A Taguchi-optimized agar phantom for temperature-based validation in electro-hyperthermia
Biomedical Physics & Engineering Express · 01.06.2026 · 12 (3) · 035010
DOI: 10.1088/2057-1976/ae5ca7 ↗Kurz gefasst
Validierungsstudie in der Zeitschrift Biomedical Physics & Engineering Express, online erschienen am 24.04.2026. Verfasst von Chih-Wu Cheng und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Ditmanson Medical Foundation Chiayi Christian Hospital, Chia-Yi City, Taiwan. Hauptthemen laut MEDLINE: Agar, Hyperthermia, Induced, Phantoms, Imaging. Die Arbeit nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Research Program of Ditmanson Medical Foundation Chia-Yi Christian Hospital.
Abstract (Originalsprache)
Objective
This study aimed to develop and optimize an agar-based phantom using the Taguchi method for temperature-based validation of electro-hyperthermia systems in a quality assurance (QA) framework. Materials and methods . This study utilized the Taguchi method with an orthogonal array to design nine agar phantom formulations with varying concentrations of three factors: agar, sodium chloride (NaCl), and sodium azide (NaN₃). Heating was performed using the Oncotherm EHY-2000 RF hyperthermia device under six different power levels (25–100 W), with each stage lasting 5 min for a total duration of 30 min. Temperature changes were measured at a depth of 5 cm within the phantom using a type T thermocouple thermometer. A pork tissue model was used as the reference standard for comparison.
Results
All phantom formulations exhibited a linear increase in temperature during the heating process. Analysis using Minitab software identified the optimal formulation, consisting of 5.0% agar, 0.1% NaCl, and 0.44% NaN₃, as producing a temperature profile most closely resembling that of the pork tissue model. Taguchi analysis indicated that agar concentration was the most significant factor influencing temperature variation. The interactions among the three formulation variables were weak, suggesting that each factor could be optimized independently. Significance. The agar phantom developed in this study features simple fabrication, reusability, and long-term storability, making it a practical tool for detecting abnormal heating in hyperthermia devices and enhancing treatment safety and accuracy. It is suited for routine QA of the Oncotherm electro-hyperthermia system.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Radiation Oncology, Ditmanson Medical Foundation Chiayi Christian Hospital, Chia-Yi City, Taiwan.
- Department of Biomedical Engineering, Chung Yuan Christian University, Taoyuan City, Taiwan.
ORCID-Profile: Chih-Wu Cheng, Wen-Tyng Li
Förderung
- Research Program of Ditmanson Medical Foundation Chia-Yi Christian Hospital
Zitieren
APA
Cheng, C. W., Li, W. T., & Tsang, Y. W. (2026). A Taguchi-optimized agar phantom for temperature-based validation in electro-hyperthermia. Biomedical Physics & Engineering Express, 12(3), 035010. https://doi.org/10.1088/2057-1976/ae5ca7
Vancouver
Cheng CW, Li WT, Tsang YW. A Taguchi-optimized agar phantom for temperature-based validation in electro-hyperthermia. Biomedical Physics & Engineering Express. 2026;12(3):035010. doi:10.1088/2057-1976/ae5ca7
BibTeX
@article{Cheng2026,
author = {Cheng, Chih-Wu and Li, Wen-Tyng and Tsang, Yuk-Wah},
title = {A Taguchi-optimized agar phantom for temperature-based validation in electro-hyperthermia},
journal = {Biomedical Physics & Engineering Express},
year = {2026},
volume = {12},
number = {3},
pages = {035010},
doi = {10.1088/2057-1976/ae5ca7}
}
EUVIMED