Fachartikel in der Zeitschrift Medical Physics, online erschienen am 13.08.2025. Verfasst von Anastasia Antoniou und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Electrical Engineering, Computer Engineering, and Informatics, Cyprus University of Technology, Limassol, Cyprus. Hauptthemen laut MEDLINE: Agar, Phantoms, Imaging, Magnetic Resonance Imaging, High-Intensity Focused Ultrasound Ablation. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 50 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Recovery and Resilience Facility, BRAINSONIC, Research and Innovation Foundation of Cyprus.
Abstract (Originalsprache)
Background
While the technical aspects of histotripsy have seen significant advancements, research on Magnetic Resonance Imaging (MRI) monitoring remains limited.
Purpose
This preliminary study explored the use of conventional T2‐Weighted (T2‐W) Turbo Spin Echo (TSE) imaging to monitor histotripsy lesions in a pure agar gel, as a potential research tool for MRI‐guided histotripsy (MRgHt).
Methods
Histotripsy experiments were conducted in a 3T MRI scanner using a 2% weight per volume agar solution in water. Pulsed Focused Ultrasound (FUS) was applied at a duty factor of 2%, with a pulse repetition period of 1s, delivering up to 2000 pulses. T2‐W TSE imaging was employed for intra‐procedural monitoring with 10s updates and post‐sonication assessment, with the first prioritizing rapid acquisition and the second focusing on resolution. Supplementary, MR thermometry tracked temperature variations throughout the procedure.
Results
Dynamic T2‐W TSE imaging provided insights into the progress of phantom disruption. Mild thermal effects served as early signs of phantom response, as mechanical effects required more time to manifest. Typical histotripsy lesion characteristics were observed, including a hyperintense core with an elongated shape and a thin hypointense border. The developed temperatures remained well below hyperthermia levels, reaching a maximum of 34°C.
Conclusions
The presented methodology, utilizing a pure agar gel within a simple, versatile setup alongside conventional T2‐W TSE imaging, holds promise for advancing MRgHt research.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Electrical Engineering, Computer Engineering, and Informatics, Cyprus University of Technology, Limassol, Cyprus.
- Diagnostic and Interventional Radiology Department, German Medical Institute, Limassol, Cyprus.
- Department of Electronics and Information Engineering, Hangzhou Dianzi University, Hangzhou, Zhejiang, China.
- 3DMI Research Group, Department of Medical Physics, School of Medicine, University of Patras, Rion-Patras, Greece.
Förderung
- Recovery and Resilience Facility
- BRAINSONIC
- Research and Innovation Foundation of Cyprus
Zitieren
APA
Antoniou, A., Evripidou, N., Georgiou, L., Christofi, A., Zhao, J., Yu, L., Li, W., Kagadis, G. C., & Damianou, C. (2025). Magnetic Resonance Imaging monitoring of histotripsy effects in agar phantom. Medical Physics, 52(8). https://doi.org/10.1002/mp.18054
Vancouver
Antoniou A, Evripidou N, Georgiou L, Christofi A, Zhao J, Yu L, et al. Magnetic Resonance Imaging monitoring of histotripsy effects in agar phantom. Medical Physics. 2025;52(8). doi:10.1002/mp.18054
BibTeX
@article{Antoniou2025,
author = {Antoniou, Anastasia and Evripidou, Nikolas and Georgiou, Leonidas and Christofi, Antonis and Zhao, Jufeng and Yu, Liyang and Li, Wenjun and Kagadis, George C and Damianou, Christakis},
title = {Magnetic Resonance Imaging monitoring of histotripsy effects in agar phantom},
journal = {Medical Physics},
year = {2025},
volume = {52},
number = {8},
doi = {10.1002/mp.18054}
}
EUVIMED