Effect of Plate Length on Construct Stiffness and Strain in a Synthetic Short-Fragment Fracture Gap Model Stabilized with a 3.5-mm Locking Compression Plate
Veterinary and Comparative Orthopaedics and Traumatology · 01.03.2025 · 38 (02) · 063-070
DOI: 10.1055/s-0044-1789263 ↗Kurz gefasst
Fachartikel in der Zeitschrift Veterinary and Comparative Orthopaedics and Traumatology, erschienen am 01.03.2025. Verfasst von Fabian N Trefny und 4 weiteren Autor:innen. Beteiligte Einrichtung: Division of Health Sciences, School of Veterinary Medicine, Murdoch University, Perth, Western Australia, Australia. Hauptthemen laut MEDLINE: Bone Plates, Fractures, Bone, Fracture Fixation, Internal. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 39 Literaturangaben.
Abstract (Originalsprache)
Objective
To evaluate the effect of 3.5-mm locking compression plate (LCP) length on construct stiffness and plate and bone model strain in a synthetic, short-fragment, fracture-gap model. Study Design Six replicates of 6-hole, 8-hole, 10-hole, and 12-hole LCP constructs on a short-fragment, tubular Delrin fracture gap model underwent four-point compression and tension bending. Construct stiffness and surface strain, calculated using three-dimensional digital image correlation, were compared across plate length and region of interest (ROI) on the construct.
Results
The 12-hole plates (80% plate–bone ratio) had significantly higher construct stiffness than 6-hole, 8-hole, and 10-hole plates and significantly lower plate strain than 6-hole plates at all ROIs. Strain on the bone model was significantly lower in constructs with 10-hole and 12-hole plates than 6-hole plates under both compression and tension bending.
Conclusion
Incremental increases in construct stiffness and incremental decreases in plate strain were only identified when comparing 6-hole, 8-hole, and 10-hole plates to 12-hole plates, and 6-hole to 12-hole plates, respectively. Strain on the bone model showed an incremental decrease when comparing 6-hole to 10-hole and 12-hole plates. A long plate offered biomechanical advantages of increased construct stiffness and reduced plate and bone model strain, over a short plate in this in vitro model.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Division of Health Sciences, School of Veterinary Medicine, Murdoch University, Perth, Western Australia, Australia.
- Department of Medical Engineering and Physics, Royal Perth Hospital, Perth, Australia.
ORCID-Profile: Mark Glyde
Zitieren
APA
Trefny, F. N., Glyde, M., Hosgood, G., Hayes, A., & Day, R. (2025). Effect of Plate Length on Construct Stiffness and Strain in a Synthetic Short-Fragment Fracture Gap Model Stabilized with a 3.5-mm Locking Compression Plate. Veterinary and Comparative Orthopaedics and Traumatology, 38(02), 063-070. https://doi.org/10.1055/s-0044-1789263
Vancouver
Trefny FN, Glyde M, Hosgood G, Hayes A, Day R. Effect of Plate Length on Construct Stiffness and Strain in a Synthetic Short-Fragment Fracture Gap Model Stabilized with a 3.5-mm Locking Compression Plate. Veterinary and Comparative Orthopaedics and Traumatology. 2025;38(02):063-070. doi:10.1055/s-0044-1789263
BibTeX
@article{Trefny2025,
author = {Trefny, Fabian N and Glyde, Mark and Hosgood, Giselle and Hayes, Alex and Day, Robert},
title = {Effect of Plate Length on Construct Stiffness and Strain in a Synthetic Short-Fragment Fracture Gap Model Stabilized with a 3.5-mm Locking Compression Plate},
journal = {Veterinary and Comparative Orthopaedics and Traumatology},
year = {2025},
volume = {38},
number = {02},
pages = {063--070},
doi = {10.1055/s-0044-1789263}
}
EUVIMED