Effect of Plate–Bone Distance and Screw Number on 2.0-mm Locking Construct Torsional Stiffness and Plate Strain in a Diaphyseal Fracture Gap Model

Alecia Evans, Mark Glyde iD, Giselle Hosgood, Rob Day

Veterinary and Comparative Orthopaedics and Traumatology · 01.07.2026 · 39 (04) · 241-247

DOI: 10.1055/a-2749-6666 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Veterinary and Comparative Orthopaedics and Traumatology, erschienen am 01.07.2026. Verfasst von Alecia Evans und 3 weiteren Autor:innen. Beteiligte Einrichtung: The Animal Hospital at Murdoch University, Perth, Australia. Hauptthemen laut MEDLINE: Bone Plates, Bone Screws, Fracture Fixation, Internal, Fractures, Bone. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 33 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

To determine the effect of plate–bone distance and screw number on 2.0-mm locking compression plate (LCP) torsional stiffness and plate strain for a fixed working length in a diaphyseal fracture gap model. In vitro biomechanical study. The LCP was tested with nine different combinations of plate–bone distance (PBD) (1, 1.5, and 3 mm) and screw number (SN) (two, three, or four screws per fragment) across the same working length. Six replicates per group were tested. Constructs were subjected to quasi-static, non-destructive torsion producing overall construct stiffness data (Nm/degree). Plate surface strain data were recorded from three-dimensional digital image correlation during testing for all constructs. A greater PBD and a low SN resulted in lower construct stiffness (p = 0.015, and P > 0.0001, respectively) in torsion. Across all PBD, there was significantly less stiffness for constructs with two screws per fragment, compared with three screws or four screws. No significant difference was detected in plate strain for any combination of PBD and SN. In this model, incrementally increasing the PBD and decreasing the SN resulted in less construct stiffness in torsion, with marked reduction in torsional stiffness in constructs with two screws per fragment compared with three or four screws per fragment.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Bone PlatesBone ScrewsFracture Fixation, InternalFractures, BoneAnimalsBiomechanical PhenomenaDiaphyses

Autor:innen und Institutionen

  • The Animal Hospital at Murdoch University, Perth, Australia.
  • Health Technology Management Unit, Royal Perth Hospital, Perth, Australia.

ORCID-Profile: Mark Glyde

Zitieren

APA

Evans, A., Glyde, M., Hosgood, G., & Day, R. (2026). Effect of Plate–Bone Distance and Screw Number on 2.0-mm Locking Construct Torsional Stiffness and Plate Strain in a Diaphyseal Fracture Gap Model. Veterinary and Comparative Orthopaedics and Traumatology, 39(04), 241-247. https://doi.org/10.1055/a-2749-6666

Vancouver

Evans A, Glyde M, Hosgood G, Day R. Effect of Plate–Bone Distance and Screw Number on 2.0-mm Locking Construct Torsional Stiffness and Plate Strain in a Diaphyseal Fracture Gap Model. Veterinary and Comparative Orthopaedics and Traumatology. 2026;39(04):241-247. doi:10.1055/a-2749-6666

BibTeX
@article{Evans2026,
  author = {Evans, Alecia and Glyde, Mark and Hosgood, Giselle and Day, Rob},
  title = {Effect of Plate–Bone Distance and Screw Number on 2.0-mm Locking Construct Torsional Stiffness and Plate Strain in a Diaphyseal Fracture Gap Model},
  journal = {Veterinary and Comparative Orthopaedics and Traumatology},
  year = {2026},
  volume = {39},
  number = {04},
  pages = {241--247},
  doi = {10.1055/a-2749-6666}
}
Kostenlos starten →