Micro CT calibration accuracy for pre-clinical studies in ion therapy

A Bückner iD, M Moglioni iD, U Weber iD, L Volz iD, O Sokol iD, M Witt iD, M Simard, M Durante iD, D Boscolo iD

Physics in Medicine & Biology · 14.05.2026 · 71 (9) · 095004

DOI: 10.1088/1361-6560/ae5ebb ↗

Kurz gefasst

Fachartikel in der Zeitschrift Physics in Medicine & Biology, online erschienen am 29.04.2026. Verfasst von A Bückner und 8 weiteren Autor:innen. Beteiligte Einrichtung: GSI Helmholtz Centre for Heavy Ion Research GmbH, Darmstadt, Germany. Hauptthemen laut MEDLINE: X-Ray Microtomography, Heavy Ion Radiotherapy. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 62 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: ERC Advanced Grant.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective. Accurate estimation of tissue stopping power ratios (SPRs) from computed tomography (CT) data is essential in charged particle therapy (CPT), yet it remains a major source of range uncertainties. While various SPR calibration methods have been developed for clinical CT scanners, they are not tailored to cone beam type Micro CT scanners commonly used in pre-clinical studies involving small animals. This work aims to assess how different clinical SPR calibration methods influence range predictions in small animal CPT pre-clinical trials, evaluated at both the voxel and organ levels. Approach. A small animal calibration phantom was imaged with a Micro CT at x-ray tube voltages between 45 kV and 70 kV. The images were used in two single energy (SECT), a stoichiometric and a simplified method, and one stoichiometric dual energy (DECT) calibration method. The accuracy of SPR predictions for each method was evaluated using simulated CT images of murine head, thorax, and abdomen virtual phantoms derived from in vivo data. Main results. Both SECT methods were found to yield comparable accuracy in the head and abdomen areas. Root mean squared (RMS) errors to the nominal SPR values were found to be 5.04% and 4.22% in the head phantom and 2.31% and 2.61% in the abdomen phantom, for the stoichiometric and simplified SECT method, respectively. RMS errors in the lung phantom were found to be 16.42% and 35.80%. The stoichiometric DECT method was found to improve accuracy in the lung phantom to 12.37% while in the head phantom it achieved 19.37% and in the abdomen 6.42%. Significance. This is the first report of treatment plan agnostic low-kV Micro CT SPR calibration accuracy. It can provide a reference for expected SPR and range uncertainties in pre-clinical CPT trials.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

X-Ray MicrotomographyHeavy Ion RadiotherapyCalibrationAnimalsPhantoms, ImagingMiceHead

Stichwörter der Autor:innen

Micro CT calibrationdual-energy CTparticle therapyrange uncertaintiessingle-energy CTsmall animal imagingstopping-power calibration

Autor:innen und Institutionen

  • GSI Helmholtz Centre for Heavy Ion Research GmbH, Darmstadt, Germany.
  • Institute for Condensed Matter Physics, TU Darmstadt, Darmstadt, Germany.
  • Institute for Medical Physics and Radiation Protection, Technische Hochschule Mittelhessen, Gießen, Germany.
  • Marburg Ion-Beam Therapy Center, Marburg, Germany.
  • Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom.

ORCID-Profile: A Bückner, M Moglioni, U Weber, L Volz, O Sokol, M Witt, M Durante, D Boscolo

Förderung

  • ERC Advanced Grant

Zitieren

APA

Bückner, A., Moglioni, M., Weber, U., Volz, L., Sokol, O., Witt, M., Simard, M., Durante, M., & Boscolo, D. (2026). Micro CT calibration accuracy for pre-clinical studies in ion therapy. Physics in Medicine & Biology, 71(9), 095004. https://doi.org/10.1088/1361-6560/ae5ebb

Vancouver

Bückner A, Moglioni M, Weber U, Volz L, Sokol O, Witt M, et al. Micro CT calibration accuracy for pre-clinical studies in ion therapy. Physics in Medicine & Biology. 2026;71(9):095004. doi:10.1088/1361-6560/ae5ebb

BibTeX
@article{Bckner2026,
  author = {Bückner, A and Moglioni, M and Weber, U and Volz, L and Sokol, O and Witt, M and Simard, M and Durante, M and Boscolo, D},
  title = {Micro CT calibration accuracy for pre-clinical studies in ion therapy},
  journal = {Physics in Medicine & Biology},
  year = {2026},
  volume = {71},
  number = {9},
  pages = {095004},
  doi = {10.1088/1361-6560/ae5ebb}
}
Kostenlos starten →