Biomechanical analysis of orthogonal and unilateral locking plate constructs in a fracture gap model

Peter J Welsh iD, Adam M Nylund, Peter J Gilbert, Nick A Smith, Lloyd V Smith iD

Veterinary Surgery · 2025

DOI: 10.1111/vsu.14206 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Veterinary Surgery, 2025. Verfasst von Peter J Welsh und 4 weiteren Autor:innen. Beteiligte Einrichtung: WestVet Emergency and Specialty Center, Garden City, Idaho, USA. Hauptthemen laut MEDLINE: Bone Plates, Fracture Fixation, Internal, Fractures, Bone. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 40 Literaturangaben. Gefördert von: Arthrex, College of Veterinary Medicine, Washington State University.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

To compare unilaterally plated (UP) constructs and orthogonally plated (OP) constructs of differing sizes (3.5 mm + 2.0 mm locking plates [OP2.0], 3.5 mm + 2.4 mm locking plates [OP2.4] and 3.5 mm + 3.0 mm locking plates [OP3.0]) under cyclic and static axial compression load to failure testing using a fracture gap model. Study design In vitro experimental biomechanical study. Sample population Acetal homopolymer (Delrin) rods stabilized with locking plates and screws (Arthrex OrthoLine).

Methods

One of four stabilization techniques (UP, OP2.0, OP2.4, OP3.0) was applied to rods with a fixed fracture gap. Constructs were fatigue tested under axial compression (90,000 cycles; 4–196 N) followed by static load to failure. Cyclic displacement was evaluated after the first, middle, and last 100 cycles. Stiffness and strength were analyzed during static axial compression load to failure.

Results

During cyclic testing, UP experienced 3.5, 3.8, and 4.1 times the gap strain of OP2.0, OP2.4, and OP3.0, respectively ( p < 0.0075). Fatigue and construct design had significant effects on displacement ( p < .0001). OP2.0, OP2.4, and OP3.0 demonstrated 2.5, 3.0, and 4.1 times the strength and 3.0, 3.6, and 4.2 times the stiffness of UP, respectively ( p < .0002).

Conclusion

In the present in vitro fracture gap model, OP constructs were stronger and stiffer than UP under dynamic and static axial compression, and OP stiffness increased with increasing implant size. Clinical significance The results of this study demonstrate objective biomechanical advantages of OP compared to UP. Based on these results, orthogonal plating can be considered when increased fixation strength and stiffness are warranted.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Bone PlatesFracture Fixation, InternalFractures, BoneBiomechanical PhenomenaAnimalsMaterials TestingBone Screws

Autor:innen und Institutionen

  • WestVet Emergency and Specialty Center, Garden City, Idaho, USA.
  • Department of Veterinary Clinical Sciences, College of Veterinary Medicine, Washington State University, Pullman, Washington, USA.
  • School of Mechanical and Materials Engineering, Voiland College of Engineering and Architecture, Washington State University, Pullman, Washington, USA.

ORCID-Profile: Peter J Welsh, Lloyd V Smith

Förderung

  • Arthrex
  • College of Veterinary Medicine, Washington State University

Zitieren

APA

Welsh, P. J., Nylund, A. M., Gilbert, P. J., Smith, N. A., & Smith, L. V. (2025). Biomechanical analysis of orthogonal and unilateral locking plate constructs in a fracture gap model. Veterinary Surgery. https://doi.org/10.1111/vsu.14206

Vancouver

Welsh PJ, Nylund AM, Gilbert PJ, Smith NA, Smith LV. Biomechanical analysis of orthogonal and unilateral locking plate constructs in a fracture gap model. Veterinary Surgery. 2025. doi:10.1111/vsu.14206

BibTeX
@article{Welsh2025,
  author = {Welsh, Peter J and Nylund, Adam M and Gilbert, Peter J and Smith, Nick A and Smith, Lloyd V},
  title = {Biomechanical analysis of orthogonal and unilateral locking plate constructs in a fracture gap model},
  journal = {Veterinary Surgery},
  year = {2025},
  doi = {10.1111/vsu.14206}
}
Kostenlos starten →