Endoscopic Polyp Size Measurement Using a Laser-Based Sizing Probe: Pilot Study
Endoscopy International Open · 23.06.2026 · 14 (CP)
DOI: 10.1055/a-2872-7167 ↗Kurz gefasst
Fachartikel in der Zeitschrift Endoscopy International Open, erschienen am 23.06.2026. Verfasst von Mohamed Zineddine Mahdadi und 13 weiteren Autor:innen. Beteiligte Einrichtung: MedicineCentre Hospitalier de l'Universite de MontréalMontrealCanada. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Background and study aims Accurate measurement of colorectal polyp size is important for planning resection and post-polypectomy surveillance. This study evaluated a through-the-scope laser-based probe for polyp sizing during colonoscopy. Patients and methods A prospective pilot study was conducted at CHUM, where polyps were measured using a laser-based probe. The primary outcome was accuracy of the laser-based probe, defined as mean bias with microscopic sizing as the reference standard. Secondary outcomes included agreement with the reference, implementation rate, technical success rate, reasons for failure, and per-polyp measurement time.
Results
Among 323 patients, 402 colorectal polyps were detected. Laser-based measurement was attempted in 237 polyps (59.0% 237/402; 95% confidence interval [CI] 54.0–63.8), of which 149 had microscopic reference measurements. The technical success rate was 40.9% (97/237; 95% CI 34.6–47.5), with a mean measurement time of 1:35. Accounting for non-en-bloc resections, 41 of 149 polyps (27.5%; 95% CI 20.7–35.5) had paired laser-microscopy measurements for the primary analysis. Laser-based measurement showed minimal bias (−0.10 mm; 95% CI −0.30 to 0.10; P = 0.328), meeting the prespecified noninferiority margin of 0.4 mm. Agreement with the reference standard was high (intraclass correlation coefficient = 0.91; 95% CI 0.83–0.95), with near-perfect categorical concordance (κ = 0.85) and 95.1% classification accuracy (39/41; 95% CI 82.2–99.2). Bland–Altman limits of agreement ranged from −1.34 to 1.14 mm.
Conclusions
Laser-based measurement provided accurate polyp sizing when obtained. However, technical success was limited by longer measurement times and device issues.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- MedicineCentre Hospitalier de l'Universite de MontréalMontrealCanada.
- Hospital Research Center (CRCHUM)University of MontrealMontrealCanada.
ORCID-Profile: Roupen Djinbachian, Daniel von Renteln
Zitieren
APA
Mahdadi, M. Z., Cristea, É., Oleksiw, M., Aleksieva, P., Michal, V., Battat, R., Djinbachian, R., Gefflot, C., Daoud, D. C., Bouchard, S., Bouin, M., Liu Chen Kiow, J., Panzini, B. P., & von Renteln, D. (2026). Endoscopic Polyp Size Measurement Using a Laser-Based Sizing Probe: Pilot Study. Endoscopy International Open, 14(CP). https://doi.org/10.1055/a-2872-7167
Vancouver
Mahdadi MZ, Cristea É, Oleksiw M, Aleksieva P, Michal V, Battat R, et al. Endoscopic Polyp Size Measurement Using a Laser-Based Sizing Probe: Pilot Study. Endoscopy International Open. 2026;14(CP). doi:10.1055/a-2872-7167
BibTeX
@article{Mahdadi2026,
author = {Mahdadi, Mohamed Zineddine and Cristea, Éric and Oleksiw, Megan and Aleksieva, Preslava and Michal, Victoire and Battat, Robert and Djinbachian, Roupen and Gefflot, Claire and Daoud, Dane Christina and Bouchard, Simon and Bouin, Mickael and Liu Chen Kiow, Jeremy and Panzini, Benoit P and von Renteln, Daniel},
title = {Endoscopic Polyp Size Measurement Using a Laser-Based Sizing Probe: Pilot Study},
journal = {Endoscopy International Open},
year = {2026},
volume = {14},
number = {CP},
doi = {10.1055/a-2872-7167}
}