Determining the Effect of Scan Time and Reconstruction Algorithm on Lesion Signal Strength and Background Noise for Oncologic FDG Studies Using a Next Generation 148cm LAFOV PET System
Nuklearmedizin - NuclearMedicine · 01.04.2026 · 65 (02) · 160-161
DOI: 10.1055/s-0046-1818335 ↗Kurz gefasst
Fachartikel in der Zeitschrift Nuklearmedizin - NuclearMedicine, erschienen am 01.04.2026. Verfasst von P Rath und 7 weiteren Autor:innen.
Abstract (Originalsprache)
Ziel/Aim: Next generation LAFOV PET offers unique sensitivity for clinical scans, but acquisition protocols are not yet standardized. We evaluated the effect of shorter scan time and novel reconstruction techniques on static image quality for oncologic FDG studies. Methodik/Methods: Using the Panorama GS 148cm LAFOV PET/CT (United Imaging Healthcare), 10 patients with various malignancies underwent a 5 min list-mode F-18-FDG scan (standard dose 3 MBq/kg bodyweight) and low-dose CT for attenuation correction. List-mode data were re-sampled to 300, 150, 60, 30 and 15sec static datasets. Each was reconstructed using standard OSEM iterative algorithm, a manufacturer-provided advanced iterative algorithm (hyper iterative, HI), and a manufacturer-provided AI-based algorithm (deep progressive reconstruction, DPR). Lesion signal strength was measured by SUVpeak and SUVmax in up to 3 lesions per patient, and change was defined relative to OSEM 300s as reference. Automated CT-based organ segmentation was employed and Coefficient of Variation (CoV) for SUV mean of the lesion-free liver parenchyma was defined as a measure of background noise. Ergebnisse/Results: CoV increases for shorter OSEM reconstruction times from 21% to 29%. For HI and DPR, it remains stable at 21-22%. Only for the shortest acquisition time of 15s, DPR yields an elevated CoV of 51%. Relative change of lesional SUVpeak was low at 1-5% between all scan times for all 3 reconstruction methods (only DPR 60s: 12%). For SUVmax, shorter OSEM scan time increases relative change up to 11%, HI and DPR reconstructions return higher lesional SUVmax by up to 18% and 29% on average, respectively. Schlussfolgerungen/Conclusion: Using standard injected doses of FDG, next-generation LAFOV PET enables a reduction of acquisition time down to 30 sec without loss of lesion signal strength or increase of background noise. Novel instrumentation-specific and AI-based reconstruction algorithms further improve lesion signal strength but may not be used interchangeably for SUV-based image analysis. Their system-specific effects on LAFOV PET accuracy warrant further investigation. Publication History Article published online: 02 April 2026 © 2026. Thieme. All rights reserved. Georg Thieme Verlag KG Oswald-Hesse-Straße 50, 70469 Stuttgart, Germany
Zitieren
APA
Rath, P., Kröhn, S., Nikoukar, L. R., Weiberg, D., Uhlemann, A., Henning, K., Ziegler, S., & Bengel, F. (2026). Determining the Effect of Scan Time and Reconstruction Algorithm on Lesion Signal Strength and Background Noise for Oncologic FDG Studies Using a Next Generation 148cm LAFOV PET System. Nuklearmedizin - NuclearMedicine, 65(02), 160-161. https://doi.org/10.1055/s-0046-1818335
Vancouver
Rath P, Kröhn S, Nikoukar LR, Weiberg D, Uhlemann A, Henning K, et al. Determining the Effect of Scan Time and Reconstruction Algorithm on Lesion Signal Strength and Background Noise for Oncologic FDG Studies Using a Next Generation 148cm LAFOV PET System. Nuklearmedizin - NuclearMedicine. 2026;65(02):160-161. doi:10.1055/s-0046-1818335
BibTeX
@article{Rath2026,
author = {Rath, P and Kröhn, S and Nikoukar, L R and Weiberg, D and Uhlemann, A and Henning, K and Ziegler, S and Bengel, F},
title = {Determining the Effect of Scan Time and Reconstruction Algorithm on Lesion Signal Strength and Background Noise for Oncologic FDG Studies Using a Next Generation 148cm LAFOV PET System},
journal = {Nuklearmedizin - NuclearMedicine},
year = {2026},
volume = {65},
number = {02},
pages = {160--161},
doi = {10.1055/s-0046-1818335}
}
EUVIMED