Consistency of Faraday cup and ionization chamber dosimetry of proton fields and the role of nuclear interactions

Jörg Wulff, Anne Paul, Claus Maximilian Bäcker, Kilian-Simon Baumann, Johannes Niklas Esser, Benjamin Koska, Beate Timmermann, Nico Gerd Verbeek, Christian Bäumer

Medical Physics · 2024

DOI: 10.1002/mp.16819 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, 2024. Verfasst von Jörg Wulff und 8 weiteren Autor:innen. Beteiligte Einrichtung: West German Proton Therapy Centre Essen, Essen, Germany. Hauptthemen laut MEDLINE: Protons, Proton Therapy. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 73 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: EU Horizon 2020.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

A Faraday cup (FC) facilitates a quite clean measurement of the proton fluence emerging from clinical spot‐scanning nozzles with narrow pencil‐beams. The utilization of FCs appears to be an attractive option for high dose rate delivery modes and the source models of Monte‐Carlo (MC) dose engines. However, previous studies revealed discrepancies of 3%–6% between reference dosimetry with ionization chambers (ICs) and FC‐based dosimetry. This has prevented the widespread use of FCs for dosimetry in proton therapy.

Purpose

The current study aims at bridging the gap between FC dosimetry and IC dosimetry of proton fields delivered with spot‐scanning treatment heads. Particularly, a novel method to evaluate FC measurements is introduced.

Methods

A consistency check is formulated, which makes use of the energy balance and the reciprocity theorem. The measurement data comprise central‐axis depth distributions of the absorbed dose of quasi‐monochromatic fields with a width of about 28.5 cm and FC measurements of the reciprocal fields with a single spot. These data are complemented by a look‐up of energy‐range tables, the average Q‐value of transmutations, and the escape energy carried away by neutrons and photons. The latter data are computed by MC simulations, which in turn are validated with measurements of the distal dose tail and neutron out‐of‐field doses. For comparison, the conventional approach of FC evaluation is performed, which computes absorbed dose from the product of fluence and stopping power. The results from the FC measurements are compared with the standard dosimetry protocols and improved reference dosimetry methods.

Results

The deviation between the conventional FC‐based dosimetry and the IC‐based one according to standard dosimetry protocols was −4.7 ( 3.3)% for a 100 MeV field and −3.6 (3.5)% for 200 MeV, thereby agreeing within the reported uncertainties. The deviations could be reduced to −4.0 ( 2.9)% and −3.0 ( 3.1)% by adopting state‐of‐the‐art reference dosimetry methods. The alternative approach using the energy balance gave deviations of only −1.9% (100 MeV) and −2.6% (200 MeV) using state‐of‐the‐art dosimetry. The standard uncertainty of this novel approach was estimated to be about 2%.

Conclusions

An alternative concept has been established to determine the absorbed dose of monoenergetic proton fields with an FC. It eliminates the strong dependence of the conventional FC‐based approach on the MC simulation of the stopping‐power and of the secondary ions, which according to the study at hand is the major contributor to the underestimation of the absorbed dose. Some contributions to the uncertainty of the novel approach could potentially be reduced in future studies. This would allow for accurate consistency tests of conventional dosimetry procedures.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ProtonsProton TherapyRadiometryComputer SimulationCalibrationMonte Carlo MethodRadiotherapy Dosage

Stichwörter der Autor:innen

Faraday cupproton therapyreference dosimetry

Substanzen

Protons

Autor:innen und Institutionen

  • West German Proton Therapy Centre Essen, Essen, Germany.
  • University Hospital Essen, Essen, Germany.
  • West German Cancer Center (WTZ), Essen, Germany.
  • Heinrich Heine University Düsseldorf, Düsseldorf, Germany.
  • Department of Radiotherapy and Radiation Oncology, Marburg University Hospital, Marburg, Germany.
  • Marburg Ion-Beam Therapy Center (MIT), Marburg, Germany.

Förderung

  • EU Horizon 2020

Zitieren

APA

Wulff, J., Paul, A., Bäcker, C. M., Baumann, K. S., Esser, J. N., Koska, B., Timmermann, B., Verbeek, N. G., & Bäumer, C. (2024). Consistency of Faraday cup and ionization chamber dosimetry of proton fields and the role of nuclear interactions. Medical Physics. https://doi.org/10.1002/mp.16819

Vancouver

Wulff J, Paul A, Bäcker CM, Baumann KS, Esser JN, Koska B, et al. Consistency of Faraday cup and ionization chamber dosimetry of proton fields and the role of nuclear interactions. Medical Physics. 2024. doi:10.1002/mp.16819

BibTeX
@article{Wulff2024,
  author = {Wulff, Jörg and Paul, Anne and Bäcker, Claus Maximilian and Baumann, Kilian-Simon and Esser, Johannes Niklas and Koska, Benjamin and Timmermann, Beate and Verbeek, Nico Gerd and Bäumer, Christian},
  title = {Consistency of Faraday cup and ionization chamber dosimetry of proton fields and the role of nuclear interactions},
  journal = {Medical Physics},
  year = {2024},
  doi = {10.1002/mp.16819}
}
Kostenlos starten →