Theoretical and experimental calibration of hot-sphere anemometers under low-pressure and simulated microgravity conditions
Physics of Fluids · 01.07.2026 · 38 (7)
DOI: 10.1063/5.0342455 ↗Kurz gefasst
Fachartikel in der Zeitschrift Physics of Fluids, online erschienen am 23.07.2026. Verfasst von Jinhui Tang und 5 weiteren Autor:innen. Die Arbeit nennt 39 Literaturangaben. Gefördert von: Discretionary Fund of National Science Center for earthquake Engineering.
Abstract (Originalsprache)
Hot-sphere anemometers exhibit significant measurement deviations in microgravity-related and low-pressure environments, creating a need for recalibration. In this study, a compact low-pressure and low-velocity wind tunnel calibration system is developed by combining heat transfer theory with ground-based experiments. Based on the Gr/Re2 criterion and Kn number analysis, it is verified that the flow remains in the continuum regime at 40 kPa. In addition, buoyancy-induced natural convection is sufficiently suppressed when the wind velocity exceeds 0.097 m/s, demonstrating the feasibility of extrapolating ground-based calibration results to microgravity-related forced-convection conditions. The results show that the initial measurement error of the probe under atmospheric pressure reaches 19.84%. After recalibration, the correlation coefficient (R2) increases to 0.9993, with an expanded uncertainty ranging from 0.036 to 0.039 m/s. At low-pressures (90–40 kPa), the measured wind velocity exhibits an attenuation trend, with a maximum attenuation rate of 66.02% relative to standard atmospheric pressure. Although thermal radiation accounts for only 4.5% of the total heat dissipation under the reference condition, its relative contribution increases under low-pressure and low-velocity conditions; therefore, it is treated as a secondary heat-loss pathway. With conductive heat loss through the wire neglected, the energy conservation equation is modified, and a forced-convection-dominated empirical calibration correlation is established, with residual heat-loss effects incorporated into the fitted coefficients. This correlation enables low-pressure wind velocities to be estimated from atmospheric-pressure baseline calibration data within the calibrated range, with a prediction error of less than 10%. This study provides a reliable calibration method for hot-sphere anemometers under microgravity-related low-pressure conditions.
Förderung
- Discretionary Fund of National Science Center for earthquake Engineering
Zitieren
APA
Tang, J., Shi, B., Liu, T., Sun, H., Liu, J., & Zhu, R. (2026). Theoretical and experimental calibration of hot-sphere anemometers under low-pressure and simulated microgravity conditions. Physics of Fluids, 38(7). https://doi.org/10.1063/5.0342455
Vancouver
Tang J, Shi B, Liu T, Sun H, Liu J, Zhu R. Theoretical and experimental calibration of hot-sphere anemometers under low-pressure and simulated microgravity conditions. Physics of Fluids. 2026;38(7). doi:10.1063/5.0342455
BibTeX
@article{Tang2026,
author = {Tang, Jinhui and Shi, Baolu and Liu, Tongyu and Sun, Hejiang and Liu, Junjie and Zhu, Ran},
title = {Theoretical and experimental calibration of hot-sphere anemometers under low-pressure and simulated microgravity conditions},
journal = {Physics of Fluids},
year = {2026},
volume = {38},
number = {7},
doi = {10.1063/5.0342455}
}
EUVIMED