Objective

To evaluate accuracy and feasibility of a handheld ultrasound machine for measuring carotid artery intima-media thicknesses (CIMT) and hemodynamic parameters by different expertise levels of ultrasound operators.

Methods

The operators were divided into three groups based on the level of their medical expertise: ultrasound technician, sonographer, and nursing staff. Operators from each group measured the CIMT and hemodynamic parameters of 25 volunteers using both handheld ultrasound and a conventional ultrasound machine. The reliability and reproducibility of handheld ultrasound measurements of CIMT and hemodynamic parameters (peak systolic velocity (PSV), end-diastolic velocity (EDV)) in operators were analyzed.

Results

After a period of training, there was no statistically significant difference between the mean CIMT measured using handheld ultrasound among the three operators (0.45 ± 0.09 mm, 0.50 ± 0.07 mm, 0.46 ± 0.08 mm, P> 0.05, respectively), as well as PSV (83.30 ± 15.42 cm/s, 76.28 ± 13.26 cm/s, 81.12 ± 21.21 cm/s, P> 0.05, respectively) and EDV (21.04 ± 4.12 cm/s, 21.87 ± 5.05 cm/s, 20.17 ± 5.90 cm/s, respectively, P> 0.05). Furthermore, there was a good repeatability and consistent of handheld ultrasound device in measuring mean CIMT in the ultrasound technician and sonographer groups (r = 0.662, 0.691, respectively, P < 0.01).

Conclusions

Under the premise of proper training, handheld ultrasound systems are feasible for rapid and primary assessment of carotid artery by operators with different levels of expertise.

Zitieren

APA

Yuzhou, S., Lin, J., Lei, S., Mengmeng, C., Desheng, S., Li, L., & Zhaojun, L. (2024). Can Different Expertise Levels of Ultrasound Operators Accurately Screen with Handheld Ultrasound?. ADVANCED ULTRASOUND IN DIAGNOSIS AND THERAPY, 8(3), 116. https://doi.org/10.37015/audt.2023.230046

Vancouver

Yuzhou S, Lin J, Lei S, Mengmeng C, Desheng S, Li L, et al. Can Different Expertise Levels of Ultrasound Operators Accurately Screen with Handheld Ultrasound?. ADVANCED ULTRASOUND IN DIAGNOSIS AND THERAPY. 2024;8(3):116. doi:10.37015/audt.2023.230046

BibTeX
@article{Yuzhou2024,
  author = {Yuzhou, Shen and Lin, Jin and Lei, Sha and Mengmeng, Cao and Desheng, Sun and Li, Liu and Zhaojun, Li},
  title = {Can Different Expertise Levels of Ultrasound Operators Accurately Screen with Handheld Ultrasound?},
  journal = {ADVANCED ULTRASOUND IN DIAGNOSIS AND THERAPY},
  year = {2024},
  volume = {8},
  number = {3},
  pages = {116},
  doi = {10.37015/audt.2023.230046}
}
Kostenlos starten →