Evaluation of image quality in five different handheld ultrasound devices and analysis of various impact factors
Ultraschall in der Medizin - European Journal of Ultrasound · 26.03.2026
DOI: 10.1055/a-2809-7627 ↗Kurz gefasst
Fachartikel in der Zeitschrift Ultraschall in der Medizin - European Journal of Ultrasound, erschienen am 26.03.2026. Verfasst von Antje Naas und 33 weiteren Autor:innen. Beteiligte Einrichtung: Else Kröner Fresenius Center for Digital Health, TU Dresden Faculty of Medicine Carl Gustav Carus, Dresden, Germany. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Handheld ultrasound (HHUS) devices are increasingly used in clinical practice due to their portability and cost-effectiveness. However, HHUS image quality remains variable, and a standardized expert-based evaluation protocol is lacking. This study compares the image quality of B-mode recordings of five HHUS models with a high-end reference system (HEUS) using a blinded assessment to identify differences, covariates, offering a standardized evaluation framework. In this blinded, multi-center study, video sequences of the liver, pancreas, and sigmoid colon were recorded in ten volunteers using 5 HHUS devices – Vscan Air, Butterfly iQ+, Philips Lumify, Clarius C3, and Clarius C3 HD3 – and one HEUS (GE Logiq E10). Twenty-two physicians with many years of abdominal ultrasound experience rated blinded clips for 5 image quality parameters. The HEUS consistently received the highest ratings. Image quality differed significantly between HHUS devices. Clarius systems showed the smallest deviation from the reference, followed by Vscan Air and Philips Lumify. All image parameters significantly correlated with the overall impression, without a single dominant factor. The patients’ BMI had no relevant influence. Examiner experience was less impactful than device familiarity and personal preference. The image quality of HEUS remained superior to all tested HHUS devices. Further subdivision of image quality aspects did not provide additional information. Furthermore, we could not identify significant covariates in our setup. These findings could justify simplified setups with reduced questionnaires and requirements for examiners.
Autor:innen und Institutionen
- Else Kröner Fresenius Center for Digital Health, TU Dresden Faculty of Medicine Carl Gustav Carus, Dresden, Germany.
- Medical Department 1, Dresden University Hospital, Dresden, Germany.
- Vodafone Chair for Mobile Communications, TU Dresden Faculty of Electrical and Computer Engineering, Dresden, Germany.
- Department of Gastroenterology, Asklepios Hospital Radeberg, Radeberg, Germany.
- Gastroenterologie, Upper Lusatia Hospital Bischofswerda Branch, Bischofswerda, Germany.
- Medical Department I, Dresden University Hospital, Dresden, Germany.
ORCID-Profile: Antje Naas, Julian Kober, Tönnis Trittler, Edgar Manfred Gustav Dorausch, Tina Gabriel, Martin Franck, Christian Jenssen, Valentin Blank, Thomas Karlas, Jochen Hampe, Moritz Herzog
Zitieren
APA
Naas, A., Kober, J., Trittler, T., Nauber, R., Dorausch, E. M. G., Kampfrath, N., Brinkmann, F., Arsova, M., Husman, J., Langanke, R., Sulk, S., Matthes, K., Gabriel, T., Teufel, A., Behrendt, P., Markova, A., Franck, M., Jenssen, C., Merkel, D., …, & Herzog, M. (2026). Evaluation of image quality in five different handheld ultrasound devices and analysis of various impact factors. Ultraschall in der Medizin - European Journal of Ultrasound. https://doi.org/10.1055/a-2809-7627
Vancouver
Naas A, Kober J, Trittler T, Nauber R, Dorausch EMG, Kampfrath N, et al. Evaluation of image quality in five different handheld ultrasound devices and analysis of various impact factors. Ultraschall in der Medizin - European Journal of Ultrasound. 2026. doi:10.1055/a-2809-7627
BibTeX
@article{Naas2026,
author = {Naas, Antje and Kober, Julian and Trittler, Tönnis and Nauber, Richard and Dorausch, Edgar Manfred Gustav and Kampfrath, Nicole and Brinkmann, Franz and Arsova, Maia and Husman, Julia and Langanke, Robert and Sulk, Stefan and Matthes, Katja and Gabriel, Tina and Teufel, Andreas and Behrendt, Patrick and Markova, Antoaneta and Franck, Martin and Jenssen, Christian and Merkel, Daniel and Blank, Valentin and Karlas, Thomas and Seeger, Marcus and Eckart, Marius Andreas and Poralla, Lukas and Lehmann, Malte and Staudacher, Jonas and Ziesch, Matthias and Aland, Julia and Herzog, Lueder and Herzog, Angelika and Jürgensen, Christian and Fettweis, Gerhard Paul and Hampe, Jochen and Herzog, Moritz},
title = {Evaluation of image quality in five different handheld ultrasound devices and analysis of various impact factors},
journal = {Ultraschall in der Medizin - European Journal of Ultrasound},
year = {2026},
doi = {10.1055/a-2809-7627}
}
EUVIMED