Image Quality of Handheld Ultrasound Devices for Lung Ultrasound in Adult Patients with Pneumonia

Simon Micheelsen Norlén iD, Morten H Lorentzen, Christian B Laursen iD, Rahul Bhatnagar iD, Stefan Posth, Søren Helbo Skaarup, Pia I Pietersen iD

Journal of Ultrasound in Medicine · 24.09.2026

DOI: 10.1002/jum.70494 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Ultrasound in Medicine, online erschienen am 24.09.2026. Verfasst von Simon Micheelsen Norlén und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Odense University Hospital, Odense, Denmark. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

Handheld ultrasound devices (HHUDs) have increased accessibility and convenience of ultrasonography, but also raised concern regarding image quality and sufficiency for clinical use. This explorative study aims to compare image quality between 3 commercially available HHUDs in patients with pneumonia, and to compare expert agreement on image quality and sonopathological findings.

Methods

Patients were scanned with 3 different HHUDs. Clips were stored, anonymized, duplicated, and randomized. Three blinded, experienced thoracic ultrasound operators independently rated the clips for image quality (1–5), sufficiency for clinical use (1–3), and presence of sonopathological findings.

Results

Eleven patients were scanned, yielding 38 ultrasound clips and 76 ratings per rater. Overall image quality was 3.60 (95% CI 3.40–3.80) for Lumify , 3.32 (95% CI 3.11–3.53) for Vscan Air , and 2.78 (95% CI 2.55–3.01) for Butterfly iQ+ . Overall agreement on image quality ratings was low (27.6%). Interrater agreement for specific sonopathological findings ranged from slight for air bronchograms (kappa 0.15–0.34) to substantial for pleural effusions (kappa 0.58–0.84).

Conclusion

This explorative study demonstrated a higher image quality in Vscan Air and Lumify compared to Butterfly iQ+ in patients with pneumonia. However, interrater agreement on image quality was low, while interrater agreement on sonopathological findings was moderate.

Schlagwörter

Stichwörter der Autor:innen

handheldimage qualitylungspneumoniaultrasound

Autor:innen und Institutionen

  • Department of Radiology, Odense University Hospital, Odense, Denmark.
  • Department of Anaesthesia and Intensive Care, Odense University Hospital, Odense, Denmark.
  • Department of Respiratory Medicine, Odense University Hospital, Odense, Denmark.
  • Centre for Advanced Lung Cancer Diagnostics (CALU), Odense University Hospital, Odense, Denmark.
  • Odense Respiratory Research Unit (ODIN), University of Southern Denmark, Odense, Denmark.
  • Academic Respiratory Unit, University of Bristol, Bristol, UK.

ORCID-Profile: Simon Micheelsen Norlén, Christian B Laursen, Rahul Bhatnagar, Pia I Pietersen

Zitieren

APA

Norlén, S. M., Lorentzen, M. H., Laursen, C. B., Bhatnagar, R., Posth, S., Skaarup, S. H., & Pietersen, P. I. (2026). Image Quality of Handheld Ultrasound Devices for Lung Ultrasound in Adult Patients with Pneumonia. Journal of Ultrasound in Medicine. https://doi.org/10.1002/jum.70494

Vancouver

Norlén SM, Lorentzen MH, Laursen CB, Bhatnagar R, Posth S, Skaarup SH, et al. Image Quality of Handheld Ultrasound Devices for Lung Ultrasound in Adult Patients with Pneumonia. Journal of Ultrasound in Medicine. 2026. doi:10.1002/jum.70494

BibTeX
@article{Norln2026,
  author = {Norlén, Simon Micheelsen and Lorentzen, Morten H and Laursen, Christian B and Bhatnagar, Rahul and Posth, Stefan and Skaarup, Søren Helbo and Pietersen, Pia I},
  title = {Image Quality of Handheld Ultrasound Devices for Lung Ultrasound in Adult Patients with Pneumonia},
  journal = {Journal of Ultrasound in Medicine},
  year = {2026},
  doi = {10.1002/jum.70494}
}
Kostenlos starten →