Technical note: Visual, rapid, scintillation point dosimetry for in vivo MV photon beam radiotherapy treatments
Medical Physics · 2024
DOI: 10.1002/mp.17071 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, 2024. Verfasst von Savannah M Decker und 6 weiteren Autor:innen. Beteiligte Einrichtung: Thayer School of Engineering, Dartmouth College, Hanover, New Hampshire, USA. Hauptthemen laut MEDLINE: Photons, Scintillation Counting. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 23 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NCI NIH HHS, NIH HHS, NIBIB NIH HHS.
Abstract (Originalsprache)
Background
While careful planning and pre‐treatment checks are performed to ensure patient safety during external beam radiation therapy (EBRT), inevitable daily variations mean that in vivo dosimetry (IVD) is the only way to attain the true delivered dose. Several countries outside the US require daily IVD for quality assurance. However, elsewhere, the manual labor and time considerations of traditional in vivo dosimeters may be preventing frequent use of IVD in the clinic.
Purpose
This study expands upon previous research using plastic scintillator discs for optical dosimetry for electron therapy treatments. We present the characterization of scintillator discs for in vivo x‐ray dosimetry and describe additional considerations due to geometric complexities.
Methods
Plastic scintillator discs were coated with reflective white paint on all sides but the front surface. An anti‐reflective, matte coating was applied to the transparent face to minimize specular reflection. A time‐gated iCMOS camera imaged the discs under various irradiation conditions. In post‐processing, background‐subtracted images of the scintillators were fit with Gaussian‐convolved ellipses to extract several parameters, including integral output, and observation angle.
Results
Dose linearity and x‐ray energy independence were observed, consistent with ideal characteristics for a dosimeter. Dose measurements exhibited less than 5% variation for incident beam angles between 0° and 75° at the anterior surface and 0–60 at the posterior surface for exit beam dosimetry. Varying the angle between the disc surface and the camera lens did not impact the integral output for the same dose up to 55°. Past this point, up to 75°, there is a sharp falloff in response; however, a correction can be used based on the detected width of the disc. The reproducibility of the integral output for a single disc is 2%, and combined with variations from the gantry angle, we report the accuracy of the proposed scintillator disc dosimeters as ±5.4%.
Conclusions
Plastic scintillator discs have characteristics that are well‐suited for in vivo optical dosimetry for x‐ray radiotherapy treatments. Unlike typical point dosimeters, there is no inherent readout time delay, and an optical recording of the measurement is saved after treatment for future reference. While several factors influence the integral output for the same dose, they have been quantified here and may be corrected in post‐processing.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Thayer School of Engineering, Dartmouth College, Hanover, New Hampshire, USA.
- Geisel School of Medicine, Dartmouth College, Hanover, New Hampshire, USA.
- Dartmouth Cancer Center, Dartmouth Health, Lebanon, New Hampshire, USA.
Förderung
- NCI NIH HHS
- NIH HHS
- NIBIB NIH HHS
- National Institutes of Health
Zitieren
APA
Decker, S. M., Bruza, P., Zhang, R., Williams, B. B., Jarvis, L. A., Pogue, B. W., & Gladstone, D. J. (2024). Technical note: Visual, rapid, scintillation point dosimetry for in vivo MV photon beam radiotherapy treatments. Medical Physics. https://doi.org/10.1002/mp.17071
Vancouver
Decker SM, Bruza P, Zhang R, Williams BB, Jarvis LA, Pogue BW, et al. Technical note: Visual, rapid, scintillation point dosimetry for in vivo MV photon beam radiotherapy treatments. Medical Physics. 2024. doi:10.1002/mp.17071
BibTeX
@article{Decker2024,
author = {Decker, Savannah M and Bruza, Petr and Zhang, Rongxiao and Williams, Benjamin B and Jarvis, Lesley A and Pogue, Brian W and Gladstone, David J},
title = {Technical note: Visual, rapid, scintillation point dosimetry for in vivo MV photon beam radiotherapy treatments},
journal = {Medical Physics},
year = {2024},
doi = {10.1002/mp.17071}
}
EUVIMED