Force Analysis Using Self-Expandable Valve Fluoroscopic Imaging: a way Through Artificial Intelligence
Journal of Cardiovascular Translational Research · 2024
DOI: 10.1007/s12265-024-10550-6 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Cardiovascular Translational Research, 2024. Verfasst von Yiming Qi und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Institute of Cardiovascular Diseases, Zhongshan Hospital, Fudan University, Shanghai , 180 Fenglin Road, Shanghai, China. Hauptthemen laut MEDLINE: Heart Valve Prosthesis, Transcatheter Aortic Valve Replacement, Aortic Valve, Prosthesis Design. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 17 Literaturangaben. Gefördert von: National Nature Science Foundation of China, Shanghai Top Priority Research Center.
Abstract (Originalsprache)
This study aimed to develop a force analysis model correlating fluoroscopic images of self-expandable valves with stress distribution. For this purpose, a nonmetallic measuring device designed to apply diverse forces at specific positions on a valve stent while simultaneously measuring force magnitude was manufactured, obtaining 465 sets of fluorescent films under different force conditions, resulting in 5580 images and their corresponding force tables. Using the XrayGLM, a mechanical analysis model based on valve fluorescence images was trained. The accuracy of the image force analysis using this model was approximately 70% (50-88.3%), with a relative accuracy of 93.3% (75-100%). This confirms that fluoroscopic images of transcatheter aortic valve replacement (TAVR) valve stents contain a wealth of mechanical information, and machine learning can be used to train models to recognize the relationship between stent images and force distribution, enhancing the understanding of TAVR complications.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiology, Institute of Cardiovascular Diseases, Zhongshan Hospital, Fudan University, Shanghai , 180 Fenglin Road, Shanghai, China.
- National Clinical Research Center for Interventional Medicine, 180 Fenglin Road, Shanghai, China.
- Department of Nursing, Zhongshan Hospital, Fudan University, 180 Fenglin Road, Shanghai, China.
- Department of Cardiology, Institute of Cardiovascular Diseases, Zhongshan Hospital, Fudan University, Shanghai , 180 Fenglin Road, Shanghai, China. peden@sina.com.
- National Clinical Research Center for Interventional Medicine, 180 Fenglin Road, Shanghai, China. peden@sina.com.
- Department of Cardiology, Institute of Cardiovascular Diseases, Zhongshan Hospital, Fudan University, Shanghai , 180 Fenglin Road, Shanghai, China. daxin_zhou@163.com.
ORCID-Profile: Daxin Zhou
Förderung
- National Nature Science Foundation of China
- Shanghai Top Priority Research Center
Zitieren
APA
Qi, Y., Zhang, X., Shen, Z., Liang, Y., Chen, S., Pan, W., Zhou, D., & Ge, J. (2024). Force Analysis Using Self-Expandable Valve Fluoroscopic Imaging: a way Through Artificial Intelligence. Journal of Cardiovascular Translational Research. https://doi.org/10.1007/s12265-024-10550-6
Vancouver
Qi Y, Zhang X, Shen Z, Liang Y, Chen S, Pan W, et al. Force Analysis Using Self-Expandable Valve Fluoroscopic Imaging: a way Through Artificial Intelligence. Journal of Cardiovascular Translational Research. 2024. doi:10.1007/s12265-024-10550-6
BibTeX
@article{Qi2024,
author = {Qi, Yiming and Zhang, Xiaochun and Shen, Zhiyun and Liang, Yixiu and Chen, Shasha and Pan, Wenzhi and Zhou, Daxin and Ge, Junbo},
title = {Force Analysis Using Self-Expandable Valve Fluoroscopic Imaging: a way Through Artificial Intelligence},
journal = {Journal of Cardiovascular Translational Research},
year = {2024},
doi = {10.1007/s12265-024-10550-6}
}
EUVIMED