Fachartikel in der Zeitschrift Magnetic Resonance in Medicine, online erschienen am 14.09.2026. Verfasst von Sheng Chen und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, University of Virginia, Charlottesville, Virginia, USA. Hauptthemen laut MEDLINE: Skull, Brain, Magnetic Resonance Imaging, Imaging, Three-Dimensional. Die Arbeit nennt 52 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Focused Ultrasound Foundation, NIH HHS, Siemens Healthineers.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To develop an MRI approach for simultaneous skull and brain temperature monitoring between sonications in transcranial MR‐guided focused ultrasound (MRgFUS) surgeries.

Methods

A 3D stack‐of‐spirals dual‐echo ultra‐short TE (UTE) sequence incorporating fat suppression, spiral deblurring, trajectory measurement, and B1 correction was implemented. A variable TE strategy enabled UTEs as short as 0.05 ms for skull T1 mapping, while a fixed late‐TE (LTE) was used for brain temperature measurement. The approach was evaluated in laboratory and clinical settings through phantom and human studies.

Results

During laboratory phantom cooling, MR‐estimated temperatures closely matched optical fiber measurements, with standard deviations of 0.60°C and 0.57°C across two gel ROIs. Cortical bone T1 showed a strong correlation with temperature (Pearson r = 0.963, p < 0.05; slope = 2.51 ms/°C). Similar results were obtained during laboratory phantom heating ( r = 0.800, p < 0.05; slope = 3.86 ms/°C). In clinical phantom heating, gel temperature and bone T1 changed concurrently, with peak pixel‐wise temperature increases of ∼12°C and peak ROI‐averaged T1 increases of ∼14 ms. In patient scans, B1 correction reduced skull T1 spatial standard deviation from 98.1 to 59.9 ms. Volunteer uncertainty maps showed mean uncertainties of 46.8 ms for skull T1, enabling detection of ∼30 ms changes with 80% power, and 0.57°C for brain temperature.

Conclusions

The proposed method enabled inter‐sonication monitoring of skull temperature changes and brain thermometry during transcranial MRgFUS. While accurate ex vivo, improvements to in vivo T1 precision are needed for clinical temperature quantification.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

SkullBrainMagnetic Resonance ImagingImaging, Three-DimensionalBody TemperatureHumansPhantoms, ImagingMaleAdultFemaleTemperature

Stichwörter der Autor:innen

3DUTEspiralthermometrytranscranial MRgFUS

Autor:innen und Institutionen

  • Department of Biomedical Engineering, University of Virginia, Charlottesville, Virginia, USA.
  • Department of Electrical and Computer Engineering, Brigham Young University, Provo, Utah, USA.
  • Department of Radiology and Medical Imaging, University of Virginia, Charlottesville, Virginia, USA.

ORCID-Profile: Sheng Chen, Steven P Allen, G Wilson Miller

Förderung

  • Focused Ultrasound Foundation
  • NIH HHS
  • Siemens Healthineers
  • NIBIB NIH HHS
  • National Institutes of Health

Zitieren

APA

Chen, S., Allen, S. P., Mugler, J. P., Miller, G. W., & Meyer, C. H. (2027). Toward Inter‐Sonication Monitoring of Skull and Brain Temperatures in Transcranial MRgFUS Using a 3D Stack‐of‐Spirals Sequence. Magnetic Resonance in Medicine, 97(1), 17-30. https://doi.org/10.1002/mrm.70530

Vancouver

Chen S, Allen SP, Mugler JP, Miller GW, Meyer CH. Toward Inter‐Sonication Monitoring of Skull and Brain Temperatures in Transcranial MRgFUS Using a 3D Stack‐of‐Spirals Sequence. Magnetic Resonance in Medicine. 2027;97(1):17-30. doi:10.1002/mrm.70530

BibTeX
@article{Chen2027,
  author = {Chen, Sheng and Allen, Steven P and Mugler, John P and Miller, G Wilson and Meyer, Craig H},
  title = {Toward Inter‐Sonication Monitoring of Skull and Brain Temperatures in Transcranial MRgFUS Using a 3D Stack‐of‐Spirals Sequence},
  journal = {Magnetic Resonance in Medicine},
  year = {2027},
  volume = {97},
  number = {1},
  pages = {17--30},
  doi = {10.1002/mrm.70530}
}
Kostenlos starten →