Characterization of the Medscint HYPERSCINT scintillation dosimetry research platform with orthovoltage photon beams and a novel quasi monoenergetic beam

Leonard Che Fru iD, Benjamin A Insley iD, Dongyeon Lee iD, Dirk A Bartkoski iD, Reza Reiazi iD, Aviad Schori iD, Ramesh Tailor iD, Mohammad Salehpour iD

Physics in Medicine & Biology · 2025

DOI: 10.1088/1361-6560/ae0efb ↗

Kurz gefasst

Fachartikel in der Zeitschrift Physics in Medicine & Biology, 2025. Verfasst von Leonard Che Fru und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Physics, Unit 1420, The University of Texas MD Anderson Cancer Center, 1515 Holcombe Boulevard, Houston, TX 77030, United Sta…. Hauptthemen laut MEDLINE: Photons, Radiometry, Scintillation Counting. Die Arbeit nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Convergent Radiotherapy and Radiosurgery Ltd, Convergent Radiosurgery and Radiotherapy Ltd..

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective . This work aimed to evaluate the dosimetric characteristics of the new HyperScint TM Research Platform 200 system (HS-RP200), with orthovoltage beams and a novel kilovoltage converging beam, converging radiotherapy and radiosurgery (CRnR) beam. Approach . Several dosimetric properties of the HS-RP200 detector with several orthovoltage beams were studied in air. Relative dosimetry of the HS-RP200 detector was also studied for use on the converging x-ray beam. Energy correction factors were compared between measurement, Monte Carlo, and analytical methods. Main results. The signal-to-noise ratio for the detector increased as the frame exposure time increased, and repeated measurements had less uncertainty when measurements were performed without stem-effect removal (WOSER). The short-term stability study of the detector’s signal showed an acceptable variance of its readings to within 1% during 1 h measurements. The detector exhibited angular independence within the measurement uncertainty of 1%. The detector exhibited excellent dose linearity response with the orthovoltage beams. During an 80 d period, repeated measurements determined that the detector possesses long-term stability within a root mean square (rms) of 2.3% and 0.6% when measurements were performed with stem effect removal and WOSER, respectively. Calculated energy correction factors implied significant fluctuations of the HS-RP200 detector with the beam quality of orthovoltage beams, ranging from 3% to 66%, depending on the divergence of the effective energy of the beam from reference quality. Monte Carlo–simulated energy correction factors agreed with measurements with a maximum difference of 5.4% when measurements were performed WOSER. Lastly, the HS-RP200 detector system exhibited good relative dosimetric characteristics with the CRnR beam. A maximum rms value of 0.019 was calculated for the normalized profiles when measurements from the HS-RP200 detector were compared to measurements from film. Significance . The HS-RP200 detector could be used as an alternative to other detectors with orthovoltage beams, including the CRnR beam.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

PhotonsRadiometryScintillation CountingMonte Carlo MethodSignal-To-Noise RatioUncertaintyRadiosurgeryRadiotherapy Dosage

Stichwörter der Autor:innen

CRnRMedscint HS-RP200converging beamdosimetrykV beamslow energy dependencescintillator detector

Autor:innen und Institutionen

  • Department of Radiation Physics, Unit 1420, The University of Texas MD Anderson Cancer Center, 1515 Holcombe Boulevard, Houston, TX 77030, United States of America.

ORCID-Profile: Leonard Che Fru, Benjamin A Insley, Dongyeon Lee, Dirk A Bartkoski, Reza Reiazi, Aviad Schori, Ramesh Tailor, Mohammad Salehpour

Förderung

  • Convergent Radiotherapy and Radiosurgery Ltd
  • Convergent Radiosurgery and Radiotherapy Ltd.

Zitieren

APA

Che Fru, L., Insley, B. A., Lee, D., Bartkoski, D. A., Reiazi, R., Schori, A., Tailor, R., & Salehpour, M. (2025). Characterization of the Medscint HYPERSCINT scintillation dosimetry research platform with orthovoltage photon beams and a novel quasi monoenergetic beam. Physics in Medicine & Biology. https://doi.org/10.1088/1361-6560/ae0efb

Vancouver

Che Fru L, Insley BA, Lee D, Bartkoski DA, Reiazi R, Schori A, et al. Characterization of the Medscint HYPERSCINT scintillation dosimetry research platform with orthovoltage photon beams and a novel quasi monoenergetic beam. Physics in Medicine & Biology. 2025. doi:10.1088/1361-6560/ae0efb

BibTeX
@article{CheFru2025,
  author = {Che Fru, Leonard and Insley, Benjamin A and Lee, Dongyeon and Bartkoski, Dirk A and Reiazi, Reza and Schori, Aviad and Tailor, Ramesh and Salehpour, Mohammad},
  title = {Characterization of the Medscint HYPERSCINT scintillation dosimetry research platform with orthovoltage photon beams and a novel quasi monoenergetic beam},
  journal = {Physics in Medicine & Biology},
  year = {2025},
  doi = {10.1088/1361-6560/ae0efb}
}
Kostenlos starten →