The PETcoil project: PET performance evaluation of two detector modules for a second generation RF-penetrable TOF-PET brain dedicated insert for simultaneous PET/MRI

Qian Dong, Chen-Ming Chang, Brian J Lee, Ilaria Sacco, Salar Sajedi, Zander Adams, Peter Fischer, Craig S Levin

Physics in Medicine & Biology · 2023

DOI: 10.1088/1361-6560/acc3f2 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Physics in Medicine & Biology, 2023. Verfasst von Qian Dong und 7 weiteren Autor:innen. Beteiligte Einrichtung: Molecular Imaging Instrumentation Laboratory, Stanford University, Stanford, CA, United States of America. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Brain. Die Arbeit wurde bisher 17-mal zitiert (Crossref) und nennt 45 Literaturangaben. Gefördert von: National Institutes of Health.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective . We are developing a portable, ‘RF-penetrable’, brain-dedicated time of flight (TOF)-PET insert (PETcoil) for simultaneous PET/MRI. Approach . In this paper, we evaluate the PET performance of two fully assembled detector modules for this insert design outside the MR room. Main results . The global coincidence time resolution, global 511 keV energy resolution, coincidence count rate, and detector temperature achieved over 2 h data collection were 242.2 ± 0.4 ps full width at half maximum (FWHM), 11.19% ± 0.02% FWHM, 22.0 ± 0.1 kcps, and 23.5 °C ± 0.3 °C, respectively. The intrinsic spatial resolutions in the axial and transaxial directions were 2.74 ± 0.01 mm FWHM and 2.88 ± 0.03 mm FWHM, respectively. Significance . These results demonstrate excellent TOF capability and the performance and stability necessary for scaling up to a full ring comprising 16 detector modules.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Magnetic Resonance ImagingBrainPositron-Emission TomographyPhantoms, ImagingTemperature

Stichwörter der Autor:innen

MR-compatiblePET insertPETcoilRF-penetrableTOFbrain imagingsimultaneous PET/MRI

Autor:innen und Institutionen

  • Molecular Imaging Instrumentation Laboratory, Stanford University, Stanford, CA, United States of America.
  • Department of Computer Science, Heidelberg University, Baden-Württemberg, Germany.

Förderung

  • National Institutes of Health

Zitieren

APA

Dong, Q., Chang, C. M., Lee, B. J., Sacco, I., Sajedi, S., Adams, Z., Fischer, P., & Levin, C. S. (2023). The PETcoil project: PET performance evaluation of two detector modules for a second generation RF-penetrable TOF-PET brain dedicated insert for simultaneous PET/MRI. Physics in Medicine & Biology. https://doi.org/10.1088/1361-6560/acc3f2

Vancouver

Dong Q, Chang CM, Lee BJ, Sacco I, Sajedi S, Adams Z, et al. The PETcoil project: PET performance evaluation of two detector modules for a second generation RF-penetrable TOF-PET brain dedicated insert for simultaneous PET/MRI. Physics in Medicine & Biology. 2023. doi:10.1088/1361-6560/acc3f2

BibTeX
@article{Dong2023,
  author = {Dong, Qian and Chang, Chen-Ming and Lee, Brian J and Sacco, Ilaria and Sajedi, Salar and Adams, Zander and Fischer, Peter and Levin, Craig S},
  title = {The PETcoil project: PET performance evaluation of two detector modules for a second generation RF-penetrable TOF-PET brain dedicated insert for simultaneous PET/MRI},
  journal = {Physics in Medicine & Biology},
  year = {2023},
  doi = {10.1088/1361-6560/acc3f2}
}