and additional work is needed, including repeated measurements, different detector types and more clinically relevant measurement locations.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

RadiosurgeryProstatic NeoplasmsNeutronsRadiometryHumansMaleRadiotherapy DosageRadiotherapy Planning, Computer-AssistedParticle AcceleratorsPhantoms, ImagingCalibrationPhotons

Stichwörter der Autor:innen

Stereotactic body radiotherapybubble detectorsflattening filter-freeneutron dosesprostate cancer

Autor:innen und Institutionen

  • BC Cancer Abbotsford, Abbotsford, British Columbia, Canada.

ORCID-Profile: Sali Di, Brandon VanGenderen

Zitieren

APA

Di, S., VanGenderen, B., Sonier, M., & Awotwi-Pratt, J. (2026). Evaluation of neutron dose equivalent in prostate stereotactic body radiotherapy by bubble detector. Biomedical Physics & Engineering Express. https://doi.org/10.1088/2057-1976/aeaba9

Vancouver

Di S, VanGenderen B, Sonier M, Awotwi-Pratt J. Evaluation of neutron dose equivalent in prostate stereotactic body radiotherapy by bubble detector. Biomedical Physics & Engineering Express. 2026. doi:10.1088/2057-1976/aeaba9

BibTeX
@article{Di2026,
  author = {Di, Sali and VanGenderen, Brandon and Sonier, Marcus and Awotwi-Pratt, Joseph},
  title = {Evaluation of neutron dose equivalent in prostate stereotactic body radiotherapy by bubble detector},
  journal = {Biomedical Physics & Engineering Express},
  year = {2026},
  doi = {10.1088/2057-1976/aeaba9}
}
Kostenlos starten →