Artificial intelligence-based polyp size measurement in gastrointestinal endoscopy using the auxiliary waterjet as a reference

Boban Sudarevic, Philipp Sodmann, Ioannis Kafetzis, Joel Troya iD, Thomas J Lux, Zita Saßmannshausen, Katja Herlod, Stefan A Schmidt iD, Markus Brand iD, Katrin Schöttker, Wolfram G Zoller, Alexander Meining und 1 weitere

Endoscopy · 2023-09 · 55 (09) · 871-876

DOI: 10.1055/a-2077-7398 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Endoscopy, online erschienen am 12.07.2023. Verfasst von Boban Sudarevic und 12 weiteren Autor:innen. Beteiligte Einrichtung: Interventional and Experimental Endoscopy (InExEn), Internal Medicine II, University Hospital Würzburg, Würzburg, Germany. Hauptthemen laut MEDLINE: Colonic Polyps, Colonography, Computed Tomographic, Colorectal Neoplasms. Die Arbeit wurde bisher 56-mal zitiert (Crossref) und nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Bavarian Center for Cancer Research, Eva Mayr-Stihl Stiftung, Fischerwerke GmbH & Co. KG.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Measurement of colorectal polyp size during endoscopy is mainly performed visually. In this work, we propose a novel polyp size measurement system (Poseidon) based on artificial intelligence (AI) using the auxiliary waterjet as a measurement reference.

Methods

Visual estimation, biopsy forceps-based estimation, and Poseidon were compared using a computed tomography colonography-based silicone model with 28 polyps of defined sizes. Four experienced gastroenterologists estimated polyp sizes visually and with biopsy forceps. Furthermore, the gastroenterologists recorded images of each polyp with the waterjet in proximity for the application of Poseidon. Additionally, Poseidon's measurements of 29 colorectal polyps during routine clinical practice were compared with visual estimates.

Results

In the silicone model, visual estimation had the largest median percentage error of 25.1 % (95 %CI 19.1 %–30.4 %), followed by biopsy forceps-based estimation: median 20.0 % (95 %CI 14.4 %–25.6 %). Poseidon gave a significantly lower median percentage error of 7.4 % (95 %CI 5.0 %–9.4 %) compared with other methods. During routine colonoscopies, Poseidon presented a significantly lower median percentage error (7.7 %, 95 %CI 6.1 %–9.3 %) than visual estimation (22.1 %, 95 %CI 15.1 %–26.9 %).

Conclusion

In this work, we present a novel AI-based method for measuring colorectal polyp size with significantly higher accuracy than other common sizing methods.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Colonic PolypsColonography, Computed TomographicColorectal NeoplasmsHumansArtificial IntelligenceColonoscopySurgical Instruments

Autor:innen und Institutionen

  • Interventional and Experimental Endoscopy (InExEn), Internal Medicine II, University Hospital Würzburg, Würzburg, Germany.
  • Department of Internal Medicine and Gastroenterology, Katharinenhospital, Stuttgart, Germany.
  • Department of Diagnostic and Interventional Radiology, University Hospital Ulm, Ulm, Germany.

ORCID-Profile: Joel Troya, Stefan A Schmidt, Markus Brand, Alexander Hann

Förderung

  • Bavarian Center for Cancer Research
  • Eva Mayr-Stihl Stiftung
  • Fischerwerke GmbH & Co. KG
  • Forum Gesundheitsstandort Baden-Württemberg
  • Dieter von Holtzbrinck Foundation GmbH

Zitieren

APA

Sudarevic, B., Sodmann, P., Kafetzis, I., Troya, J., Lux, T. J., Saßmannshausen, Z., Herlod, K., Schmidt, S. A., Brand, M., Schöttker, K., Zoller, W. G., Meining, A., & Hann, A. (2023). Artificial intelligence-based polyp size measurement in gastrointestinal endoscopy using the auxiliary waterjet as a reference. Endoscopy, 55(09), 871-876. https://doi.org/10.1055/a-2077-7398

Vancouver

Sudarevic B, Sodmann P, Kafetzis I, Troya J, Lux TJ, Saßmannshausen Z, et al. Artificial intelligence-based polyp size measurement in gastrointestinal endoscopy using the auxiliary waterjet as a reference. Endoscopy. 2023;55(09):871-876. doi:10.1055/a-2077-7398

BibTeX
@article{Sudarevic2023,
  author = {Sudarevic, Boban and Sodmann, Philipp and Kafetzis, Ioannis and Troya, Joel and Lux, Thomas J and Saßmannshausen, Zita and Herlod, Katja and Schmidt, Stefan A and Brand, Markus and Schöttker, Katrin and Zoller, Wolfram G and Meining, Alexander and Hann, Alexander},
  title = {Artificial intelligence-based polyp size measurement in gastrointestinal endoscopy using the auxiliary waterjet as a reference},
  journal = {Endoscopy},
  year = {2023},
  volume = {55},
  number = {09},
  pages = {871--876},
  doi = {10.1055/a-2077-7398}
}