A novel angle-independent neutron detector based on CR-39

Liuyu Cui iD, Weihai Zhuo, Bo Chen iD

Journal of Radiological Protection · 01.12.2026 · 46 (4) · 041509

DOI: 10.1088/1361-6498/aeb271 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Radiological Protection, online erschienen am 30.10.2026. Verfasst von Liuyu Cui und 2 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Radiation Medicine, Fudan University, Shanghai 200032, People's Republic of China. Hauptthemen laut MEDLINE: Neutrons, Radiometry. Die Arbeit nennt 28 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Natural Science Foundation of China.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

CR-39 offers numerous advantages for neutron detection, yet its strong angular dependence has long compromised the accuracy of measurements. To address this issue, this work proposes a novel cubic neutron detector consisting of six CR-39 elements. By taking the average response of the six elements as the measurement metric, the angular dependence is effectively suppressed. To accommodate the broad energy range of realistic neutron fields, a compact dual-converter structure is designed, in which polyethylene and boron nitride converters are coupled to the two faces of CR-39 element for fast- and thermal-neutron detection, respectively. Simulation results show that the angular variation in detector response remains below 12% for both fast- and thermal-neutron detection over the full solid angle, whereas a conventional single CR-39 element exhibits a response variation exceeding 300%. This superior angular independence has also been verified experimentally.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

NeutronsRadiometryEquipment DesignPolyethyleneEquipment Failure AnalysisPolyethylene Glycols

Stichwörter der Autor:innen

CR-39angular dependenceneutron detector

Substanzen

CR 39PolyethylenePolyethylene Glycols

Autor:innen und Institutionen

  • Institute of Radiation Medicine, Fudan University, Shanghai 200032, People's Republic of China.
  • Institute of Modern Physics, Department of Nuclear Science and Technology, Fudan University, Shanghai 200433, People's Republic of China.

ORCID-Profile: Liuyu Cui, Bo Chen

Förderung

  • National Natural Science Foundation of China

Zitieren

APA

Cui, L., Zhuo, W., & Chen, B. (2026). A novel angle-independent neutron detector based on CR-39. Journal of Radiological Protection, 46(4), 041509. https://doi.org/10.1088/1361-6498/aeb271

Vancouver

Cui L, Zhuo W, Chen B. A novel angle-independent neutron detector based on CR-39. Journal of Radiological Protection. 2026;46(4):041509. doi:10.1088/1361-6498/aeb271

BibTeX
@article{Cui2026,
  author = {Cui, Liuyu and Zhuo, Weihai and Chen, Bo},
  title = {A novel angle-independent neutron detector based on CR-39},
  journal = {Journal of Radiological Protection},
  year = {2026},
  volume = {46},
  number = {4},
  pages = {041509},
  doi = {10.1088/1361-6498/aeb271}
}