EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Dosimetric applicator characterization in hyperthermia: an automated quality assurance procedure based on relative specific absorption rate measurements

Timoteo Daniel Herrera iD, Remko Zweije, H Petra Kok, Johannes Crezee

Strahlentherapie und Onkologie · 2025

DOI: 10.1007/s00066-025-02489-7 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Strahlentherapie und Onkologie, 2025. Verfasst von Timoteo Daniel Herrera und 3 weiteren Autor:innen. Beteiligte Einrichtung: Radiation Oncology, Amsterdam UMC location University of Amsterdam, Meibergdreef 9, 1105AZ, Amsterdam, The Netherlands. T.D.Herrera@amsterdamumc.nl. Hauptthemen laut MEDLINE: Quality Assurance, Health Care, Radiometry, Hyperthermia, Induced, Absorption, Radiation. Die Arbeit nennt 39 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: HORIZON EUROPE Marie Sklodowska-Curie Actions.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Ensuring reproducible treatments and therapeutic temperature rises is pivotal for wider clinical application of hyperthermia. We propose a fast automated quality assurance (QA) procedure using E‑field measurements for relative specific absorption rate (SAR) applicator characterization, with a similar workflow to radiotherapy dosimetry.

Methods

Our procedure was demonstrated for 434 MHz superficial contact flexible microstrip applicators (CFMA) with similar antennas and geometry to the ALBA ON 4000D system. Applicators were placed on a liquid saline phantom. A Cartesian robot performed E‑field measurements moving E‑field sensors through the phantom. We analysed the influence of spatial resolution on effective field size (EFS) measurements. Using volume measurements, we established effective penetration depth (EPD) variation across the aperture. We evaluated repeatability, measuring central planes on eight different days. Applicators/phantoms were characterized with flat and curved setups, using standard clinical and more excessive bolus thicknesses.

Results

Procedures take ~40 min per setup condition. One-centimetre spatial resolution appears sufficient for QA. EPD showed errors < 5% when determined from high-SAR regions (~80–90% of the maximum). The EFS and EPD variability between different days was < 5%. Increasing EPD and decreasing EFS with increasing curvature was observed when using the clinical bolus thickness, with more homogeneous SAR distributions for curved than for flat setups. Excessive bolus thickness resulted in irregular SAR distributions and larger EFS and EPD variations.

Conclusion

The proposed QA procedure is characterized by fast, practical and reproducible measurements that are suitable for efficiently evaluating various setup conditions. This flexible workflow can also be used with other radiative hyperthermia systems.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Quality Assurance, Health CareRadiometryHyperthermia, InducedAbsorption, RadiationHumansPhantoms, ImagingReproducibility of ResultsEquipment DesignEquipment Failure AnalysisSensitivity and Specificity

Stichwörter der Autor:innen

Effective field sizeEffective penetration depthElectric field measurementMicrowave hyperthermia systemsRepeatability evaluation

Autor:innen und Institutionen

  • Radiation Oncology, Amsterdam UMC location University of Amsterdam, Meibergdreef 9, 1105AZ, Amsterdam, The Netherlands. T.D.Herrera@amsterdamumc.nl.
  • Treatment and quality of life, Cancer biology and immunology, Cancer Center Amsterdam, Amsterdam, The Netherlands. T.D.Herrera@amsterdamumc.nl.
  • Radiation Oncology, Amsterdam UMC location University of Amsterdam, Meibergdreef 9, 1105AZ, Amsterdam, The Netherlands.
  • Treatment and quality of life, Cancer biology and immunology, Cancer Center Amsterdam, Amsterdam, The Netherlands.

ORCID-Profile: Timoteo Daniel Herrera

Förderung

  • HORIZON EUROPE Marie Sklodowska-Curie Actions

Zitieren

APA

Herrera, T. D., Zweije, R., Kok, H. P., & Crezee, J. (2025). Dosimetric applicator characterization in hyperthermia: an automated quality assurance procedure based on relative specific absorption rate measurements. Strahlentherapie und Onkologie. https://doi.org/10.1007/s00066-025-02489-7

Vancouver

Herrera TD, Zweije R, Kok HP, Crezee J. Dosimetric applicator characterization in hyperthermia: an automated quality assurance procedure based on relative specific absorption rate measurements. Strahlentherapie und Onkologie. 2025. doi:10.1007/s00066-025-02489-7

BibTeX
@article{Herrera2025,
  author = {Herrera, Timoteo Daniel and Zweije, Remko and Kok, H Petra and Crezee, Johannes},
  title = {Dosimetric applicator characterization in hyperthermia: an automated quality assurance procedure based on relative specific absorption rate measurements},
  journal = {Strahlentherapie und Onkologie},
  year = {2025},
  doi = {10.1007/s00066-025-02489-7}
}
Kostenlos starten →