Water calorimetry-based beam quality correction factors for carbon and helium ion beams
Physics in Medicine & Biology · 28.07.2026 · 71 (14) · 145006
DOI: 10.1088/1361-6560/ae794e ↗Kurz gefasst
Fachartikel in der Zeitschrift Physics in Medicine & Biology, online erschienen am 16.07.2026. Verfasst von Lies Verpoest und 14 weiteren Autor:innen. Beteiligte Einrichtung: UCLouvain-Institut de Recherche Expérimentale et Clinique-Molecular Imaging Radiotherapy and Oncology (MIRO), Brussels, Belgium. Hauptthemen laut MEDLINE: Helium, Water, Carbon, Calorimetry. Die Arbeit nennt 44 Literaturangaben. Gefördert von: Waalse Gewest, HORIZON EUROPE Excellent Science.
Abstract (Originalsprache)
Objective. To determine beam quality correction factors, ℏ , in single-layer scanned carbon and helium ion beams using water calorimetry, thereby reducing uncertainties in dosimetry for light ion therapy. Approach. Water calorimetry measurements were performed under harmonized conditions at two synchrotron-based ion beam therapy centers (MedAustron, Austria and HIT, Germany). Measurements were conducted in the plateau region of single-layer scanned beams. Carbon ion beams covered nominal energies from 213.4–402.8 MeV/u, while helium ion beams were measured at 196.3 and 198.8 MeV/u. Multiple cylindrical (IBA FC65-G, PTW 30013) and plane-parallel (IBA PPC05, PPC40; PTW 34001, 34045) ionization chamber types were investigated. Results. ℏ -factors were determined with standard uncertainties of approximately 1%, depending on chamber type and measurement condition. For all investigated chambers, inter-center agreement was observed within the combined standard uncertainties. Chamber-to-chamber variability was below 0.5% for chamber types for which two samples were available, except for the IBA PPC05, for which differences up to 1.3% were observed. For carbon ion beams, the measured ℏ -factors are consistent with previously published water calorimetry-based results. However, systematic deviations up to 3% persist between experimentally determined and Monte Carlo-derived ℏ values. The measured ℏ -factors for helium ions show trends that are broadly similar to those observed in carbon ion beams across all chamber types. Significance. This work expands the experimental ℏ -database for light ions by providing new carbon ion data and the first experimental ℏ -factors for helium ion beams. These results provide essential experimental data for future refinement of dosimetry recommendations, and highlight persistent discrepancies between the experimental and Monte Carlo-based ℏ -factors that require further investigation.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- UCLouvain-Institut de Recherche Expérimentale et Clinique-Molecular Imaging Radiotherapy and Oncology (MIRO), Brussels, Belgium.
- Dosimetry Business Unit, Ion Beams Applications S.A., Louvain La Neuve, Belgium.
- Heidelberg Ion-Beam Therapy Center (HIT), Department of Radiation Oncology, Heidelberg University Hospital (UKHD), Heidelberg, Germany.
- Clinical Cooperation Unit Translational Radiation Oncology, German Cancer Research Center (DKFZ), Heidelberg, Germany.
- Division of Medical Physics in Radiation Oncology, German Cancer Research Center (DKFZ), Heidelberg, Germany.
- National Center for Radiation Research in Oncology (NCRO), Heidelberg Institute for Radiation Oncology (HIRO), Heidelberg, Germany.
ORCID-Profile: Lies Verpoest, Kevin Souris, Thomas Tessonnier, Iván Muñoz, Stephan Brons, Oliver Jäkel, Kilian-Simon Baumann, Hermann Fuchs, Peter Kuess, Hugo Palmans
Förderung
- Waalse Gewest
- HORIZON EUROPE Excellent Science
Zitieren
APA
Verpoest, L., Rossomme, S., Souris, K., Tessonnier, T., Mairani, A., Muñoz, I., Brons, S., Jäkel, O., Baumann, K. S., Fuchs, H., Kuess, P., Douralis, A., Lourenço, A., Palmans, H., & Lee, J. (2026). Water calorimetry-based beam quality correction factors for carbon and helium ion beams. Physics in Medicine & Biology, 71(14), 145006. https://doi.org/10.1088/1361-6560/ae794e
Vancouver
Verpoest L, Rossomme S, Souris K, Tessonnier T, Mairani A, Muñoz I, et al. Water calorimetry-based beam quality correction factors for carbon and helium ion beams. Physics in Medicine & Biology. 2026;71(14):145006. doi:10.1088/1361-6560/ae794e
BibTeX
@article{Verpoest2026,
author = {Verpoest, Lies and Rossomme, Séverine and Souris, Kevin and Tessonnier, Thomas and Mairani, Andrea and Muñoz, Iván and Brons, Stephan and Jäkel, Oliver and Baumann, Kilian-Simon and Fuchs, Hermann and Kuess, Peter and Douralis, Alexandros and Lourenço, Ana and Palmans, Hugo and Lee, John},
title = {Water calorimetry-based beam quality correction factors for carbon and helium ion beams},
journal = {Physics in Medicine & Biology},
year = {2026},
volume = {71},
number = {14},
pages = {145006},
doi = {10.1088/1361-6560/ae794e}
}
EUVIMED